1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884450771310), received between Feb 20, 2025 and Feb 20, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 2, 2026Class IIOpen, Classified

    catheter may experience resistance when being advanced over the guidewire

    Z-2002-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 20, 2025Fracture; Material Separation; Fracture; Material Separation

    THE ACCOUNT ALLEGES THAT A BRIDGE GUIDE WIRE FROM ONE OF THE PREP KITS BROKE APART INSIDE THE PATIENT. THE TIP OF THE WIRE GOT STUCK IN THE IVC AND WHEN THE EP EXTRACTOR PULLED BACK THE GUIDE WIRE IT SLINKIED APART. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY SNARE THE PIECE THAT BROKE OFF.

    Patient problems: Foreign Body In Patient

    MAUDE report 21420479 ↗

FDA clearance and classification

The FDA regulates product code OEQ (angiography/angioplasty kit) under 21 CFR 870.1650 as a Class II device. 21 CFR 870.1650 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Bridge™ Prep Kit 00884450771310

Has the Bridge™ Prep Kit 00884450771310 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2002-2026 (Class II), began on Apr 2, 2026: catheter may experience resistance when being advanced over the guidewire. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Bridge™ Prep Kit 00884450771310?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 20, 2025. That report lists device problems “Fracture” (1) and “Material Separation” (1); patient problem “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Bridge™ Prep Kit 00884450771310?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Bridge™ Prep Kit 00884450771310, and is it still sold?

The FDA's device database lists Merit Medical Systems, Inc. as the company responsible for it (brand Bridge™ Prep Kit, model 00884450771310). Its record shows it as in commercial distribution.