Skip to content

Comprehensive® Reverse Shoulder 180550Recalls & reported problems

Biomet Orthopedics, LLC · Class 2

All models of Comprehensive Reverse Shoulder (56)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Biomet Comprehensive reverse shoulder implant is a semi-constrained shoulder replacement. The 44 reports mostly describe the implant loosening within the bone, bone fracture, and parts detaching or wearing away, with patients reporting pain and bone loss that sometimes led to revision surgery. Two deaths and 41 injuries are reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
2
Injuries reported
40
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00880304677135), received between Nov 26, 2019 and Feb 23, 2026 — classified as 2 deaths, 40 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (13 reports).

19 other recalls cover devices cleared under the same FDA submission (K080642, K130390, K132239); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 4 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem20
  • Loosening of Implant Not Related to Bone-Ingrowth13
  • Detachment of Device or Device Component5
  • Fracture5
  • Unintended Movement4
  • Material Erosion3
  • Malposition of Device2
  • Unstable2

Patient problems

  • Pain13
  • Failure of Implant11
  • Inadequate Osseointegration6
  • Osteopenia/ Osteoporosis6
  • Bone Fracture(s)5
  • Erosion5
  • Metal Related Pathology4
  • Unspecified Infection4

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT UNDERWENT REVISION USING A CUSTOM IMPLANT APPROXIMATELY 1 YEAR AND 1 MONTH POST IMPLANTATION DUE TO THE PATIENT'S GLENOID BONE FRACTURING. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME

    Patient problems: Bone Fracture(s)

    MAUDE report 24415374 ↗
  • InjuryFeb 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT UNDERWENT REVISION USING A CUSTOM IMPLANT APPROXIMATELY 1 YEAR AND 1 MONTH POST IMPLANTATION DUE TO THE PATIENT'S GLENOID BONE FRACTURING. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME

    Patient problems: Bone Fracture(s)

    MAUDE report 24415332 ↗
  • InjuryNov 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED FROM A CLINICAL STUDY THAT A PATIENT HAD AN LEFT TOTAL SHOULDER. SUBSEQUENTLY, THE PATIENT HAS REPORTED AN INCREASE IN SHOULDER PAIN, AT TIMES SEVERE, WITH AN ONSET DATE OF APPROXIMATELY 6 YEARS POSTOP. IT IS UNKNOWN IF REVISION SURGERY IS RECOMMENDED OR PLANNED. NO INTRAOPERATIVE COMPLICATIONS. NO FURTHER INFORMATION REGARDING TREATMENT OR INTERVENTION HAS BEEN PROVIDED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Pain

    MAUDE report 23610473 ↗
  • InjuryNov 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED FROM A CLINICAL STUDY THAT A PATIENT HAD AN LEFT TOTAL SHOULDER. SUBSEQUENTLY, THE PATIENT HAS REPORTED AN INCREASE IN SHOULDER PAIN, AT TIMES SEVERE, WITH AN ONSET DATE OF APPROXIMATELY 6 YEARS POSTOP. IT IS UNKNOWN IF REVISION SURGERY IS RECOMMENDED OR PLANNED. NO INTRAOPERATIVE COMPLICATIONS. NO FURTHER INFORMATION REGARDING TREATMENT OR INTERVENTION HAS BEEN PROVIDED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Pain

    MAUDE report 23610469 ↗
  • InjuryAug 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY 6 MONTHS POST IMPLANTATION DUE TO IMPLANT DISASSOCIATION AND MASSIVE GLENOID EROSION. THE PATIENT WAS CONVERTED TO A HEMIARTHROPLASTY AND WILL BE IMPLANTED WITH A CUSTOM BASEPLATE ON AN UNKNOWN DATE IN THE FUTURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Erosion

    MAUDE report 22903774 ↗
  • DeathJul 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PASSED AWAY IN RECOVERY FOLLOWING A REVERSE TOTAL SHOULDER REPLACEMENT WITH THE COMPREHENSIVE SYSTEM. AS FAR AS THE SURGEON WAS CONCERNED, THE OPERATION WAS COMPLETED AS EXPECTED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22530379 ↗
  • DeathJul 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PASSED AWAY IN RECOVERY FOLLOWING A REVERSE TOTAL SHOULDER REPLACEMENT WITH THE COMPREHENSIVE SYSTEM. AS FAR AS THE SURGEON WAS CONCERNED, THE OPERATION WAS COMPLETED AS EXPECTED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22530377 ↗
  • InjuryJun 26, 2025Material Erosion; Material Erosion

    IT WAS REPORTED A PATIENT HAD AN INITIAL LEFT TOTAL ANATOMICAL SHOULDER ARTHROPLASTY ON AN UNKNOWN DATE WITH UNKNOWN PRODUCTS. APPROXIMATELY ONE (1) YEAR AND NINE (9) MONTHS AGO, THE PATIENT WAS REVISED TO A COMPREHENSIVE REVERSE TOTAL SHOULDER ARTHROPLASTY DUE TO GLENOID LOOSENING AND FAILED ROTATOR CUFF. SUBSEQUENTLY, THE PATIENT HAD FALLEN APPROXIMATELY ONE (1) YEAR AND FOUR (4) MONTHS POST-IMPLANTATION AND THEN BEGAN TO HAVE PAIN. RADIOGRAPHIC IMAGING DISPLAYED MALALIGNMENT OF THE GLENOSPHERE, FRACTURED SCREW, LUCENCY AROUND THE INFERIOR LOCKING SCREW, RADIO-DENSE BODIES SUPERIOR IN THE JOINT, AND LIKELY BONY FRAGMENTS INFERIORLY. APPROXIMATELY ONE (1) MONTH AGO, THE PATIENT UNDERWENT A REVISION WHERE A LITTLE BIT OF METALLOSIS OF THE SOFT TISSUE AROUND THE JOINT CAPSULE WAS ENCOUNTERED AND THE CENTRAL SCREW WAS FRACTURED. CANCELLOUS ALLOGRAFT WAS IMPLANTED AND THE PATIENT WAS CONVERTED TO A HEMIARTHROPLASTY WITH PLANS FOR A POSSIBLE REVISION IN THE FUTURE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Aseptic loosening; Metal Related Pathology; Pain; Radiolucent Lines

    MAUDE report 22342444 ↗
  • InjuryJun 26, 2025Material Erosion; Material Erosion

    IT WAS REPORTED A PATIENT HAD AN INITIAL LEFT TOTAL ANATOMICAL SHOULDER ARTHROPLASTY ON AN UNKNOWN DATE WITH UNKNOWN PRODUCTS. APPROXIMATELY ONE (1) YEAR AND NINE (9) MONTHS AGO, THE PATIENT WAS REVISED TO A COMPREHENSIVE REVERSE TOTAL SHOULDER ARTHROPLASTY DUE TO GLENOID LOOSENING AND FAILED ROTATOR CUFF. SUBSEQUENTLY, THE PATIENT HAD FALLEN APPROXIMATELY ONE (1) YEAR AND FOUR (4) MONTHS POST-IMPLANTATION AND THEN BEGAN TO HAVE PAIN. RADIOGRAPHIC IMAGING DISPLAYED MALALIGNMENT OF THE GLENOSPHERE, FRACTURED SCREW, LUCENCY AROUND THE INFERIOR LOCKING SCREW, RADIO-DENSE BODIES SUPERIOR IN THE JOINT, AND LIKELY BONY FRAGMENTS INFERIORLY. APPROXIMATELY ONE (1) MONTH AGO, THE PATIENT UNDERWENT A REVISION WHERE A LITTLE BIT OF METALLOSIS OF THE SOFT TISSUE AROUND THE JOINT CAPSULE WAS ENCOUNTERED AND THE CENTRAL SCREW WAS FRACTURED. CANCELLOUS ALLOGRAFT WAS IMPLANTED AND THE PATIENT WAS CONVERTED TO A HEMIARTHROPLASTY WITH PLANS FOR A POSSIBLE REVISION IN THE FUTURE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Aseptic loosening; Metal Related Pathology; Pain; Radiolucent Lines

    MAUDE report 22342443 ↗
  • InjuryMay 28, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT SHOULDER ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT WAS CONSIDERED FOR A PMI PRODUCT AND REVISED DUE TO LOOSENING AND INSTABILITY OF THE IMPLANT. DURING THE REVISION, IT WAS NOTED THAT THE CENTRAL SCREW WAS FRACTURED. ALL COMPONENTS WERE REMOVED AND REPLACED; HOWEVER, A PIECE OF THE FRACTURED CENTRAL SCREW COULD NOT BE REMOVED AS IT WAS RETAINED IN THE GLENOID.

    Patient problems: Aseptic loosening

    MAUDE report 22101937 ↗
  • InjuryMar 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED FROM A CLINICAL STUDY THAT A PATIENT HAD A RIGHT REVERSE TOTAL SHOULDER ARTHROPLASTY PERFORMED APPROXIMATELY 5.5 YEARS AGO. THE PATIENT DID VERY WELL THROUGHOUT THE STUDY UNTIL REPORTING SEVERE PAIN AND INSTABILITY WITH ACTIVITIES AT THE 5 YEAR FOLLOW UP. NO TREATMENT OR INTERVENTION HAS BEEN REPORTED TO DATE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Pain

    MAUDE report 21734201 ↗
  • InjuryMar 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED FROM A CLINICAL STUDY THAT A PATIENT HAD A RIGHT REVERSE TOTAL SHOULDER ARTHROPLASTY PERFORMED APPROXIMATELY 5.5 YEARS AGO. THE PATIENT DID VERY WELL THROUGHOUT THE STUDY UNTIL REPORTING SEVERE PAIN AND INSTABILITY WITH ACTIVITIES AT THE 5 YEAR FOLLOW UP. NO TREATMENT OR INTERVENTION HAS BEEN REPORTED TO DATE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Pain

    MAUDE report 21734195 ↗
  • InjuryMar 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS FURTHER REPORTED THAT THE PATIENT WAS GETTING DRESSED AND FELT A POPPING SENSATION WHICH LEAD TO THE REVISION PROCEDURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 21504804 ↗
  • InjuryFeb 12, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY DUE TO A LOOSENING OF THE DEVICE. THERE WAS HARM OR INJURY TO PATIENT AS THE PATIENT HAD TO UNDERGO ADDITIONAL SURGICAL PROCEDURE. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 21369889 ↗
  • InjuryJan 7, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SHOULDER ARTHROPLASTY APPROXIMATELY ONE MONTH AGO. SUBSEQUENTLY, THE PATIENT WAS REVISED 8 DAYS LATER DUE TO INSTABILITY AND LOOSENING. PATIENT HAD NEW STEM AND TAPER PUT IN. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 21095603 ↗
  • InjuryJan 7, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SHOULDER ARTHROPLASTY APPROXIMATELY ONE MONTH AGO. SUBSEQUENTLY, THE PATIENT WAS REVISED 8 DAYS LATER DUE TO INSTABILITY AND LOOSENING. PATIENT HAD NEW STEM AND TAPER PUT IN. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 21095594 ↗
  • InjuryDec 3, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Inadequate Osseointegration; Osteopenia/ Osteoporosis

    MAUDE report 20831452 ↗
  • InjuryDec 3, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Inadequate Osseointegration; Osteopenia/ Osteoporosis

    MAUDE report 20831432 ↗
  • InjuryNov 1, 2024Detachment of Device or Device Component; Unintended Movement; Detachment of Device or Device Component; Unintended Movement

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISASSOCIATION. IT WAS ALSO REPORTED THAT AN UNKNOWN AMOUNT OF SCREWS WERE LOOSE FROM THE BASE PLATE. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20587301 ↗
  • InjuryNov 1, 2024Detachment of Device or Device Component; Unintended Movement; Detachment of Device or Device Component; Unintended Movement

    IT WAS REPORTED THAT THAT THE PATIENT WAS REVISED DUE TO DISASSOCIATION. IT WAS ALSO REPORTED THAT AN UNKNOWN AMOUNT OF SCREWS WERE LOOSE FROM THE BASE PLATE. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20587294 ↗
  • InjuryNov 1, 2024Detachment of Device or Device Component; Unintended Movement; Detachment of Device or Device Component; Unintended Movement

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISASSOCIATION. IT WAS ALSO REPORTED THAT AN UNKNOWN NUMBER OF SCREWS WERE LOOSE FROM THE BASE PLATE. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20587283 ↗
  • InjuryNov 1, 2024Detachment of Device or Device Component; Unintended Movement; Detachment of Device or Device Component; Unintended Movement

    IT WAS REPORTED THAT THAT THE PATIENT WAS REVISED DUE TO DISASSOCIATION. IT WAS ALSO REPORTED THAT AN UNKNOWN AMOUNT OF SCREWS WERE LOOSE FROM THE BASE PLATE. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20587277 ↗
  • MalfunctionOct 17, 2024Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THAT DURING SURGERY THAT THE ANCHOR WOULD NOT DEPLOY. THERE WAS NO HARM OR INJURY TO PATIENT. A DELAY OF APPROXIMATELY 15 MINUTES WAS REPORTED WHILE ANOTHER DEVICE WAS RETRIEVED. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20474239 ↗
  • InjurySep 27, 2024Fracture; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A PATIENT EXPERIENCED BASEPLATE LOOSENING FROM THE GLENOID FOLLOWING AN INITIAL ARTHROSCOPY PROCEDURE. THE SURGEON INDICATED THE PATIENT'S ANATOMY, WITH VERY LITTLE BONE IN THE GLENOID, LIKELY CAUSED THE FAILURE. IMAGING CONFIRMED THIS LOOSENING APPROXIMATELY FIFTEEN (15) MONTHS LATER. ADDITIONALLY, ONE (1) OF THE PERIPHERAL SCREWS WAS FOUND TO BE FRACTURED DURING RADIOGRAPH EVALUATION. A FUTURE REVISION IS PLANNED ON AN UNKNOWN DATE. NO OTHER INTERVENTIONS OR HARMS HAVE BEEN REPORTED DUE TO THIS EVENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 20321584 ↗
  • InjuryMay 21, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Bacterial Infection; Failure of Implant; Pain

    MAUDE report 19358803 ↗

FDA clearance and classification

Cleared through 510(k) K130390 on Oct 9, 2013 (decision: Substantially Equivalent). Submitted by Biomet Manufacturing, Inc. as “Comprehensive Convertible Glenoid”. The FDA review took 236 days. FDA record ↗

Cleared through 510(k) K132239 on Sep 6, 2013 (decision: Substantially Equivalent). Submitted by Biomet Manufacturing Corp as “Comprehensive Reverse Shoulder Screws”. The FDA review took 50 days. FDA record ↗

Cleared through 510(k) K080642 on Jul 9, 2008 (decision: Substantially Equivalent). Submitted by Biomet, Inc. as “Comprehensive Reverse Shoulder”. The FDA review took 125 days. FDA record ↗

The FDA regulates product code PAO (prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K080642, K130390, K132239. One submission often covers many models, so check the product and lot details in each record.

  • Z-1131-2021 · Jan 15, 2021 · Terminated
    Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
  • Z-0619-2021 · Nov 25, 2020 · Terminated
    Comprehensive Reverse Shoulder System, Item numbers 115394, 115395, 115396, 115398, 180550, 180551, 180552, 180553, 180554, 180555, 180556, 180557, 180558, 180559, 180560, 180562
  • Z-2144-2020 · Apr 17, 2020 · Terminated
    Extremities Item Number: 1)113954 Hybrid Glenoid Glenoid Base, 4 MM 2) XL-115366 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 41 MM, Standard 3) XL-115364 Comp
  • Z-1750-2020 · Mar 13, 2020 · Terminated
    Comprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the gle
  • Z-1859-2017 · Feb 17, 2017 · Terminated
    various polyethylene implants Extremities, Shoulders and Elbows BioModular Shoulder System; Comprehensive Shoulder System; Comprehensive Reverse Shoulder System Product Usage: Sh
  • Z-1103-2017 · Dec 20, 2016 · Terminated
    Comprehensive Reverse Shoulder System Humeral Tray Model 115340
  • Z-1548-2013 · May 15, 2013 · Terminated
    Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
  • Z-2545-2011 · Apr 22, 2011 · Terminated
    Comprehensive reverse shoulder humeral tray with locking ring, 44 mm standard, sterile, single use, REF 115340, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Sho
  • Z-2547-2011 · Apr 22, 2011 · Terminated
    Custom solar comprehensive reverse humeral tray with lock ring 44 mm +5mm, sterile, REF CP561659, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product
  • Z-2549-2011 · Apr 22, 2011 · Terminated
    Custom comprehensive reverse humeral tray set w/ lock rings 44 mm + 15 mm & + 18 mm sterile, REF CP561705, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder
  • Z-2546-2011 · Apr 22, 2011 · Terminated
    Custom bi-angular to reverse shoulder humeral tray 44 mm sterile, REF CP561548, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated fo
  • Z-2544-2011 · Apr 22, 2011 · Terminated
    Comprehensive reverse shoulder humeral tray with locking ring, 44 mm +5 mm, sterile, single use, REF 115345, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulde
  • Z-2542-2011 · Apr 22, 2011 · Terminated
    Bio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder
  • Z-2541-2011 · Apr 22, 2011 · Terminated
    Bio-modular reverse shoulder humeral tray with lock ring, 44 mm 10 degree angle, sterile, single use, REF 11-113685, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse S
  • Z-2543-2011 · Apr 22, 2011 · Terminated
    Comprehensive reverse shoulder humeral tray with locking ring, 44 mm +10 mm, sterile, single use, REF 115348, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Should

…and 4 more.

Common questions about the Comprehensive® Reverse Shoulder 180550

Has the Comprehensive® Reverse Shoulder 180550 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00880304677135), based on the FDA recall file dated Nov 27, 2026. However, 19 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Comprehensive® Reverse Shoulder 180550?

The FDA's MAUDE database lists 43 adverse event reports for this device, received between Nov 26, 2019 and Feb 23, 2026. In those reports, the most common specific device problems are “Loosening of Implant Not Related to Bone-Ingrowth” (13), “Detachment of Device or Device Component” (5) and “Fracture” (5); the most common patient problems are “Pain” (13), “Failure of Implant” (11) and “Inadequate Osseointegration” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Comprehensive® Reverse Shoulder 180550?

2 reports are classified as deaths, 40 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Comprehensive® Reverse Shoulder 180550 FDA approved?

It was cleared through 510(k) K080642 on Jul 9, 2008; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Comprehensive® Reverse Shoulder 180550, and is it still sold?

The FDA's device database lists Biomet Orthopedics, LLC as the company responsible for it (brand Comprehensive® Reverse Shoulder, model 180550). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented devices

All prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page