This Biomet G7 acetabular component is used in hip replacement. The 23 reports mostly describe threaded inserter and impactor parts breaking, deforming or cross-threading, making them hard to attach or remove, with no patient involvement mentioned. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00880304532694), received between Jun 12, 2020 and May 27, 2026 — classified as 23 malfunctions.

The most frequently reported specific device problem is “Fracture” (7 reports).

3 other recalls cover devices cleared under the same FDA submission (K121874); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture7
  • Material Deformation5
  • Separation Failure5
  • Damaged Thread3
  • Difficult to Remove3
  • Degraded1
  • Difficult to Insert1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2026Fracture

    IT WAS REPORTED THAT THREADS FROM A STRAIGHT SHELL INSERTER BROKE OFF AND REMAINED WITHIN THE ACETABULAR CUP APEX HOLE DURING USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25303197 ↗
  • MalfunctionSep 5, 2025Damaged Thread; Damaged Thread

    IT WAS REPORTED THAT THE ACETABULAR SHELL INSERTER HANDLE DID NOT SECURELY ATTACH DUE TO THE THREADS. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22976205 ↗
  • MalfunctionJun 26, 2025Damaged Thread; Damaged Thread

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22335333 ↗
  • MalfunctionOct 9, 2024Difficult to Insert; Damaged Thread; Difficult to Insert; Damaged Thread

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20409170 ↗
  • MalfunctionAug 14, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THE MODULAR HANDLE CROSS THREADED CAUSING THE SHELL TRIALS TO BE DIFFICULT TO REMOVE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19983743 ↗
  • MalfunctionJul 5, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE LINER IMPACTOR THREADS WERE CROSS THREADED AND IT CAUSED G7 IMPACTORS TO BECOME DAMAGED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19679174 ↗
  • MalfunctionMar 19, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18935040 ↗
  • MalfunctionMar 8, 2024Material Deformation; Separation Problem; Material Deformation; Separation Problem

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18864989 ↗
  • MalfunctionNov 29, 2023Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT WHEN THE SURGEON TRIED TO LOOSEN THE IMPACTOR FROM THE SHELL, IT WAS UNABLE TO BE REMOVED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18226517 ↗
  • MalfunctionJul 13, 2023Fracture; Material Deformation; Fracture; Material Deformation

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17316862 ↗
  • MalfunctionMar 30, 2023Separation Failure; Separation Failure

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16645971 ↗
  • MalfunctionNov 4, 2022Separation Failure; Material Deformation; Separation Failure; Material Deformation

    IT WAS REPORTED THAT DURING A HIP PROCEDURE, THE SURGEON SURGEON IMPACTED THE CUP BUT WAS NOT SATISFIED WITH THE POSITION SO THE IMPACTOR WAS ATTACHED AGAIN. AFTER SUCCESSFULLY REPOSITIONING THE CUP, THE SURGEON WAS NOT ABLE TO GET THE IMPACTOR HANDLE TO RELEASE FROM THE CUP. POTENTIALLY THE THREADS WERE DAMAGED. THE IMPACTOR WAS EVENTUALLY ABLE TO BE REMOVED FROM THE CUP WITHOUT FURTHER INCIDENT. THERE WAS NO HARM OR INJURY TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15729337 ↗
  • MalfunctionAug 25, 2022Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT DURING AN INITIAL HIP ARTHROPLASTY, THE CUP INSERTER WOULD NOT DISENGAGE FROM THE CUP. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15298039 ↗
  • MalfunctionJun 21, 2022Separation Failure; Separation Failure

    IT WAS REPORTED THAT DURING A HIP PROCEDURE, THE SHELL INSERTER COULD NOT BE DISENGAGED FROM THE SHELL. ATTEMPTS TO UNSCREW THE HANDLE LED TO THE IMPLANT SPINNING AND LOSS OF PRIMARY FIXATION. THE SHELL WAS REMOVED FROM THE PATIENT AND A NEW SHELL AND INSERTER WERE USED TO COMPLETE THE SURGERY. THERE WAS NO REPORTED HARM OR INJURY TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14756387 ↗
  • MalfunctionMay 13, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL HIP ARTHROPLASTY THE THREADS OF THE INSERTER FRACTURED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION AND ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14397860 ↗
  • MalfunctionMay 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL HIP ARTHROPLASTY, THE THREADS OF THE INSERTER FRACTURED INTO THE CUP. THE SURGEON HAD TO REMOVE THE CUP FROM THE PATIENT'S BONE WITH ANOTHER INSERTER. WHILE DOING THIS, THE SECOND INSERTER'S THREADS STRIPPED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION AND ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14366112 ↗
  • MalfunctionMar 23, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE INSTRUMENT FRACTURED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13870376 ↗
  • MalfunctionDec 9, 2021Separation Failure; Separation Failure

    IT WAS REPORTED THAT THE G7 CUP WAS STUCK ON INSERTER HANDLE SO CUP WAS HARD TO SEAT WITH PRESS FIT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12969619 ↗
  • MalfunctionMay 19, 2021Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11847716 ↗
  • MalfunctionApr 14, 2021Difficult to Remove; Difficult to Remove

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11667276 ↗
  • MalfunctionApr 5, 2021Separation Failure; Separation Failure

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11614479 ↗
  • MalfunctionApr 5, 2021Material Deformation; Material Deformation

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11612546 ↗
  • MalfunctionJun 12, 2020Degraded; Degraded

    NO ADDITIONAL INFORMATION ON REPORTED EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10147839 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121874. One submission often covers many models, so check the product and lot details in each record.

  • Z-0007-2023 · Aug 25, 2022 · Open, Classified
    G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
  • Z-1838-2014 · May 7, 2014 · Terminated
    G7 Acetabular System Positioning Guide Post, orthopedic surgical instrument for hip prosthesis. PN:110003500.
  • Z-1837-2014 · May 7, 2014 · Terminated
    G7 Acetabular System Positioning Guide Rod, orthopedic surgical instrument for hip prosthesis PN 110003458.

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →