Embosphere® 00884450116715Recalls & reported problems
Biosphere Medical · Class 2 · Implant
All models of Embosphere (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00884450116715), received between Jun 16, 2021 and Jul 14, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Illegible Information” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Aug 10, 2026Open, Classified
Microspheres used in embolization of tumors, prostatic arteries, arteriovenous malformations, and blood vessels may have holes in packaging that may affect the sterile barrier, which if used could lead to infection.
Z-3305-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Illegible Information6
- Adverse Event Without Identified Device or Use Problem1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Thrombosis/Thrombus1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 14, 2026Fluid/Blood Leak
CUSTOMER REPORTED THAT THERE WAS NO LIQUID IN THE SYRINGE IN THE ORIGINAL PACKAGING. THE MICROSPHERE WAS FOUND TO BE IN A DRY STATE, AND THE LIQUID LEAK WAS OBSERVED. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25812269 ↗MalfunctionApr 14, 2026Illegible Information
THE CUSTOMER REPORTED THE EXPIRY DATE (YEAR) ON THE PRODUCT LABEL CONFUSED THE CUSTOMER BECAUSE IT LOOKS LIKE 2026 INSTEAD OF 2025, HENCE THE HOSPITAL REJECTED AND RETURNED THE UNUSED PRODUCT TO DISTRIBUTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24873400 ↗MalfunctionApr 14, 2026Illegible Information
THE CUSTOMER REPORTED THE EXPIRY DATE (YEAR) ON THE PRODUCT LABEL CONFUSED THE CUSTOMER BECAUSE IT LOOKS LIKE 2026 INSTEAD OF 2025, HENCE THE HOSPITAL REJECTED AND RETURNED THE UNUSED PRODUCT TO DISTRIBUTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24870851 ↗MalfunctionApr 14, 2026Illegible Information
THE CUSTOMER REPORTED THE EXPIRY DATE (YEAR) ON THE PRODUCT LABEL CONFUSED THE CUSTOMER BECAUSE IT LOOKS LIKE 2026 INSTEAD OF 2025, HENCE THE HOSPITAL REJECTED AND RETURNED THE UNUSED PRODUCT TO DISTRIBUTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24870843 ↗MalfunctionApr 14, 2026Illegible Information
THE CUSTOMER REPORTED THE EXPIRY DATE (YEAR) ON THE PRODUCT LABEL CONFUSED THE CUSTOMER BECAUSE IT LOOKS LIKE 2026 INSTEAD OF 2025, HENCE THE HOSPITAL REJECTED AND RETURNED THE UNUSED PRODUCT TO DISTRIBUTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24870836 ↗MalfunctionApr 14, 2026Illegible Information
THE CUSTOMER REPORTED THE EXPIRY DATE (YEAR) ON THE PRODUCT LABEL CONFUSED THE CUSTOMER BECAUSE IT LOOKS LIKE 2026 INSTEAD OF 2025, HENCE THE HOSPITAL REJECTED AND RETURNED THE UNUSED PRODUCT TO DISTRIBUTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24870835 ↗MalfunctionApr 14, 2026Illegible Information
THE CUSTOMER REPORTED THE EXPIRY DATE (YEAR) ON THE PRODUCT LABEL CONFUSED THE CUSTOMER BECAUSE IT LOOKS LIKE 2026 INSTEAD OF 2025, HENCE THE HOSPITAL REJECTED AND RETURNED THE UNUSED PRODUCT TO DISTRIBUTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24870830 ↗InjuryJun 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE ACCOUNT ALLEGES THAT A TUMOR EMBOLIZATION WAS PERFORMED USING EMBOLIC DEVICES WITHIN THE MIDDLE MENINGEAL ARTERY 2 DAYS BEFORE SCHEDULED CRANIOTOMY SURGERY. LATER ON A CRANIOTOMY INVOLVING THE CONTRALATERAL SIDE WAS PERFORMED, FOLLOWED BY RUMOR RESECTION (SIMPSON GRADE, II). DURAL DEFECTS WERE CLOSED WITH DURAGEN. PATHOLOGICAL DIAGNOSIS WAS TRANSITIONAL MENINGIOMA (WHO GRADE, I). NO NEW NEUROLOGICAL SYMPTOMS APPEARED, BUT SENSORY DISTURBANCE AND PARESIS IN THE LEFT UPPER AND LOWER EXTREMITIES DEVELOPED 8 DAYS AFTER SURGERY. MRI SHOWED CEREBRAL INFARCTION IN THE RIGHT PARIETAL LOBE. CEREBRAL VENOUS THROMBOSIS (CVT) WAS SUSPECTED, AND DRIP INFUSION OF HEPARIN WAS STARTED. TWELVE HOURS LATER, HOWEVER, SEIZURES OCCURRED ON THE LEFT SIDE OF THE BODY, AND MOVEMENT DISORDER (MANUAL MUSCLE TESTING, 2/5) AND SENSORY DISTURBANCE ON THE LEFT SIDE OF THE BODY, LEFT UNILATERAL SPATIAL NEGLECT AND APRAXIA DEVELOPED. MRI SHOWED THAT THE INFARCT AREA IN THE RIGHT TEMPORAL LOBE HAD BECOME HEMORRHAGIC, AND MAGNETIC RESONANCE VENOGRAPHY (MRV) REVEALED STENOSIS OF THE SUPERIOR SAGITTAL SINUS JUST BENEATH THE SITE OF CRANIOTOMY. SUPERIOR SAGITTAL SINUS THROMBOSIS WAS THUS DIAGNOSED. DRIP INFUSION O
Patient problems: Thrombosis/Thrombus
MAUDE report 12013323 ↗
Other AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS devices
- HydroPearl HP2S0600Microvention, Inc. · Agents, embolic, for treatment of uterine fibroids1 recall0 reports
- Embosphere® 00884450109519Biosphere Medical · AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS1 recall0 reports
- Embosphere® 00884450098561Biosphere Medical · AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS8 reports
- Embosphere® 00884450098424Biosphere Medical · AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS8 reports
- Embosphere® 00884450106051Biosphere Medical · AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS3 reports
- Embosphere® 00884450116296Biosphere Medical · AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS1 report
- Cook G09665Cook Incorporated · AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS1 report
- Bearing nsPVA® 00884450123317Merit Medical Systems, Inc. · AGENTS,EMBOLIC,FOR TREATMENT OF UTERINE FIBROIDS1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page