DuraMax® (REF H787103028185)Recalls & reported problems
Merit Medical Systems, Inc. · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00884450786178), received between Jul 2, 2025 and Mar 30, 2026 — classified as 8 injuries, 1 malfunction.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Feb 13, 2026Open, Classified
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.
Z-1578-2026 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Patient Device Interaction Problem7
- Migration4
- Migration or Expulsion of Device4
- Break1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Unspecified Infection3
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2026Break
CUSTOMER REPORTS THE SHEATH BROKE AT THE WRONG LOCATION; PHYSICIAN HAD TO RETRIEVE IT FROM WITHIN THE PATIENT. NO ADDITIONAL DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24735053 ↗MalfunctionMar 23, 2026Separation Failure
THE CUSTOMER REPORTS THAT THE SHEATH DID NOT PEEL AND BROKE. THERE WAS NO PATIENT INJURY AND THE REST OF CATHETER WAS SUCCESSFULLY PLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24670106 ↗InjuryJul 2, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IN A RANDOMIZED CLINICAL STUDY EVALUATING THE PERFORMANCE OF THE DURAMAX® TUNNELED HEMODIALYSIS CATHETER IDENTIFIED A CASE OF CATHETER-RELATED INFECTION. THESE INFECTIONS NECESSITATED EARLY CATHETER REMOVAL. THE INCIDENCE RATE OF CRI IN THE DM GROUP WAS 1.14 EVENTS PER 1,000 CATHETER DAYS, WHICH WAS NOTABLY HIGHER THAN THE COMPARATOR GROUP'S RATE OF 0.67 EVENTS PER 1,000 CATHETER DAYS. ALL CATHETERS WERE IMPLANTED WITHOUT PROCEDURAL COMPLICATIONS. THIS REPORT IS BEING SUBMITTED AS A SERIOUS INJURY DUE TO THE NEED FOR MEDICAL INTERVENTION AND EARLY DEVICE REMOVAL.
Patient problems: Unspecified Infection
MAUDE report 22384400 ↗InjuryJul 2, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IN A RANDOMIZED CLINICAL STUDY EVALUATING THE PERFORMANCE OF THE DURAMAX® TUNNELED HEMODIALYSIS CATHETER IDENTIFIED A CASE OF CATHETER-RELATED INFECTION. THESE INFECTIONS NECESSITATED EARLY CATHETER REMOVAL. THE INCIDENCE RATE OF CRI IN THE DM GROUP WAS 1.14 EVENTS PER 1,000 CATHETER DAYS, WHICH WAS NOTABLY HIGHER THAN THE COMPARATOR GROUP'S RATE OF 0.67 EVENTS PER 1,000 CATHETER DAYS. ALL CATHETERS WERE IMPLANTED WITHOUT PROCEDURAL COMPLICATIONS. THIS REPORT IS BEING SUBMITTED AS A SERIOUS INJURY DUE TO THE NEED FOR MEDICAL INTERVENTION AND EARLY DEVICE REMOVAL.
Patient problems: Unspecified Infection
MAUDE report 22384398 ↗InjuryJul 2, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IN A RANDOMIZED CLINICAL STUDY EVALUATING THE PERFORMANCE OF THE DURAMAX® TUNNELED HEMODIALYSIS CATHETER IDENTIFIED A CASE OF CATHETER-RELATED INFECTION. THESE INFECTIONS NECESSITATED EARLY CATHETER REMOVAL. THE INCIDENCE RATE OF CRI IN THE DM GROUP WAS 1.14 EVENTS PER 1,000 CATHETER DAYS, WHICH WAS NOTABLY HIGHER THAN THE COMPARATOR GROUP'S RATE OF 0.67 EVENTS PER 1,000 CATHETER DAYS. ALL CATHETERS WERE IMPLANTED WITHOUT PROCEDURAL COMPLICATIONS. THIS REPORT IS BEING SUBMITTED AS A SERIOUS INJURY DUE TO THE NEED FOR MEDICAL INTERVENTION AND EARLY DEVICE REMOVAL.
Patient problems: Unspecified Infection
MAUDE report 22384396 ↗InjuryJul 2, 2025Migration or Expulsion of Device; Patient Device Interaction Problem; Migration; Migration or Expulsion of Device; Patient Device Interaction Problem; Migration
IN A RANDOMIZED CLINICAL STUDY EVALUATING THE PERFORMANCE OF THE DURAMAX® TUNNELED HEMODIALYSIS CATHETER IDENTIFIED A CASE OF CATHETER DISLODGEMENT. ALL DEVICES WERE PLACED WITHOUT TECHNICAL DIFFICULTY OR PROCEDURAL COMPLICATIONS. HOWEVER, THE DM CATHETER GROUP EXHIBITED A DISPROPORTIONATELY HIGHER RATE OF DISLODGEMENT COMPARED TO THE COMPARATOR DEVICE GROUP, WHICH HAD ONLY ONE DISLODGEMENT, AND THAT PATIENT HAD A PRIOR DISLODGEMENT HISTORY WITH A DM CATHETER. THESE DISLODGEMENT EVENTS REQUIRED EARLY CATHETER REMOVAL. THIS REPORT IS BEING SUBMITTED AS A POTENTIAL DEVICE MALFUNCTION THAT MAY LEAD TO SERIOUS INJURY IF THE ISSUE WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22384393 ↗InjuryJul 2, 2025Migration or Expulsion of Device; Patient Device Interaction Problem; Migration; Migration or Expulsion of Device; Patient Device Interaction Problem; Migration
IN A RANDOMIZED CLINICAL STUDY EVALUATING THE PERFORMANCE OF THE DURAMAX® TUNNELED HEMODIALYSIS CATHETER IDENTIFIED A CASE OF CATHETER DISLODGEMENT. ALL DEVICES WERE PLACED WITHOUT TECHNICAL DIFFICULTY OR PROCEDURAL COMPLICATIONS. HOWEVER, THE DM CATHETER GROUP EXHIBITED A DISPROPORTIONATELY HIGHER RATE OF DISLODGEMENT COMPARED TO THE COMPARATOR DEVICE GROUP, WHICH HAD ONLY ONE DISLODGEMENT, AND THAT PATIENT HAD A PRIOR DISLODGEMENT HISTORY WITH A DM CATHETER. THESE DISLODGEMENT EVENTS REQUIRED EARLY CATHETER REMOVAL. THIS REPORT IS BEING SUBMITTED AS A POTENTIAL DEVICE MALFUNCTION THAT MAY LEAD TO SERIOUS INJURY IF THE ISSUE WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22384392 ↗InjuryJul 2, 2025Migration or Expulsion of Device; Patient Device Interaction Problem; Migration; Migration or Expulsion of Device; Patient Device Interaction Problem; Migration
IN A RANDOMIZED CLINICAL STUDY EVALUATING THE PERFORMANCE OF THE DURAMAX® TUNNELED HEMODIALYSIS CATHETER IDENTIFIED A CASE OF CATHETER DISLODGEMENT. ALL DEVICES WERE PLACED WITHOUT TECHNICAL DIFFICULTY OR PROCEDURAL COMPLICATIONS. HOWEVER, THE DM CATHETER GROUP EXHIBITED A DISPROPORTIONATELY HIGHER RATE OF DISLODGEMENT COMPARED TO THE COMPARATOR DEVICE GROUP, WHICH HAD ONLY ONE DISLODGEMENT, AND THAT PATIENT HAD A PRIOR DISLODGEMENT HISTORY WITH A DM CATHETER. THESE DISLODGEMENT EVENTS REQUIRED EARLY CATHETER REMOVAL. THIS REPORT IS BEING SUBMITTED AS A POTENTIAL DEVICE MALFUNCTION THAT MAY LEAD TO SERIOUS INJURY IF THE ISSUE WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22384384 ↗InjuryJul 2, 2025Migration or Expulsion of Device; Patient Device Interaction Problem; Migration; Migration or Expulsion of Device; Patient Device Interaction Problem; Migration
IN A RANDOMIZED CLINICAL STUDY EVALUATING THE PERFORMANCE OF THE DURAMAX® TUNNELED HEMODIALYSIS CATHETER IDENTIFIED A CASE OF CATHETER DISLODGEMENT. ALL DEVICES WERE PLACED WITHOUT TECHNICAL DIFFICULTY OR PROCEDURAL COMPLICATIONS. HOWEVER, THE DM CATHETER GROUP EXHIBITED A DISPROPORTIONATELY HIGHER RATE OF DISLODGEMENT COMPARED TO THE COMPARATOR DEVICE GROUP, WHICH HAD ONLY ONE DISLODGEMENT, AND THAT PATIENT HAD A PRIOR DISLODGEMENT HISTORY WITH A DM CATHETER. THESE DISLODGEMENT EVENTS REQUIRED EARLY CATHETER REMOVAL. THIS REPORT IS BEING SUBMITTED AS A POTENTIAL DEVICE MALFUNCTION THAT MAY LEAD TO SERIOUS INJURY IF THE ISSUE WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22384381 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.