IntelliBridge EC10 ModuleRecalls & reported problems

Philips Medizin Systeme Böblingen GmbH · The IntelliBridge EC10 module acquires real-time information from bedside devices e.g. mechanical ventilators, anesthesia machines, and multi-measurement devices for display on Philips IntelliVue patient monitors. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884838015951), received between Feb 21, 2020 and May 18, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Defective Alarm” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 15, 2025Open, Classified

    Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.

    Z-1106-2025 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Defective Alarm2
  • Device Alarm System2
  • Application Program Freezes or Fails to Launch1
  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2026Application Program Freezes or Fails to Launch

    THE CUSTOMER REPORTED THAT THE DEVICE IS INTERMITTENTLY, AND THE VITALS COMING IN FREEZES AND DOES NOT CROSS OVER. THE DEVICE WAS IN CLINICAL USE AT TIME OF EVENT, NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25211688 ↗
  • MalfunctionSep 29, 2025Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT THE CUSTOMER IS USING INTELLIBRIDGE MODULES FOR DATA TRANSMISSION. SINCE HE RECENTLY PURCHASED CAPSULE E DEVICES TO DIGITIZE THE VENTILATION DATA HE IS NO LONGER RECEIVING ALARMS ON THE CONTROL CENTER VIA THE INTELLIBRIDGE MODULES. PATIENT INVOLVEMENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23166219 ↗
  • MalfunctionDec 2, 2024Defective Alarm; Defective Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE INTELLIBRIDGE EC10 MODULE SYSTEM INDICATING AN UNSPECIFIC ALARM ISSUE. IT IS UNKNOWN IF THE DEVICE WAS NOT IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20819319 ↗
  • MalfunctionJan 31, 2023Device Alarm System; Device Alarm System

    THE CUSTOMER REPORTED THAT AN ALARM GENERATED IN YELLOW ON A PRISMAX EQUIPMENT IS TRANSCRIBED IN BLUE ON AN MX750 MONITOR. THE CUSTOMER EXPECTS THE "DYLS EMPTY BAG" ALARM TO BE CONSIDERED AS A CRITICAL ALARM AND NOT A TECHNICAL ALARM. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16270575 ↗
  • MalfunctionDec 8, 2021Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE VALUE FOR THE INSPIRED TIME DISPLAYED ON THE INTELLIVUE MP70 PATIENT MONITOR WAS MUCH HIGHER THAN THE ACTUAL READING ON THE MAQUET SERVO-I VENTILATOR WHEN USING THE INTELLIBRIDGE EC10 MODULE. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12960015 ↗
  • MalfunctionFeb 21, 2020Device Alarm System

    THE CUSTOMER REPORTED PROBLEMS WITH THE ALARM TRANSFER ON THEIR INTELLIBRIDGE EC10 MODULE. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9738637 ↗

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →