Skip to content

Flex-Neck® 00884450330210Recalls & reported problems

Merit Medical Systems, Inc. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884450330210), received between Mar 8, 2021 and Mar 17, 2021 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Fracture2
  • Detachment of Device or Device Component1
  • Material Separation1

Patient problems

  • Extravasation1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 17, 2021Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES THAT POST-PLACEMENT OF A PERITONEAL DIALYSIS CATHETER ON (B)(6) 2021. THE CUSTOMERS HOMECARE NURSE REPORTED ON 02-23-2021 THAT THE CATHETER HAD DETACHED OUTSIDE THE PATIENT BODY AND A BAD ODOR AND DRAINAGE WAS NOTED. THE PATIENT WAS TRANSFERRED TO THE HOSPITAL IMMEDIATELY FOR A PD CATHETER REPLACEMENT. ANTIBIOTICS WERE ADMINISTERED FOR THIS PATIENT. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11509943 ↗
  • InjuryMar 8, 2021Fracture; Detachment of Device or Device Component

    FEMALE PATIENT IN HER 50¿S, WITH A NEWLY INSERTED PERITONEAL DIALYSIS CATHETER. RETURNED FOR REMOVAL OF THE PERITONEAL DIALYSIS CATHETER 5 DAYS LATER, BECAUSE THE CATHETER WAS FRACTURED, LEAVING ONLY THE INTERNAL SEGMENT. THE PHYSICIAN'S VISUAL INSPECTION IDENTIFIED THE FRACTURE WAS AT THE LEVEL OF THE SUPERFICIAL CATHETER CUFF. DAY 5: UNPLANNED PROCEDURE FOR REMOVAL OF CATHETER. DAY 5: CIPRO X 10 DAYS, FOLLOW UP VISIT 2 WEEKS AND NEW CATHETER TO BE INSERTED AFTER SITE IS HEALED. DAY 15: PER DR. - HE FEELS THERE PROBABLY WAS A DEFECT IN THE CATHETER BECAUSE WHEN LOOKING BACK AT THE PICTURES IT LOOKS LIKE THERE WAS SOME EXTRAVASATION WHICH HE THOUGHT ORIGINALLY WAS JUST FLOWING BACK THE TRACK.

    Patient problems: Extravasation

    MAUDE report 11434936 ↗

Other Catheter, peritoneal, long-term indwelling devices

All Catheter, peritoneal, long-term indwelling devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page