Skip to content

IntelliVue Multi Measurement Server X2 865039Recalls & reported problems

Philips Medizin Systeme Böblingen GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Philips IntelliVue patient monitor tracks vital signs and sounds alarms, and its 504 reports mostly describe alarms not sounding. Other reports describe blood pressure and heart rate readings that seemed wrong. There are 7 deaths and 14 injuries reported. 1 open recall names this device, which began in October 2025 because the monitors might not sound an alarm.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
506
Deaths reported
7
Injuries reported
14
Recalls naming it
1

Summary

The FDA's MAUDE database contains 506 adverse event reports that list this exact device (DI 00884838000261), received between Feb 23, 2018 and Sep 29, 2026 — classified as 7 deaths, 14 injuries, 485 malfunctions.

The most frequently reported specific device problem is “No Audible Alarm” (215 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 10 reportsDec 2023: 7 reportsJan 2024: 4 reportsFeb 2024: 14 reportsMar 2024: 14 reportsApr 2024: 9 reportsMay 2024: 5 reportsJun 2024: 12 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 5 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 4 reportsDec 2025: 5 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 31, 2025Class IIOpen, Classified

    Potential issue where the IntelliVue monitors did not alarm.

    Z-0859-2026 · Root cause: Use error · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Audible Alarm215
  • No Audible Prompt/Feedback202
  • Defective Alarm16
  • Device Alarm System11
  • Incorrect Measurement10
  • Delayed Alarm7
  • Audible Prompt/Feedback Problem6
  • Display or Visual Feedback Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions450
  • No Known Impact Or Consequence To Patient28
  • Insufficient Information11
  • No Patient Involvement4
  • Cardiac Arrest3
  • Hypoxia2
  • Low Oxygen Saturation2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 25 most recent of 506 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931438 ↗
  • MalfunctionSep 15, 2026Incorrect Measurement

    Philips received a complaint on an IntelliVue Multi Measurement Server X2 indicating that SpO2 measurements could not be obtained and there is a suspicion that the SpO2 measurement was inaccurate. During remote troubleshooting replacement of the sensor cable did not resolve the issue. The device was reported to be in clinical use but not in use on a patient. No adverse event to the patient or user was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50777524 ↗
  • MalfunctionSep 3, 2026No Audible Prompt/Feedback

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50600077 ↗
  • MalfunctionAug 12, 2026No Audible Prompt/Feedback

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50226107 ↗
  • MalfunctionJun 30, 2026Imprecise or Erratic Results

    PHILIPS RECEIVED A COMPLAINT ON AN INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THEY WERE GETTING AN INCORRECT ARTERIAL READING. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE OR HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25670232 ↗
  • MalfunctionApr 17, 2026Low Readings

    PHILIPS RECEIVED A COMPLAINT ON THE INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THE HEART RATE WAS SHOWING SLOW AND ABNORMAL. THE DEVICE WAS IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24915913 ↗
  • MalfunctionApr 14, 2026Incorrect Measurement

    PHILIPS RECEIVED A COMPLAINT ON THE INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THE ARTERIAL BLOOD PRESSURE PERFORMANCE TESTS SHOWED VALUES OUTSIDE THE SPECIFIED RANGE. THE DEVICE WAS IN CLINICAL USE AT TIME OF EVENT, NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24879856 ↗
  • MalfunctionApr 1, 2026Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT NON-INVASIVE BLOOD PRESSURE (NIBP) VALUES WERE OUTSIDE THE SPECIFIED VALUE RANGE. IT IS UNKNOWN IF THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24754716 ↗
  • MalfunctionMar 10, 2026Device Fell

    IT WAS REPORTED THAT THE GCX ARM BROKE OFF AND THE X2 FELL OFF AND HIT THE FLOOR. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24560487 ↗
  • MalfunctionFeb 16, 2026No Audible Prompt/Feedback

    PHILIPS RECEIVED A COMPLAINT ON THE INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THEY REQUIRE SPARE SPEAKER. IT IS UNKNOWN IF THE DEVICE COULD PRODUCE AN AUDIBLE SOUND. THE DEVICE WAS NOT IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24349369 ↗
  • MalfunctionJan 14, 2026No Audible Alarm

    IT WAS REPORTED THAT A SPEAKER MALFUNCTION OCCURRED AND THE SPEAKER EMITS NO SOUND. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24068027 ↗
  • InjuryJan 7, 2026Delayed Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THE DESATURATION (DESAT) LIMIT WAS SET TO 70%, AND IT TOOK 15-20 SECONDS BEFORE THE MONITOR ALARMED FOR DESAT WHEN THE PATIENT'S SATURATION WAS AT 35%.

    Patient problems: Low Oxygen Saturation

    MAUDE report 23992977 ↗
  • MalfunctionJan 7, 2026No Audible Alarm

    PHILIPS RECEIVED A COMPLAINT ON AN INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THE SPEAKER WAS NOT WORKING; THERE WAS NO AUDIO ALARM. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23991011 ↗
  • MalfunctionDec 30, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    PHILIPS RECEIVED A COMPLAINT ON AN INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THE NON-INVASIVE BLOOD PRESSURE MEASUREMENT VALUES WERE UNSTABLE. IT WAS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23923078 ↗
  • MalfunctionDec 20, 2025Defective Alarm; Defective Alarm

    THE CUSTOMER REPORTED AN ALARM ISSUE. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23858446 ↗
  • MalfunctionDec 19, 2025No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED THERE WAS A BROKEN SPEAKER. IT WAS CONFIRMED THERE WAS NO AUDIO COMING FROM THE UNIT. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23847542 ↗
  • MalfunctionDec 15, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    THE CUSTOMER REPORTED THAT NBP VALUE IS STRANGE. IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23804495 ↗
  • MalfunctionDec 9, 2025No Audible Alarm; No Audible Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THE SPEAKER WAS DEFECTIVE. THERE WAS NO SOUND. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23763090 ↗
  • MalfunctionNov 25, 2025No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED A SPEAKER MALFUNCTION ERROR MESSAGE AND NO SOUND AFTER COMING BACK FROM BENCH REPAIR. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23646578 ↗
  • MalfunctionNov 24, 2025No Audible Alarm; No Audible Prompt/Feedback; No Audible Alarm; No Audible Prompt/Feedback

    THE CUSTOMER REPORTED THAT SPEAKER IS DEFECTIVE. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23632957 ↗
  • MalfunctionNov 10, 2025No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THE DEVICE WAS PRESENTED WITH A SPEAKER MALFUNCTION ERROR MESSAGE AND THERE WAS NO SOUND COMING FROM THE DEVICE. THE DEVICE WAS NOT IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23511087 ↗
  • MalfunctionNov 3, 2025No Audible Alarm; No Audible Alarm

    PHILIPS RECEIVED A COMPLAINT ON INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT WHILE PERFORMING PRODUCT TRAINING, A SPEAKER EQUIPMENT MALFUNCTION INOP WAS DISPLAYED AND THERE WAS NO AUDIBLE SOUND WHEN THE DEVICE WAS TURNED ON. THE DEVICE WAS NOT IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23450639 ↗
  • MalfunctionOct 22, 2025No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THERE WAS A SPEAKER MALFUNCTION ERROR SHOWN ON THE SCREEN. IT WAS CONFIRMED AUDIO WAS NOT PRESENT AT THE TIME OF THE EVENT. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23360421 ↗
  • MalfunctionOct 21, 2025Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT THE DEVICE COULD NOT MEASURE SPO2 AND THERE WAS NO CONFIRMED INOP ERROR MESSAGE DISPLAYED. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23347076 ↗
  • MalfunctionSep 2, 2025No Audible Alarm; No Audible Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE INTELLIVUE MULTI MEASUREMENT SERVER X2 INDICATING THAT THE DEVICE DISPLAYS A SPEAKER MALFUNCTION INOP ERROR MESSAGE AND DOES NOT EMIT AUDIBLE SOUND. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22931279 ↗

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page