Taperloc Complete Micro Primary Femoral 51-106110Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Taperloc Complete Micro Primary Femoral (35)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00880304514294), received between Jun 17, 2022 and Sep 25, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert2
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Ambulation or Postural Difficulties1
  • Failure of Implant1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING IMPLANTATION, THE STEM COATING WAS TOO THICK AND WAS CAUSING THE STEM TO SIT PROUD. THE SURGEON REMOVED THE HEAD AND TAPER SLEEVE AND HAD TO RE-BROACH TWO SIZES ABOVE THE IMPLANT THAT WAS DETERMINED AS THE CORRECT SIZE. EVENTUALLY FIT THE STEM INTO THE CANAL. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20300971 ↗
  • MalfunctionJun 17, 2022Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING AN INITIAL HIP ARTHROPLASTY, THE STEM WOULD NOT SEAT INTO THE FEMUR AFTER SEVERAL BROACHING ATTEMPTS. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14734286 ↗
  • InjuryJun 17, 2022Migration; Migration

    IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 3 MONTHS POST IMPLANTATION DUE TO PAIN AND DIFFICULTY WALKING. DURING THE PROCEDURE, THE FEMORAL COMPONENT WAS NOTED TO HAVE SUBSIDED. THE CUP AND LINER REMAINED IMPLANTED WHILE THE FEMORAL COMPONENTS WERE REMOVED AND REPLACED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Pain

    MAUDE report 14734021 ↗