Skip to content

COMPREHENSIVE SHOULDER SYSTEM 113611Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00880304554122), received between Feb 21, 2023 and Feb 21, 2023 — classified as 1 injury.

1 FDA recall record name this device's identifier.

6 other recalls cover devices cleared under the same FDA submission (K060692); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 7, 2022Terminated

    The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

    Z-0671-2022 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Hypersensitivity/Allergic reaction1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED APPROXIMATELY 3 MONTHS POST IMPLANTATION DUE TO AN SUSPECTED NICKEL ALLERGY. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 16410305 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K060692. One submission often covers many models, so check the product and lot details in each record.

  • Z-1543-2020 · Feb 17, 2020 · Terminated
    Comprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component o
  • Z-0591-2019 · Oct 10, 2018 · Terminated
    Comprehensive Mini Stem, Item No. 113631:
  • Z-0980-2018 · Dec 8, 2017 · Terminated
    Humeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
  • Z-0979-2018 · Dec 8, 2017 · Terminated
    Humeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
  • Z-0299-2018 · Jun 5, 2017 · Terminated
    COMP PRIMARY STEM 8MM MINI, model # 113628
  • Z-1755-2015 · Apr 16, 2015 · Terminated
    Comprehensive Shoulder System, 4 mm, 122 mm long, Primary Shoulder Stem, Ti-6Al-4V, Standard Length Porous Plasma. Shoulder joint replacement components include humeral stems, hum

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.