COMPREHENSIVE SHOULDER SYSTEM 113617Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of Comprehensive Shoulder System (68)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00880304554184), received between Dec 28, 2023 and Dec 28, 2023 — classified as 1 injury.
1 FDA recall record name this device's identifier.
5 other recalls cover devices cleared under the same FDA submission (K060692); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Dec 8, 2017Class IITerminated
Zimmer Biomet is conducting a medical device recall for two lots of Comprehensive Humeral stems. The lots were commingled during the manufacturing process and were etched with incorrect lot numbers. The size difference between the two stem sizes is 1.1 inches. It is visually recognizable by the user that the box label does not match the contents of the box.
Z-0980-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Joint Laxity1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE DUE TO INSTABILITY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Joint Laxity
MAUDE report 18411131 ↗
FDA clearance and classification
Cleared through 510(k) K060692 on May 30, 2006 (decision: Substantially Equivalent). Submitted by Biomet Manufacturing Corp as “Comprehensive Primary Shoulder Stems”. The FDA review took 76 days. FDA record ↗
The FDA regulates product code MBF (prosthesis, shoulder, semi-constrained, metal/polymer, uncemented) under 21 CFR 888.3670 as a Class II device. 21 CFR 888.3670 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K060692. One submission often covers many models, so check the product and lot details in each record.
- Z-1543-2020 · Feb 17, 2020 · TerminatedComprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component o
- Z-0591-2019 · Oct 10, 2018 · TerminatedComprehensive Mini Stem, Item No. 113631:
- Z-0979-2018 · Dec 8, 2017 · TerminatedHumeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
- Z-0299-2018 · Jun 5, 2017 · TerminatedCOMP PRIMARY STEM 8MM MINI, model # 113628
- Z-1755-2015 · Apr 16, 2015 · TerminatedComprehensive Shoulder System, 4 mm, 122 mm long, Primary Shoulder Stem, Ti-6Al-4V, Standard Length Porous Plasma. Shoulder joint replacement components include humeral stems, hum
Common questions about the COMPREHENSIVE SHOULDER SYSTEM 113617
Has the COMPREHENSIVE SHOULDER SYSTEM 113617 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0980-2018 (Class II), began on Dec 8, 2017: Zimmer Biomet is conducting a medical device recall for two lots of Comprehensive Humeral stems. The lots were commingled during the manufacturing process and were etched with incorrect lot numbers. The size… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the COMPREHENSIVE SHOULDER SYSTEM 113617?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 28, 2023. That report lists patient problem “Joint Laxity” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the COMPREHENSIVE SHOULDER SYSTEM 113617?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the COMPREHENSIVE SHOULDER SYSTEM 113617 FDA approved?
It was cleared through 510(k) K060692 on May 30, 2006. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the COMPREHENSIVE SHOULDER SYSTEM 113617 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the COMPREHENSIVE SHOULDER SYSTEM 113617, and is it still sold?
The FDA's device database lists Biomet Orthopedics, LLC as the company responsible for it (brand COMPREHENSIVE SHOULDER SYSTEM, model 113617). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED devices
- COMPREHENSIVE SHOULDER SYSTEM 113629Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall2 reports
- Comprehensive® Shoulder System 113607Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall1 report
- BIO-MODULAR SHOULDER SYSTEM 113863Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall0 reports
- BIO-MODULAR SHOULDER SYSTEM 113867Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall0 reports
- COPELAND SHOULDER SYSTEM 11-114648Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall0 reports
- COPELAND SHOULDER SYSTEM 11-114641Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall0 reports
- COPELAND SHOULDER SYSTEM 11-114644Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall0 reports
- COPELAND SHOULDER SYSTEM 11-114643Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page