Skip to content

Merit Maestro® 00884450357903Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

All models of Merit Maestro (27)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 6, 2023Class IIOpen, Classified

    The sterility of microcatheter and infusion system devices cannot be guaranteed.

    Z-0627-2024 · Root cause: Package design/selection · Recall details · FDA record ↗

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page