Skip to content

Merit Medical® 00884450184233Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884450184233), received between May 13, 2026 and May 13, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Product Quality Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Product Quality Problem1

Patient problems

  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2026Product Quality Problem

    CUSTOMER REPORTS A PPATIENT HAD RIGHT RADIAL AND RIGHT FEMORAL ARTERY ACCESSES FOR CHRONIC TOTAL OCCLUSION INTERVENTION. DURING PROCEDURE, RADIAL AND FEMORAL ARTERIAL TRANSDUCED BLOOD PRESSURE SIGNIFICANTLY DROPPED SIMULTANEOUSLY FROM 160S TO 60S. THERE WAS A DISCREPANCY BETWEEN NON-INVASIVE BLOOD PRESSURE AND TRANSDUCED PRESSURE. DUE TO LOW TRANSDUCED PRESSURE, PRESSORS AND INOTROPES WERE ADMINISTERED. PATIENT WAS INTUBATED. CHANGED OUT ACIST TRANSDUCER, SO BELIEVE PROBLEM WAS IN MALFUNCTIONING CONSUMABLE/DISPOSABLE MANIFOLD.

    Patient problems: Low Blood Pressure/ Hypotension

    MAUDE report 25163776 ↗

Other Cardiovascular procedure kit devices

All Cardiovascular procedure kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page