Skip to content

basixCOMPAK™ 00884450638347Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

All models of basixCOMPAK (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884450638347), received between Sep 3, 2025 and Jan 28, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 4, 2025Open, Classified

    Inflation device handle may detach from the syringe during procedure.

    Z-1123-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Material Separation3
  • Failure to Deflate1
  • Inflation Problem1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2026Inflation Problem; Detachment of Device or Device Component; Separation Problem

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO DEFLATE WITH THE INFLATION DEVICE, THE HANDLE SEPARATED FROM THE REST OF THE DEVICE. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24199299 ↗
  • MalfunctionSep 3, 2025Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    ADDITIONAL INFORMATION WAS RECEIVED ABOUT THE REPORTED EVENT. IT WAS REPORTED THE HANDLE OF THE INFLATION DEVICE DETACHED. THE NURSE USING THE DEVICE WAS MANIPULATING THE DEVICE WHEN THE HANDLE FELL OFF. THERE WAS NO REPORT OF THE INABILITY FOR A BALLOON TO DEFLATE FOR THIS REPORTED EVENT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22952884 ↗
  • MalfunctionSep 3, 2025Material Separation; Detachment of Device or Device Component; Failure to Deflate; Material Separation; Detachment of Device or Device Component; Failure to Deflate

    ADDITIONAL INFORMATION WAS RECEIVED IN A MEETING ON 15OCT2025. THE FRANCE TEAM CONFIRMED THEY HAD AN INABILITY TO DEFLATE BALLOON OCCUR ON ONE UNIT DURING A CORONARY ANGIOPLASTY OF THE LEFT CORONARY TRUNK. THE HANDLE STAYED ATTACHED BUT WOULD NOT DEFLATE. THEY WERE ABLE TO USE ANOTHER INFLATOR, AND THE PATIENT SUFFERED NO HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22952880 ↗
  • MalfunctionSep 3, 2025Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    OUR ANGIOGRAPHY DEPARTMENT HAS ENCOUNTERED A PROBLEM WITH THE MERIT INFLATOR REFERENCE IN4352 I3129671 (ISSUE ON 07/01/2025). THE INFLATOR HANDLE ENDED UP IN THE NURSE'S HAND. NO PATIENT DETAILS OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22952877 ↗

Other Syringe, Balloon Inflation, Exempt devices

All Syringe, Balloon Inflation, Exempt devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page