Skip to content

BioSentry® ARecalls & reported problems

Merit Medical Systems, Inc. · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884450808122), received between Mar 30, 2026 and Mar 30, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Pneumothorax1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Activation, Positioning or Separation Problem

    CUSTOMER REPORTS THE OLD STYLE PLUG NOT ALLOWING THE PLUG TO GO INTO THE INTRODUCER PROPERLY. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24735017 ↗
  • MalfunctionMar 30, 2026Activation, Positioning or Separation Problem

    CUSTOMER REPORTS THE GEL PLUG DID NOT DEPLOY. THE PRODUCT WAS THROWN AWAY AND A NEW ONE WAS OPENED. NO PATIENT HARM. NO ADDITIONAL DETAILS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24718011 ↗
  • InjuryMar 30, 2026Activation, Positioning or Separation Problem

    THE CUSTOMER REPORTS THE PHYSICIAN HAD FINISHED TAKING ALL BIOPSY SAMPLES AND THEY WERE READY TO DEPLOY THE BIOSENTRY PLUG. THEY NOTICED THE PATIENT ALREADY HAD A PNEUMO SO THEY WERE READY TO DEPLOY THE PLUG AND FINISH THE CASE. THEY OPENED UP ONE BIOSENTRY PLUG AND NOTICED THE PUSHER WOULD NOT ADVANCE THROUGH THE ADAPTOR SO IT COULD NOT DEPLOY. THEY OPENED UP ANOTHER BIOSENTRY PLUG OF THE SAME LOT, BUT NOTICED THIS ONE WAS NOT ABLE TO ADVANCE ALL THE WAY THROUGH THE ADAPTOR EITHER. AT THIS POINT THEY OPENED UP AN OLDER LOT THEY HAD ON HAND TO COMPLETE THE PROCEDURE. THE PATIENT ALSO REQUIRED A BLOOD PATCH. NO ADDITIONAL DETAILS PROVIDED.

    Patient problems: Pneumothorax

    MAUDE report 24717433 ↗

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.