Skip to content

G7 CURVED ACETABULAR SHELL INSERTER 110003453Recalls & reported problems

Biomet Orthopedics, LLC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00880304543324), received between Aug 5, 2024 and Feb 11, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

3 other recalls cover devices cleared under the same FDA submission (K121874); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Material Deformation1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2026Fracture

    IT WAS REPORTED THAT A CURVED IMPACTOR BROKE WHILE IMPACTING AN ACETABULAR CUP. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24313356 ↗
  • MalfunctionJan 22, 2026Fracture

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144700 ↗
  • MalfunctionJan 22, 2026Fracture

    IT WAS REPORTED THAT THE IMPACTOR PLATE BROKE OFF THE CURVED CUP INSERTER DURING USE, AND THE PROCEDURE WAS CONTINUED WITH A STRAIGHT VERSION. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT, AND NO ADDITIONAL SURGERY TIME WAS ENCOUNTERED. IT WAS REPORTED THAT NO FURTHER INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144607 ↗
  • MalfunctionFeb 12, 2025Fracture; Separation Problem; Fracture; Separation Problem

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21364684 ↗
  • MalfunctionNov 12, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE OFFSET CUP INSERTER WAS DAMAGED PRIOR TO SURGERY. THE INTERLOCKING SCREW OF THE DEVICE WAS CROSS THREADED AND COULD NOT BE USED IN SURGERY. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20662459 ↗
  • MalfunctionAug 5, 2024Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19908305 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121874. One submission often covers many models, so check the product and lot details in each record.

  • Z-0007-2023 · Aug 25, 2022 · Open, Classified
    G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
  • Z-1837-2014 · May 7, 2014 · Terminated
    G7 Acetabular System Positioning Guide Rod, orthopedic surgical instrument for hip prosthesis PN 110003458.
  • Z-1838-2014 · May 7, 2014 · Terminated
    G7 Acetabular System Positioning Guide Post, orthopedic surgical instrument for hip prosthesis. PN:110003500.

Other Hip, semi-constrained, cemented, metal/polymer + additive, cemented devices

All Hip, semi-constrained, cemented, metal/polymer + additive, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page