Skip to content

Merit Maestro® 00884450358160Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

All models of Merit Maestro (27)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884450358160), received between Dec 8, 2021 and Dec 8, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 6, 2023Class IIOpen, Classified

    The sterility of microcatheter and infusion system devices cannot be guaranteed.

    Z-0627-2024 · Root cause: Package design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 8, 2021Fracture; Fracture

    THE ACCOUNT ALLEGES DURING A FIBROIDS EMBOLIZATION PROCEDURE, THE RADIPAQUE TIP OF THE MICROCATHETER CAME LOOSE OUTSIDE THE BODY OF THE PATIENT. THE DEVICE WAS REPLACED WITH A NEW UNIT. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12960602 ↗

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page