G7 Acetabular Liner 010000928Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of G7 Acetabular Liner (85)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00880304527041), received between Mar 3, 2020 and Sep 24, 2026 — classified as 8 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated5
- Difficult to Insert2
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
- Off-Label Use1
- Patient-Device Incompatibility1
- Unstable1
Patient problems
- Joint Dislocation5
- Aseptic loosening1
- Failure of Implant1
- Hypersensitivity/Allergic reaction1
- Insufficient Information1
- Joint Laxity1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that at an unknown time post-implantation, the patient underwent a hip revision due to unknown reasons. Head and liner were exchanged.Attempts have been made and no further information has been provided.
Patient problems: Insufficient Information
MAUDE report 50882240 ↗InjuryFeb 9, 2026Patient-Device Incompatibility
IT WAS REPORTED THAT APPROXIMATELY TWO YEARS POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO INSTABILITY, LOOSENING, AND A POSSIBLE METAL ALLERGY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Aseptic loosening; Hypersensitivity/Allergic reaction; Joint Laxity
MAUDE report 24298573 ↗InjuryApr 2, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT APPROXIMATELY EIGHT MONTHS POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO MULTIPLE DISLOCATIONS DUE TO SUBOPTIMAL CUP POSITIONING. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Joint Dislocation
MAUDE report 21750643 ↗InjuryMay 8, 2024Off-Label Use; Device Dislodged or Dislocated; Off-Label Use; Device Dislodged or Dislocated
IT WAS REPORTED THE PATIENT UNDERWENT A PRIMARY HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT DEVELOPED AN INFECTION AND HAD A WASHOUT TWO WEEKS (2) POST IMPLANTATION. THE PATIENT UNDERWENT A REVISION SURGERY ELEVEN (11) DAYS AND THE COMPETITOR HEAD AND ZIMMER BIOMET LINER WERE EXCHANGED. THE PATIENT WAS REVISED FOR A SECOND TIME TWO YEARS LATER, DUE TO DISLOCATION. COMPETITOR HEAD AND ZIMMER BIOMET LINER WERE EXCHANGED. NO KNOWN CONTRIBUTING CONDITIONS RELATED TO THE EVENT.
Patient problems: Joint Dislocation
MAUDE report 19272662 ↗InjuryJul 14, 2023Unstable; Unstable
THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17325365 ↗InjuryJun 27, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THE PATIENT UNDERWENT A HIP REVISION APPROXIMATELY TWO DAYS POST IMPLANTATION DUE TO A DISLOCATION. NO ADDITIONAL INFORMATION.
Patient problems: Joint Dislocation
MAUDE report 14829362 ↗InjuryNov 2, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Joint Dislocation
MAUDE report 12737888 ↗InjurySep 3, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Joint Dislocation
MAUDE report 12420840 ↗MalfunctionJun 1, 2021Difficult to Insert; Difficult to Insert
HOLD FOR MH IT WAS REPORTED THAT IT WAS NOT POSSIBLE TO LOCK THE LINER. NO ADVERSE EVENT HAS BEEN REPORTED AS A RESULT OF THIS MALFUNCTION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11912452 ↗MalfunctionMar 3, 2020Difficult to Insert; Material Deformation; Difficult to Insert; Material Deformation
IT WAS REPORTED THAT DURING A HIP PROCEDURE, A LINER WAS DRIVEN INTO A SHELL AND WAS UNABLE TO BE SECURED. ONCE THE LINER WAS TAKEN OUT, A THREAD-LIKE MATERIAL CAME OUT ALLOWING THE LINER TO SEAT WITHIN THE SHELL. THERE WAS NO REPORTED HARM OR INJURY TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9779589 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page