Avalon CL Toco+ MP Transducer Avalon Cableless Toco+ MP TransducerRecalls & reported problems

Philips Medizin Systeme Böblingen GmbH · Avalon Cableless Toco+ MP Transducer for use with Avalon CL Basestation 866074. Toco Transducer with Avalon Smart Pulse measurement and connector for maternal ECG, fetal ECG or IUP measurement. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00884838034235), received between Mar 4, 2020 and Sep 28, 2026 — classified as 5 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement5
  • Insufficient Device Problem Information2
  • Overheating of Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Alarm1
  • Device Reprocessing Problem1
  • Unintended Electrical Shock1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Blister2
  • Injury1
  • Insufficient Information1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924051 ↗
  • MalfunctionJul 21, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25886817 ↗
  • MalfunctionJun 12, 2026Incorrect Measurement

    PHILIPS RECEIVED A COMPLAINT ON THE AVALON CABLELESS TOCO+ MP TRANSDUCER INDICATING THAT THE DEVICE IS NOT READING CORRECTLY WHEN PRESSURE IS APPLIED. THE DEVICE WAS NOT IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25490956 ↗
  • InjuryOct 21, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED, THE TRANSDUCERS ARE CRUMBLING WHEN CLEANED WITH OXIVIR EXCEL WIPES. THIS REPORTED ISSUE IS OCCURRING AT SEVERAL CUSTOMER HOSPITAL SITES. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23349560 ↗
  • InjuryOct 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED BLISTERS WERE PRESENT ON THE PATIENT UNDER THE ULTRASOUND TRANSDUCER DURING USE. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Blister

    MAUDE report 23312353 ↗
  • MalfunctionOct 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED BLISTERS FORMED UNDER THE TRANSDUCER DURING USE. THE BLISTERING WAS RAISED, CLEAR, FLUID FILLED AND OCCURRED UNDER THE SMOOTH PART OF THE TRANSDUCER. THE TRANSDUCER WAS IN PLACE APPROXIMATELY 3-4 HOURS. THE PHYSICIAN WAS NOTIFIED BUT THE BLISTERING WAS NOT CLASSIFIED IN A BURN CATEGORY. THERE WAS NO TREATMENT PROVIDED, AND THE BURN WAS EXPECTED TO HEAL COMPLETELY. THE TRANSDUCER WAS MOVED TO ANOTHER AREA ON THE ABDOMEN, BUT IT WAS NOT NOTED BY THE USER IF THE TRANSDUCER WAS HOT TO THE TOUCH. THE SPECIFIC TRANSDUCER COULD NOT BE LOCATED FOR ANALYSIS.

    Patient problems: Blister

    MAUDE report 23299875 ↗
  • MalfunctionDec 19, 2023Defective Alarm; Defective Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE AVCL-TOCO REPLACEMENT XDR ISM INDICATING THAT THE CL TRANSDUCER LED TURNS CYAN, WITH AN INTERMITTENT AUDIBLE ALARM. IT WAS REPORTED THAT THE AVALON CL BASE DID NOT DETECT THIS TRANSDUCER, AND THERE WAS NO INOP MESSAGE ON THE MONITOR TO STATE THE BATTERY ON THIS TRANSDUCER WAS EMPTY. THE DEVICE WAS IN USE AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18355845 ↗
  • InjuryOct 31, 2022Overheating of Device; Overheating of Device

    DURING A PATIENT SCAN, THE UNIT OVERHEATED AND BURNED THE PATIENT. AS A RESULT THE PATIENT REQUIRED TREATMENT A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15704532 ↗
  • InjuryOct 28, 2022Overheating of Device; Overheating of Device

    THE CUSTOMER REPORTED THAT THE UNIT OVERHEATED AND BURNED THE PATIENT. THE PATIENT REQUIRED TREATMENT.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 15691176 ↗
  • MalfunctionJun 21, 2022Unintended Electrical Shock; Unintended Electrical Shock

    THE CUSTOMER REPORTED THAT THE UNIT NEEDS TO BE EVALUATED FOR SAFETY INCIDENT. PATIENT CLAIMED TO HAVE BEEN SHOCKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14759773 ↗
  • MalfunctionJun 17, 2022Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER REPORTED THAT THE DEVICE GIVES IRREGULAR AND UNSTABLE PRESSURE VALUES. IT IS UNKNOWN IF THE DEVICE WAS IN USE AT TIME OF EVENT AND THERE WAS NO ADVERSE EVENT TO THE PATIENT/USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14738518 ↗
  • MalfunctionSep 23, 2021Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER STATED THAT A MOTHER (YOUNG WOMAN) WAS MONITORED AND HER INDUCTION OF LABOR. MOTHER HAS A BABY WITH AUDIBLE ECTOPIC HEART BEATS. BABY WAS BEING EXTERNALLY MONITORED. ON VAGINAL EXAMINATION A FETAL SCALP ELECTRODE WAS PLACED TO PROVIDE SUPERIOR MONITORING GIVEN THE CLINICAL CONDITION OF THE BABY / ALLOW GREATER UNINHIBITED MATERNAL MOVEMENT. LARGE DARK LINES WERE SEEN ON THE MONITOR. PATIENT WAS IN ACTIVE LABOUR WHICH CONTINUED. DUE TO NOT BEING ABLE TO MONITOR FHR ACCURATELY A CODE RED - TRANSFER TO OR FOR +/- CESAREAN SECTION, WAS CALLED. PATIENT TAKEN TO THEATRE FOR DELIVERY. THERE WAS NO HARM TO MOTHER OR BABY BUT A CODE RED INTERVENTION WAS REQUIRED. PATIENT WAS TAKEN TO THEATRE BUT DID NOT REQUIRE A CESAREAN SECTION, FORCEPS DELIVERY, AND MOTHER AND BABY WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12520512 ↗
  • InjuryMar 4, 2020Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT AFTER CHILDBIRTH, THERE WAS A LARGE WOUND FOUND ON MOTHER'S BELLY. IT IS DIFFICULT TO DETERMINE IF IT WAS CAUSED BY PRESSURE, HEAT OR HYPERSENSITIVITY. THERE WAS A CONTINUOUS REGISTRATION FOR 17 HOURS, AND IT IS UNCLEAR IF THE TRANSDUCER WAS REPOSITIONED. THE WOUND WAS DISCOVERED IN CONNECTION WITH THE POST CARE A FEW HOURS AFTER BIRTH.

    Patient problems: Injury

    MAUDE report 9784993 ↗

Other MONITOR, UTERINE CONTRACTION, EXTERNAL (FOR USE IN CLINIC) devices

All MONITOR, UTERINE CONTRACTION, EXTERNAL (FOR USE IN CLINIC) devices →