Skip to content

NA 11131 (REF 11131-000045)Recalls & reported problems

Physio-Control, Inc. · Sterilizable Internal Defibrillation Paddles, LP15 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00883873863497), received between Sep 3, 2024 and Jun 30, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver Shock/Stimulation” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 9, 2024Open, Classified

    Due to customer complaints and out of box failures, the firm is conducting a voluntary recall as a result of product failures (cracks and over-molding separation of the sterilizable internal defibrillation paddles.

    Z-0142-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Failure to Deliver Shock/Stimulation5
  • Fracture1
  • Inappropriate/Inadequate Shock/Stimulation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Cardiac Arrest1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2026Inappropriate/Inadequate Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S INTERNAL PADDLES FAILED THE CONTINUITY TEST. IN THIS STATE THE PADDLES MAY NOT BE ABLE TO DELIVER THE APPROPRIATE DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25665569 ↗
  • MalfunctionMay 27, 2026Failure to Deliver Shock/Stimulation; Fracture

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S INTERNAL PADDLES DID NOT DELIVER A SHOCK. A 2ND SET WAS USED TO DELIVER THE SHOCK. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25298632 ↗
  • MalfunctionOct 22, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE INTERNAL PADDLES OF THEIR DEVICE DID NOT WORK. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT HAVE BEEN AVAILABLE, IF NEEDED. THERE WERE NO REPORTS OF HARM TO THE PATIENT AS A RESULT OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356177 ↗
  • MalfunctionJan 2, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT DELAY IN THE DELIVERY OF THE SHOCK DURING INTERVENTION ON A PATIENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THE PATIENT ASSOCIATED WITH THE REPORTED EVENT DID NOT SURVIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21060343 ↗
  • MalfunctionSep 3, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE INTERNAL PADDLES HAVE FAILED. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20131295 ↗
  • MalfunctionSep 3, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE PADDLES DID NOT DELIVER ENERGY WHEN THE SHOCK BUTTON WAS PUSHED. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: Cardiac Arrest

    MAUDE report 20126502 ↗

Other Dc-defibrillator, low-energy, (including paddles) devices

All Dc-defibrillator, low-energy, (including paddles) devices →

Sources and updates