SureSigns VM4 SureSigns VM 4 Patient MonitorRecalls & reported problems
Philips Medical Systems Hsg · SureSigns VM 4 Patient Monitor · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00884838000032), received between Jan 15, 2019 and Sep 27, 2023 — classified as 18 malfunctions.
The most frequently reported specific device problem is “No Audible Prompt/Feedback” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 13, 2018Class IIOpen, Classified
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
Z-2895-2018 · Root cause: Labeling design · Recall details · FDA record ↗
Most reported problems
Device problems
- No Audible Prompt/Feedback4
- Application Program Freezes or Fails to Launch2
- Fire2
- Appropriate Device Problem Term/Code Not Available1
- Audible Prompt/Feedback Problem1
- Defective Alarm1
- Inaudible or Unclear Audible Prompt/Feedback1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- Insufficient Information2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 27, 2023Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
THE CUSTOMER REPORTED THAT THE SCREEN TURNS ON AND THEN FREEZES. IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17830453 ↗MalfunctionAug 22, 2023Defective Alarm; Defective Alarm
THE CUSTOMER REPORTED THAT THERE WAS A SPEAKER MALFUNCTION. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17597804 ↗MalfunctionMay 17, 2023No Audible Prompt/Feedback; No Audible Prompt/Feedback
ADDITIONAL INFORMATION WAS RECEIVED THAT THE DEVICE WAS NOT IN CLINICAL USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16950101 ↗MalfunctionDec 21, 2022Fire; Fire
THE CUSTOMER REPORTED THAT THEIR VM4 SURGEONS MONITOR CAUGHT FIRE DUE TO A BATTERY ISSUE. THERE WAS NO PATIENT HARM REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16030141 ↗MalfunctionDec 20, 2022Fire; Fire
THE CUSTOMER REPORTED THAT THEIR SURESIGNS VM MONITOR HAD A BATTERY ISSUE WHICH LED TO A FIRE. THE DEVICE WAS IN USE HOWEVER NO PATIENT OR USER HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16013931 ↗MalfunctionNov 8, 2022No Audible Prompt/Feedback; No Audible Prompt/Feedback
IT WAS REPORTED THERE WAS AN AUDIO SPEAKER ISSUE IN THE MONITOR. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE REPORTED EVENT. THERE WAS NO PATIENT INVOLVEMENT OR HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15753330 ↗MalfunctionSep 9, 2022Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED TO PHILIPS THAT VM4 WAS MALFUNCTIONING WHILE THE ON OFF KEY WAS PRESSED. THE COMPLAINT DEVICE WAS IN USE AT THE TIME THAT THE ISSUE WAS DISCOVERED. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15390820 ↗MalfunctionAug 30, 2022No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT THE DEVICE DOES NOT SOUND AN ALARM. PATIENT INVOLVEMENT IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15319666 ↗MalfunctionJul 21, 2022Audible Prompt/Feedback Problem; Audible Prompt/Feedback Problem
IT WAS REPORTED TO PHILIPS THAT SURESIGNS VM4 PATIENT MONITOR HAD SPEAKER MALFUNCTION. THE DEVICE WAS NOT BEING USED ON A PATIENT AT THE TIME OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15077745 ↗MalfunctionApr 6, 2022Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
THE CUSTOMER CALLED THE CUSTOMER CARE SOLUTION CENTER AND ALLEGED THAT THE MONITOR CRASHED. THE DEVICE WAS NOT IN USE AT THE TIME THAT THIS ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14024835 ↗MalfunctionFeb 10, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED TO PHILIPS THAT SPO2 SIGNAL DETECTED AFTER PATIENT EXPIRING. THE DEVICE WAS IN USE AT TIME O EVENT, THE PATIENT EXPIRED.
Patient problems: Insufficient Information
MAUDE report 13509598 ↗MalfunctionJan 6, 2022Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED TO PHILIPS THAT THE CUSTOMER COMPLAINT OF TEMPERATURE VARIATION, SPORADIC DELAY OF READING. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13176090 ↗MalfunctionNov 10, 2021No Audible Prompt/Feedback; No Audible Prompt/Feedback
IT WAS REPORTED TO PHILIPS THAT LOUDSPEAKER SYSTEM DEFECTIVE. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF EVENT, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12786922 ↗MalfunctionMar 18, 2021No Device Output; No Device Output
THE CUSTOMER CONTACTED THE CUSTOMER CARE SOLUTION CENTER AND ALLEGED THAT THE COMPLAINT DEVICE NO LONGER TURNED ON AND THAT WHEN IT WAS CONNECTED THE MAIN INDICATOR TURNED ON FOR A FEW SECONDS WITH A COLOR PATTERN, BUT NOTHING ON THE SCREEN. THE CUSTOMER ALSO INDICATED THAT THE LOUDSPEAKER ON THE DEVICE NO LONGER WORKED DUE TO A DAMAGED CONNECTION AND REQUESTED SUPPORT. THE COMPLAINT DEVICE WAS NOT IN CLINICAL USE AT THE TIME THAT THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11515939 ↗MalfunctionOct 28, 2020Inaudible or Unclear Audible Prompt/Feedback
THE CUSTOMER CONTACTED THE CUSTOMER CARE SOLUTION CENTER AND ALLEGED A SPEAKER MALFUNCTION ON THE COMPLAINT DEVICE AND DETERMINED THAT THERE WAS NO SOUND COMING FROM THE DEVICE AND REQUESTED SUPPORT. THE COMPLAINT DEVICE WAS NOT IN CLINICAL USE AT THE TIME THAT THE ISSUE WAS DISCOVERED.
Patient problems: Insufficient Information
MAUDE report 10747426 ↗MalfunctionOct 27, 2020No Audible Prompt/Feedback
THE CUSTOMER CONTACTED THE CUSTOMER CARE SOLUTION CENTER AND ALLEGED THAT THE COMPLAINT DEVICE HAD NO SOUND AND REQUESTED SUPPORT. THE COMPLAINT DEVICE WAS NOT IN CLINICAL USE AT THE TIME THAT THE ISSUE WAS DISCOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10740264 ↗MalfunctionMar 17, 2020Therapeutic or Diagnostic Output Failure
THE CUSTOMER CONTACTED THE CUSTOMER CARE SOLUTION CENTER AND ALLEGED THAT THE COMPLAINT DEVICE FAILED THE AUDIO TEST AND WAS VERY QUIET. THE CUSTOMER WAS UNABLE TO INCREASE THE VOLUME OF THE COMPLAINT DEVICE. THE COMPLAINT DEVICE WAS NOT IN CLINICAL USE AT THE TIME THAT THE ISSUE WAS DISCOVERED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9842567 ↗MalfunctionJan 15, 2019Appropriate Device Problem Term/Code Not Available
THE CUSTOMER REPORTED SPEAKER NOT WORKING. THE DEVICE WAS NOT CONNECTED TO A PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8248090 ↗
Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page