Vanguard Knee System CP113642Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of Vanguard Knee System (147)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00880304653771), received between Feb 25, 2020 and Apr 21, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K080204, K171054); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth1
- Noise, Audible1
- Tear, Rip or Hole in Device Packaging1
- Unstable1
Patient problems
- Loss of Range of Motion1
- No Patient Involvement1
- Pain1
- Swelling/ Edema1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 21, 2022Unstable; Noise, Audible; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Noise, Audible; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED VIA A CLINICAL STUDY THAT A PATIENT UNDERWENT AN INITIAL RIGHT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, TEN MONTHS POST IMPLANTATION, THE PATIENT WAS REVISED DUE TO PAIN, STIFFNESS, SWELLING, CLICKING, AND INSTABILITY. A CT SCAN PRIOR TO THE REVISION NOTED THAT THE TIBIAL COMPONENT HAD INTERNALLY ROTATED. DURING THE PROCEDURE, THE SURGEON FOUND THAT THERE WAS CEMENT PRESENT ON THE BONE AND TIBIAL IMPLANT, WHICH WAS EASILY REMOVED. THE FEMORAL IMPLANT HAD NO CEMENT ATTACHED. THE PATELLAR COMPONENT WAS LEFT INTACT BUT ALL OTHER COMPONENTS WERE REVISED WITHOUT COMPLICATION.
Patient problems: Loss of Range of Motion; Pain; Swelling/ Edema
MAUDE report 14166365 ↗MalfunctionFeb 25, 2020Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Patient Involvement
MAUDE report 9752282 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080204, K171054. One submission often covers many models, so check the product and lot details in each record.
- Z-0938-2019 · Dec 21, 2018 · TerminatedVanguard Knee System Series-A Standard Patella sizes 34mm Item Number: 184766
- Z-2289-2018 · May 24, 2018 · TerminatedVanguard Complete Knee System; E1 Vanguard PS Tibial Bearing; Model Numbers EP-183664, EP-183642 Product Usage: The product is intended for use in knee joint replacement arthropla
- Z-2177-2018 · May 18, 2018 · TerminatedBiomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10) 385340 Product is used in Total
- Z-2183-2018 · May 11, 2018 · TerminatedVanguard Knee System PS Tibial Bearing, Part Number 183621 Intended for use in knee joint replacement procedures.
Other Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices
- iTotal® PS TPS111111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive327 reports
- ATTUNE 1506-00-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive287 reports
- ATTUNE 1506-00-008Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive233 reports
- iTotal® Cruciate Retaining (CR) Knee Replacement System TCR121111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive150 reports
- ATTUNE 1506-70-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall2 deaths74 reports
- ATTUNE 1516-40-506Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive121 reports
- ATTUNE 1506-70-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall70 reports
- ATTUNE 1516-40-605Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive120 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page