LIFEPAK CR Plus and LIFEPAK Express defibrillators 80403Recalls & reported problems
Physio-Control, Inc. · PACKAGE,LPCR PLUS,AUTO,ENGLISH,AHA2005, FDA
All models of LIFEPAK CR Plus and LIFEPAK Express defibrillators (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00883873820513), received between Jun 4, 2019 and Jun 4, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Power Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Power Problem1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2019Power Problem; Power Problem
A CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8667943 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page