SureSigns VM6 SureSigns VM 6 Patient Monitor (REF 863065)Recalls & reported problems
Philips Medical Systems Hsg · SureSigns VM 6 Patient Monitor · Class 2
All models of SureSigns VM6 (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00884838000056), received between Jun 2, 2021 and Jul 17, 2024 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 13, 2018Class IIOpen, Classified
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
Z-2895-2018 · Root cause: Labeling design · Recall details · FDA record ↗
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch4
- No Audible Alarm3
- No Audible Prompt/Feedback2
- Display or Visual Feedback Problem1
- Inappropriate Audible Prompt/Feedback1
- Incorrect Measurement1
- Protective Measures Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2024No Audible Alarm; No Audible Alarm
PHILIPS RECEIVED A COMPLAINT ON THE SURESIGNS VM 6 PATIENT MONITOR INDICATING THAT THERE WAS AN AUDIO ERROR "SYSTEM ERROR 1036", WHICH INDICATED THAT THE SPEAKER WAS NOT PRODUCING SOUND. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19760004 ↗MalfunctionMay 23, 2023No Audible Prompt/Feedback; No Audible Prompt/Feedback
IT WAS REPORTED THERE IS A SPEAKER FAILURE, IT IS POSSIBLE THE LOUDSPEAKER IS DAMAGED. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16984000 ↗MalfunctionFeb 10, 2023No Audible Alarm; No Audible Alarm
PHILIPS RECEIVED A COMPLAINT ON THE SURESIGNS VM 6 PATIENT MONITOR SN (B)(6) INDICATING AN AUDIO ERROR "DEFECTIVE SPEAKER". THE DEVICE WAS NOT IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED. A PHILIPS FIELD SERVICE ENGINEER (FSE) WENT ONSITE TO EVALUATE THE DEVICES IN QUESTION.THE BIOMEDICAL ENGINEER (BMED) ONSITE CONFIRMED AN AUDIO ERROR. THE (FSE) CONFIRMED THE CUSTOMERS ALLEGATION AND FOUND THE SPEAKER TO BE COMPLETELY OUT OF SOUND. THE (FSE) REPLACED THE MAINBOARD AND SOUND SPEAKER FOR THEIR MALFUNCTION. BASED ON THE INFORMATION AVAILABLE AND THE TESTING CONDUCTED, THE CAUSE OF THE REPORTED PROBLEM WAS CONFIRMED.BASED ON THE INFORMATION AVAILABLE AND RESULTS OF ADDITIONAL ANALYSIS, NO FURTHER ACTION IS NECESSARY AT THIS TIME. AFTER SPEAKER ASSEMBLY AND MAINBOARD REPLACEMENT THE DEVICE WAS RETURNED TO FUNCTIONAL USE WITH NO FURTHER ISSUES IDENTIFIED. THE DEVICE REMAINS AT THE CUSTOMER SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16352166 ↗MalfunctionJan 20, 2023No Audible Alarm; No Audible Alarm
THE CUSTOMER REPORTED AN AUDIO ERROR INDICATING A DEFECTIVE SPEAKER. THE DEVICE WAS IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED. A PHILIPS FIELD SERVICE ENGINEER (FSE) WENT ONSITE TO EVALUATE THE DEVICES IN QUESTION. THE FOLLOWING TECHNICAL INVESTIGATION WAS PERFORMED BY INTERVIEWING THE BIOMEDICAL ENGINEER ONSITE (BME): THE (BME) IDENTIFIED AN ISSUE WITH THE MAINBOARD. AN ONSITE VISIT WAS SCHEDULED TO REPAIR THE UNIT. THE (FSE) CONFIRMED THE UNIT WAS COMPLETELY OUT OF SOUND AND THERE WAS A SPEAKER MALFUNCTION INOP. BASED ON THE INFORMATION AVAILABLE AND THE TESTING CONDUCTED, THE CAUSE OF THE REPORTED PROBLEM WAS CONFIRMED. THE REPORTED PROBLEM WAS CONFIRMED. THE MAINBOARD WAS REPLACED TO RESOLVE THE ISSUE AND THE DEVICE WAS RETURNED TO FUNCTIONAL USE WITH NO FURTHER ISSUES IDENTIFIED. THE DEVICE REMAINS AT THE CUSTOMER SITE. NO FURTHER INVESTIGATION OR ACTION IS WARRANTED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16218076 ↗MalfunctionJul 19, 2022Inappropriate Audible Prompt/Feedback; Inappropriate Audible Prompt/Feedback
IT WAS REPORTED TO PHILIPS THAT SURESIGNS VM6 PATIENT MONITOR SHOWED AUDIO ABNORMAL MESSAGE.THE DEVICE WAS NOT IN USE AT THE TIME THE ISSUE WAS DISCOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15057152 ↗MalfunctionJul 6, 2022Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED TO PHILIPS THAT SURESIGNS VM6 PATIENT MONITOR LOCKED PRESENTING SYSTEM ERROR 1036; IT WAS ONLY POSSIBLE TO TURN OFF EQUIPMENT BY REMOVING THE BATTERY. THE DEVICE WAS NOT IN USE AT THE TIME THE ISSUE WAS DISCOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14928471 ↗MalfunctionJul 1, 2022Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED TO PHILIPS THAT SURESIGNS VM 6 PATIENT MONITOR HAD SPO2 ISSUE. THE DEVICE WAS NOT IN USE AT THE TIME THE ISSUE WAS DISCOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14890189 ↗MalfunctionJun 10, 2022Display or Visual Feedback Problem; Application Program Freezes or Fails to Launch; Display or Visual Feedback Problem; Application Program Freezes or Fails to Launch
IT WAS REPORTED TO PHILIPS THAT VM6 US (B)(4) SCREEN FROZE. PATIENT INVOLVEMENT IS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 14655021 ↗MalfunctionApr 29, 2022No Audible Prompt/Feedback; No Audible Prompt/Feedback
IT WAS REPORTED TO PHILIPS THAT VM6 NEEDED REPAIR. AFTER GOOD FAITH EFFORT RESPONSE WAS RECEIVED ON 20APR2022, PHILIPS BECAME AWARE THAT VM6 HAD AUDIO DEFECT. ACCORDING TO THE THIRD GOOD FAITH EFFORT RESPONSE THE DEVICE WAS DURING A TEST OF DEVICE PERFORMED BY NURSE AT THE TIME THE ISSUE WAS DISCOVERED. NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14250176 ↗MalfunctionNov 22, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED TO GERMAY COMPETENT AUTHORITY (BFARM USER REPORT (B)(4)) THAT WHEN ACTIVATING THE RESPIRATION DISPLAY, THE SCREEN "FREEZES" PERMANENTLY. PATIENT MONITORING WAS NO LONGER POSSIBLE.NO PARAMETERS COULD BE CHANGED, ALARMS WERE NOT DISPLAYED OR REPORTED.THE MONITOR COULD ONLY BE OPERATED AGAIN AFTER A HARDWARE RESET.AT THE TIME OF THE EVENT, ECG AND SPO2 SENSORS WERE CONNECTED.THE DEVICE WAS TAKEN OUT OF SERVICE. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12852023 ↗MalfunctionNov 22, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED TO GERMANY COMPETENT AUTHORITY (BFARM USER REPORT (B)(4)) THAT WHEN ACTIVATING THE RESPIRATION DISPLAY, THE SCREEN "FREEZES" PERMANENTLY. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12851837 ↗MalfunctionJun 2, 2021Protective Measures Problem; Protective Measures Problem
THE CUSTOMER CONTACTED THE CUSTOMER CARE SOLUTION CENTER AND ALLEGED THAT THE MONITOR ON THE COMPLAINT DEVICE FREEZES WHEN ADJUSTING PARAMETERS AND THAT THE SCREEN HAD FROZEN. THE CUSTOMER THEN REQUESTED SUPPORT. IT WAS ORIGINALLY UNKNOWN IF THE COMPLAINT DEVICE WAS IN CLINICAL USE AT THE TIME THAT THE ISSUE WAS DISCOVERED AS THIS INFORMATION WAS NOT PROVIDED. INFORMATION WAS LATER PROVIDED WHICH INDICATED THAT THE COMPLAINT DEVICE WAS IN USE AT THE TIME THAT THIS ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11921333 ↗
Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page