Skip to content

The Tether™ 211H7037Recalls & reported problems

Zimmer Biomet Inc · 7.0MM DIA X 37.5MM VBT ASSY · Class N

All models of The Tether (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00880304865273), received between Jan 24, 2022 and Jan 24, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 24, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED A REVISION SURGERY WAS PERFORMED BECAUSE THE SURGEON BELIEVED THE TETHER CORD HAD BROKEN POST-OPERATIVELY. HOWEVER, DURING THE REVISION, THE CORD WAS FOUND TO BE INTACT, BUT SOME OF THE SET SCREWS DID NOT APPEAR TO BE LOCKED AND WERE LOOSE DURING REMOVAL, ESPECIALLY AT L3. THE CORD AND SET SCREWS WERE REMOVED AND REPLACED. THIS IS REPORT FIVE OF SEVEN FOR THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 13334447 ↗
  • InjuryJan 24, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED A REVISION SURGERY WAS PERFORMED BECAUSE THE SURGEON BELIEVED THE TETHER CORD HAD BROKEN POST-OPERATIVELY. HOWEVER, DURING THE REVISION, THE CORD WAS FOUND TO BE INTACT, BUT SOME OF THE SET SCREWS DID NOT APPEAR TO BE LOCKED AND WERE LOOSE DURING REMOVAL, ESPECIALLY AT L3. THE CORD AND SET SCREWS WERE REMOVED AND REPLACED. THIS IS REPORT FOUR OF SEVEN FOR THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 13334443 ↗

Other Vertebral Body Tethering System devices

All Vertebral Body Tethering System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page