Skip to content

basixCOMPAK™ 00884450498538Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884450498538), received between Dec 12, 2025 and Dec 12, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Inflation Problem1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2025Inflation Problem; Separation Failure; Detachment of Device or Device Component; Inflation Problem; Separation Failure; Detachment of Device or Device Component

    THE ACCOUNT ALLEGES THAT DURING A SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE, VASCULAR DILATION WAS PERFORMED USING A KANEKA PTA BALLOON. THE BALLOON WAS PRESSURIZED TO 30 ATM ON TWO OCCASIONS WITHOUT ISSUE. DURING THE THIRD PRESSURIZATION, A POPPING SOUND WAS REPORTED, AND THE PRESSURE GAUGE READING DROPPED TO 0 ATM. THE PLUNGER REMAINED ENGAGED AND DID NOT RETRACT. THE OPERATOR THEN DEPRESSURIZED THE DEVICE BY GRASPING THE LOCK/RELEASE BAR AND PULLING THE PLUNGER HANDLE. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON OF THE SAME PART NUMBER AND LOT. THE PROCEDURE WAS COMPLETED WITHOUT REPORTED PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23791676 ↗

Other Syringe, Balloon Inflation, Exempt devices

All Syringe, Balloon Inflation, Exempt devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.