EPIQ 5 Diagnostic Ultrasound System 989605408541Recalls & reported problems

Philips Ultrasound LLC · EPIQ 5 Diagnostic Ultrasound System · Class 2

All models of EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Philips EPIQ 5 ultrasound system is used for diagnostic imaging. The 71 reports mostly describe hardware breaking — control panel arms dropping or not locking, probe lenses cracking with electronics exposed, and probes leaking oil — plus some poor image quality; no deaths and no injuries were reported. Two completed recalls involved software: one from October 2023 over a version that could show a delayed image, and one from August 2021 over a defect that could freeze the system.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
68
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 00884838047686), received between Apr 12, 2019 and Sep 30, 2026 — classified as 1 injury, 67 malfunctions.

The most frequently reported specific device problem is “Unintended System Motion” (24 reports).

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 3 reportsSep 2025: 2 reportsOct 2025: 6 reportsNov 2025: 5 reportsDec 2025: 4 reportsJan 2026: 3 reportsFeb 2026: 3 reportsMar 2026: 4 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 9, 2023Class IICompleted

    Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.

    Z-0283-2024 · Root cause: Software design · Recall details · FDA record ↗

  2. Aug 6, 2021Class IICompleted

    Due to a software defect that can intermittently cause the system to lock-up which exiting Review Mode while performing an exam.

    Z-2483-2021 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unintended System Motion24
  • Material Integrity Problem15
  • Fracture6
  • Poor Quality Image5
  • Break3
  • Patient Data Problem3
  • Crack2
  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions60
  • No Known Impact Or Consequence To Patient3
  • Congenital Defect/Deformity2
  • No Consequences Or Impact To Patient2
  • Hyperextension1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Poor Quality Image

    Patient problems: Insufficient Information

    MAUDE report 50952445 ↗
  • MalfunctionSep 24, 2026Crack

    During service and testing at the manufacturing site, it was found the 3D9-3v transducer casing was cracked. The transducer was associated with an EPIQ 5W Ultrasound System at the customer site. The customer had initially reported to Philips the transducer had a cracked strain relief. A replacement transducer was shipped directly to the customer to address their concerns. The reported issue of cracked casing was found at the manufacturer site and there was no patient or user harm reported.Philips has started an investigation, and upon completion, a follow up report will be submitted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886260 ↗
  • MalfunctionMay 8, 2026Unintended System Motion

    A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ 5G ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS ENGINEER PROVIDED INSTRUCTIONS TO THE BIOMED ON PROPER PROCEDURE TO INSPECT AND SECURE CPA BEARING ON THE BACK OF THE CONTROL PANEL ASSEMBLY TO RESOLVE THE ISSUE. THE BIOMED IS TAKING RESPONSIBILITY FOR THE REPAIR. A THOROUGH ENGINEERING INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING. THE SYSTEM HAS RETURNED TO SERVICE WITH NO SIMILAR ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25131049 ↗
  • MalfunctionApr 6, 2026Material Puncture/Hole

    IT WAS REPORTED THAT THE INTRACAVITY MODEL 3D9-3V PROBE WAS LEAKING OIL FROM THE LENS AREA. THE PROBE WAS ASSOCIATED WITH THE EPIQ 5G ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE CUSTOMER'S COMPLAINT WAS ADDRESSED BY PROVIDING INFORMATION FOR A REPLACEMENT TRANSDUCER. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792899 ↗
  • MalfunctionMar 31, 2026Fracture

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THIS WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24744669 ↗
  • MalfunctionMar 3, 2026Fracture

    A CUSTOMER REPORTED THEIR C10-3V INTRACAVITY TRANSDUCER EXHIBITED SOME DAMAGE TO THE LENS TIP AREA EXPOSING A PORTION OF THE METAL LAYER UNDERNEATH. THIS PROBE IS ASSOCIATED WITH THE EPIQ 5 ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM. A QUALIFIED SERVICE ENGINEER EVALUATED THE TRANSDUCER TO CONFIRM THE REPORTED PROBLEM, AND THE TRANSDUCER WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24503556 ↗
  • MalfunctionMar 3, 2026Unintended System Motion

    IT WAS REPORTED THE CONTROL PANEL ARM ON AN EPIQ 5G ULTRASOUND SYSTEM LOWERS QUICKER THAN NORMAL. THIS ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER REPLACED THE CONTROL PANEL ARM GAS STRUTS TO RESOLVE THE CUSTOMER¿S IMMEDIATE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24496888 ↗
  • MalfunctionMar 2, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ 5G ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE SYSTEM WITHING THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. THE SERVICE ENGINEER CLEANED THE LOCKING MECHANISM TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24483272 ↗
  • MalfunctionFeb 27, 2026Fracture

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THIS WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE TRANSDUCER WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24470918 ↗
  • MalfunctionFeb 10, 2026Fracture

    A CUSTOMER REPORTED THAT THE C10-3V PROBE HAD A HOLE IN THE LENS WITH EXPOSED METAL. THE PROBE IS USED WITH THE EPIQ 5 ULTRASOUND SYSTEM. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE SERVICE ENGINEER CONFIRMED THE REPORTED ISSUE AND REPLACED THE PROBE TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305223 ↗
  • MalfunctionFeb 4, 2026Unintended System Motion

    A CUSTOMER REPORTED THE CONTROL PANEL ARM ON THEIR EPIQ 5G ULTRASOUND SYSTEM DROPPED RAPIDLY. THERE WAS NO PATIENT OR USER HARM. THE ACTUATION AND NON-ACTUATION GAS STRUTS WERE SHIPPED TO THE CUSTOMER TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24257107 ↗
  • MalfunctionJan 28, 2026Break

    IT WAS REPORTED THAT THE C10-3V TRANSDUCER HAD A HOLD IN THE LENS, RESULTING IN EXPOSED METAL. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ 5 ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS REMOTE SERVICE ENGINEER INITIATED THE REPLACEMENT PROCESS TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24203738 ↗
  • MalfunctionJan 15, 2026Fracture

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THE TRANSDUCER WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE SERVICE ENGINEER PROVIDED THE CUSTOMER WITH INFORMATION FOR A REPLACEMENT TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24081072 ↗
  • MalfunctionJan 6, 2026Material Integrity Problem

    A CUSTOMER REPORTED THEIR C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED METAL. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ 5G ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THIS ISSUE WAS FOUND OUTSIDE OF CLINICAL USE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. A PHILIPS SERVICE ENGINEER REPLACED THE TRANSDUCER TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23973812 ↗
  • MalfunctionDec 24, 2025Fracture; Fracture

    IT WAS REPORTED THE C9-4V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THIS WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. INFORMATION WAS PROVIDED TO THE CUSTOMER FOR A REPLACEMENT PART TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888286 ↗
  • MalfunctionDec 22, 2025Break; Break

    A CUSTOMER REPORTED THEIR INTRACAVITY C10-3V TRANSDUCER HAD LENS DAMAGE, EXPOSING METAL COMPONENTS. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ 5 ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23863238 ↗
  • MalfunctionDec 10, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED THEIR EPIQ 5 ULTRASOUND SYSTEM LOCKED UP MULTIPLE TIMES DURING A BREAST BIOPSY REQUIRING THE TECH TO REBOOT THE SYSTEM EACH TIME. THERE WAS NO REPORTED PATIENT OR USER HARM AS A RESULT OF THE ISSUE. NO ADDITIONAL INFORMATION COULD BE OBTAINED AT THIS TIME. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23770652 ↗
  • MalfunctionDec 3, 2025Material Integrity Problem; Material Integrity Problem

    A CUSTOMER REPORTED THEIR C10-3V TRANSDUCER¿S SCAN HEAD WAS DAMAGED, RESULTING IN EXPOSED METAL. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ 5 ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. A PHILIPS SERVICE ENGINEER INITIATED THE REPLACEMENT PROCESS TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23709267 ↗
  • MalfunctionNov 14, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THIS WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE SERVICE ENGINEER CONFIRMED THE REPORTED ISSUE AND A REPLACEMENT TRANSDUCER WAS SHIPPED DIRECTLY TO THE CUSTOMER TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23556150 ↗
  • MalfunctionNov 12, 2025Material Integrity Problem; Material Integrity Problem

    A CUSTOMER REPORTED THEIR 3D9-3V TRANSDUCER WAS LEAKING FLUID. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ 5W ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THERE WAS NO PATIENT NOR USER HARM ASSOCIATED WITH THIS EVENT. A PHILIPS FIELD SERVICE ENGINEER REPLACED THE TRANSDUCER TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23535865 ↗
  • MalfunctionNov 4, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD DAMAGE TO THE LENS. THIS WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THIS ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE SERVICE ENGINEER CONFIRMED THE REPORTED ISSUE AND REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23467267 ↗
  • MalfunctionNov 4, 2025Crack; Crack

    A CUSTOMER REPORTED THAT THE C10-3V TRANSDUCER LENS WAS BROKEN. THIS TRANSDUCER WAS ASSOCIATED WITH AN EPIQ 5 ULTRASOUND SYSTEM. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23466814 ↗
  • MalfunctionNov 4, 2025Material Integrity Problem; Material Integrity Problem

    A CUSTOMER REPORTED THEIR 3D9-3V TRANSDUCER WAS LEAKING FLUID. THIS PROBE IS ASSOCIATED WITH THE EPIQ 5 ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE, AND THERE WAS NO PATIENT OR USER HARM. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23465270 ↗
  • MalfunctionOct 30, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS. THIS WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE CUSTOMER¿S IMMEDIATE CONCERNS WERE ADDRESSED BY PROVIDING INFORMATION TO REPLACE THE TRANSDUCER. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23425632 ↗
  • MalfunctionOct 29, 2025Material Integrity Problem; Material Integrity Problem

    A CUSTOMER REPORTED THEIR C10-3V INTRACAVITY TRANSDUCER HAD HOLES IN THE LENS EXPOSING ELECTRONICS. THIS PROBE IS ASSOCIATED WITH THE EPIQ 5G ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE, AND THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER EVALUATED THE TRANSDUCER TO CONFIRM THE REPORTED PROBLEM AND REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23416638 ↗

Other System, imaging, pulsed echo, ultrasonic devices

All System, imaging, pulsed echo, ultrasonic devices →