Skip to content

IntelliSpace Critical Care and Anesthesia Intellispace CC & AnesthesiaRecalls & reported problems

Philips Medical Systems Hsg · IntelliSpace Critical Care and Anesthesia · Class 2

All models of IntelliSpace Critical Care and Anesthesia (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00884838085565), received between Jun 6, 2019 and May 5, 2026 — classified as 3 injuries, 15 malfunctions.

The most frequently reported specific device problem is “Application Program Problem: Medication Error” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem: Medication Error4
  • Data Problem2
  • Device Handling Problem2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Output Problem2
  • Appropriate Device Problem Term/Code Not Available1
  • Device Displays Incorrect Message1
  • Failure to Read Input Signal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Known Impact Or Consequence To Patient4
  • Drug Resistant Bacterial Infection1
  • Hypersensitivity/Allergic reaction1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 5, 2026Device Handling Problem

    IT WAS REPORTED THAT IT IS DIFFICULT TO CHART ALLERGY IN ICCA AND MOREOVER THE DISPLAY IS TRUNCATED. IF THERE ARE SEVERAL ALLERGIES DOCUMENTED, IT DISPLAYS 3 DOTS BUT NOT ALL THE CONTENT. IN CONFIGURATION, ALLERGY INTERVENTION IS HARDCODED AND THERE IS NO POSSIBILITY TO ADD ALLERGY AND INTOLERANCE WITH DIFFERENT DOCUMENTATION INSIDE. A PATIENT'S PLANNED SURGERY WAS CANCELLED AND RESCHEDULED DUE TO THE FACT THAT A LATEX ALLERGY WAS NOT SEEN (DISPLAY TRUNCATED). NO ACTUAL HARM IS REPORTED HOWEVER, AN ADVERSE EVENT FOR PATIENT IS DESCRIBED AS A TYPE OF HARM. LATER, IT WAS CONFIRMED THAT THERE WAS NO HARM TO THE PATIENT. MEDICAL ATTENTION AND PREVENTIVE MEDICAL ATTENTION BOTH ARE MARKED AS YES BUT INFORMATION ABOUT MEDICAL CONDITION WAS NOT PROVIDED. GOOD FAITH EFFORTS WERE MADE TO GET MORE INFORMATION ABOUT MEDICAL ATTENTION AND PREVENTIVE MEDICAL ATTENTION, BUT FURTHER INFORMATION COULD NOT BE RECEIVED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 25084157 ↗
  • InjuryMay 5, 2026Device Handling Problem

    IT WAS REPORTED THAT ALLERGY INTERVENTION IS HARD CODED IN ICCA. THERE IS NO POSSIBILITY TO INSERT ALLERGY AND INTOLERANCE WITH DIFFERENT DOCUMENTATION INSIDE. THE CUSTOMER IS NOT HAPPY WITH IT BECAUSE IT IS TOO COMPLEX TO FILL. IT IS ALSO MENTIONED THAT, IT CANNOT BE REPLACED IN PATIENT CHART BY AN OTHER INTERVENTION. IF USER FILL ANOTHER INTERVENTION (WITHOUT REMINDERS), THE CRITICAL INFORMATION OF ALLERGY WOULD NOT BE SEEN. THOUGH PATIENT HARM IS NOT REPORTED, ANAPHYLACTIC SHOCK IS PROVIDED AS IMPACT TO THE PATIENT. AN ATTEMPT HAS BEEN MADE TO CONFIRM THE IMPACT TO THE PATIENT BUT NO FURTHER INFORMATION WAS PROVIDED BY THE CUSTOMER.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 25083979 ↗
  • MalfunctionAug 23, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT ICCA HAS AN INTAKE MISCALCULATION ISSUE. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17603210 ↗
  • InjuryJun 13, 2023Application Program Problem: Medication Error; Application Program Problem: Medication Error

    THE CUSTOMER REPORTED THAT THE INTELLISPACE CC & ANESTHESIA INDICATING THAT PRESCRIPTION ON ICCA SOFTWARE ON 21/04 OF LEVOFLOXACIN DOES NOT APPEAR IN THE NURSING CARE PLAN. THE DEVICE WAS IN USE AT TIME OF EVENT, AN ADVERSE EVENT WAS REPORTED.

    Patient problems: Drug Resistant Bacterial Infection

    MAUDE report 17119978 ↗
  • MalfunctionMay 22, 2023Labelling, Instructions for Use or Training Problem; Application Program Problem: Medication Error; Labelling, Instructions for Use or Training Problem; Application Program Problem: Medication Error

    THE CUSTOMER REPORTED THAT SOME TIME SCHEDULES DID NOT DISPLAY ON PIPE/TAZO ORDER, SO THE PATIENT DIDN'T RECEIVE THE DRUG WHEN IT SHOULD HAVE BEEN. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16974701 ↗
  • MalfunctionMay 17, 2023Data Problem; Data Problem

    THE CUSTOMER REPORTED THAT THE NURSE AT THE BEDSIDE DID NOT AUTOMATICALLY SEE PRESCRIBED MEDICATION. THEREFORE, PRESCRIBED SCHEDULES CAN POTENTIALLY BE MISSED. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16950199 ↗
  • MalfunctionMay 17, 2023Data Problem; Data Problem

    THE CUSTOMER REPORTED THAT WHEN A USER SAVES DATA FOR A GIVEN PATIENT, THE UPDATE DOES NOT PUSH ALL OTHER CLIENTS WHO HAVE THE RECORD OPEN; THE CLINICAL TEAMS DO NOT HAVE THE THE LATEST DATA AVAILABLE AT THE BED SPACE. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16950196 ↗
  • MalfunctionApr 29, 2023Failure to Read Input Signal; Failure to Read Input Signal

    PHILIPS RECEIVED A COMPLAINT ON INTELLISPACE CC & ANESTHESIA INDICATING THATIM961311 DOES NOT ALLOW TO ASSIGN THE PERFUSIONS THAT THEY HAVE SCHEDULED AND ASSIGNS PERFUSIONS THAT THEY HAVE NOT HAD SCHEDULED. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16841140 ↗
  • MalfunctionApr 26, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT THE ICCA DRUG CHART WAS MISSING SCHEDULED MEDICATIONS, RESULTING IN MISSED DOSES FOR CRITICAL MEDICATIONS. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16820825 ↗
  • MalfunctionFeb 22, 2023Inaccurate Delivery; Inaccurate Delivery

    THE CUSTOMER REPORTED THAT FOR BED 7 IP-003 AND IP-015, INCORRECT VOLUME WAS INFUSED. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED. A PHILIPS REMOTE SERVICE ENGINEER (RSE) SPOKE TO THE CUSTOMER. THE RSE FOUND THERE WERE TWO PUMPS THAT ENCOUNTERED MULTIPLE DISCONNECTION ISSUES; IP-015 FROM 20 FEB 2023 2350H TO 21 FEB 2023 0306H(LOCATED AT CHANNEL 4), WHICH WAS THEN REPLACED BY IP-003(FROM CHANNEL 2 TO CHANNEL 4). ASIDE FROM THE MULTIPLE DISCONNECTIONS, THE RSE ALSO DETECTED A ¿RESET OF TOTAL INFUSED VOLUME¿ SIX TIMES BETWEEN 2350H TO 0943H (PLUS THE PUMP REPLACEMENT AT 0306H AND THE PUMP RE-ASSIGNMENT AFTER FLUID LABEL CHANGE). THE RSE INFORMED THE NURSE TO HELP TO MONITOR THE ISSUE, AND ASKED THE NURSE TO TRY AND SWAP THE LINK BRIDGE CURRENTLY IN USE BY IP-003 IN CASE THE ISSUE HAPPENS AGAIN FOR TROUBLESHOOTING/ISOLATION PURPOSES. IF THE ISSUE PERSISTS AFTER REPLACING/SWAPPING THE LINK BRIDGE, THE RSE WILL PROCEED TO REQUEST AN LINK BRIDGE (EC80) REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16417756 ↗
  • MalfunctionFeb 17, 2023Application Program Problem: Medication Error; Application Program Problem: Medication Error

    PHILIPS RECEIVED A COMPLAINT BY THE CUSTOMER ON INTELLISPACE CC & ANESTHESIA INDICATING INCORRECT SPEED DUMP. IT GOES AT A SPEED OF 1ML/H AND IN THE ICCA IT SHOWS 40ML/H. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16396746 ↗
  • MalfunctionJan 19, 2023Incorrect, Inadequate or Imprecise Result or Readings; Infusion or Flow Problem; Incorrect, Inadequate or Imprecise Result or Readings; Infusion or Flow Problem

    THE CUSTOMER REPORTED THAT THE FLUID MANAGEMENT IN ANESTHESIA IS NOT CORRECTLY ASSIGNED. ROOM 1 AND ROOM 9 SEEM TO BE SWAPPED. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED. A PHILIPS REMOTE SERVICE ENGINEER (RSE) REMOTELY CONNECTED TO THE CONTROL CENTER AT THE CUSTOMER SITE AND DETERMINED THE INFUSION PUMPS WOULD REQUIRE CONNECTION VIA THE PATIENT INFORMATION CENTER IX (PIIC IX). THE DEVICE REMAINS AT THE CUSTOMER SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16205613 ↗
  • MalfunctionMay 6, 2020Application Program Problem: Medication Error

    THE CUSTOMER ALLEGED ¿LEVEL OF THE DRUG CIRCUIT FOR THE SYRINGE PUMP RELAYS: TRACEABILITY OF ERRONEOUS ADMINISTRATION WITH INCORRECT VALUES OF SPEED OR NAME OF DRUGS USED TOTALLY WRONG". THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 10029565 ↗
  • MalfunctionNov 22, 2019Device Displays Incorrect Message

    THE CUSTOMER IS REPORTING THAT A MEDICATION WAS INCORRECTLY DISPLAYING 'OVERDUE' IN THE MAR. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9362255 ↗
  • MalfunctionOct 17, 2019Output Problem

    THE CUSTOMER ALLEGED LAB REQUESTS HAVE NOT BEEN PRINTING. NO ADVERSE EVENT OCCURRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9204422 ↗
  • MalfunctionOct 17, 2019Output Problem

    THE CUSTOMER ALLEGED LAB REQUESTS HAVE NOT BEEN PRINTING. NO ADVERSE EVENT OCCURRED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9204305 ↗
  • MalfunctionJul 10, 2019Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER ALLEGED END OF 2018 THE PROPERTY ¿ROUTE¿ WAS REMOVED FROM THE INTERVENTIONS. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR PATIENT HARM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8775149 ↗
  • MalfunctionJun 6, 2019Patient Data Problem

    THE CUSTOMER ALLEGED A PRESCRIPTION IS NOT DISPLAYED AFTER SAVING, BUT IS PRESENT, AND ONLY VISIBLE AFTER ANOTHER LOGIN. THE DEVICE WAS IN USE ON A PATIENT AND THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8674671 ↗

Other DISPLAY, CATHODE-RAY TUBE, MEDICAL devices

All DISPLAY, CATHODE-RAY TUBE, MEDICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page