Avalon Ultrasound TransducerRecalls & reported problems
Philips Medizin Systeme Böblingen GmbH · The device is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Philips Avalon ultrasound transducer is used to monitor fetal and other vital signs. The 41 reports mostly describe inaccurate or out-of-range heart rate readings, including when monitoring twins or multiples, and mention 2 deaths and 2 injuries. There is 1 open recall started in May 2024 for inaccurate fetal heart rate measurements when monitoring multiples. Reports fell to 0 in the last 12 months from 14 in the prior 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 41
- Deaths reported
- 2
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 00884838093195), received between May 8, 2024 and Oct 30, 2025 — classified as 2 deaths, 2 injuries, 37 malfunctions.
The most frequently reported specific device problem is “High Readings” (28 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 22, 2024Class IIOpen, Classified
Potential for inaccurate fetal heart rate measurements when monitoring multiples.
Z-2239-2024 · Root cause: Component change control · Recall details · FDA record ↗
Most reported problems
Device problems
- High Readings28
- Incorrect Measurement10
- Incorrect, Inadequate or Imprecise Result or Readings7
- Insufficient Device Problem Information2
- High Test Results1
- Low Readings1
- No Audible Alarm1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions38
- Fetal Distress2
- Insufficient Information2
Recent adverse event reports
The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
PHILIPS RECEIVED A COMPLAINT ON AN AVALON US TRANSDUCER INDICATING THAT THE VALUE IS NOT IN TOLERANCE. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23423292 ↗MalfunctionSep 12, 2025Incorrect Measurement; Incorrect Measurement
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT THE SENSOR DOES NOT GIVE THE CORRECT VALUES. IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23040248 ↗MalfunctionFeb 26, 2025Low Readings; Low Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT THE HEART RATE MONITORED FROM THE TRANSDUCER IS LOW. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21468984 ↗MalfunctionFeb 18, 2025High Readings; High Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT THERE WERE INACCURATE FETAL HEART RATE (FHR) MEASUREMENTS WHEN MONITORING MULTIPLE FETUSES. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF EVENT. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21399623 ↗MalfunctionFeb 18, 2025High Readings; High Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT THERE WERE INACCURATE FETAL HEART RATE (FHR) MEASUREMENTS WHEN MONITORING MULTIPLE FETUSES. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF EVENT. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21399142 ↗MalfunctionFeb 18, 2025High Readings; High Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT THERE WERE INACCURATE FETAL HEART RATE (FHR) MEASUREMENTS WHEN MONITORING MULTIPLE FETUSES. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF EVENT. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21397392 ↗MalfunctionFeb 18, 2025High Readings; High Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT THERE WERE INACCURATE FETAL HEART RATE (FHR) MEASUREMENTS WHEN MONITORING MULTIPLE FETUSES. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF EVENT. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21395613 ↗MalfunctionDec 27, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN A PAIR OF TWINS IS BEING MONITORED ONE OF THE TRANSDUCERS WILL PICK UP A ABNORMAL RHYTHM. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025774 ↗MalfunctionDec 27, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN A PAIR OF TWINS IS BEING MONITORED ONE OF THE TRANSDUCERS WILL PICK UP A ABNORMAL RHYTHM. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025771 ↗MalfunctionDec 27, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN A PAIR OF TWINS IS BEING MONITORED ONE OF THE TRANSDUCERS WILL PICK UP A ABNORMAL RHYTHM. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025766 ↗MalfunctionDec 27, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN A PAIR OF TWINS IS BEING MONITORED ONE OF THE TRANSDUCERS WILL PICK UP A ABNORMAL RHYTHM. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025580 ↗MalfunctionDec 27, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN A PAIR OF TWINS IS BEING MONITORED ONE OF THE TRANSDUCERS WILL PICK UP A ABNORMAL RHYTHM. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025459 ↗MalfunctionDec 27, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN A PAIR OF TWINS IS BEING MONITORED ONE OF THE TRANSDUCERS WILL PICK UP A ABNORMAL RHYTHM. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025110 ↗MalfunctionDec 3, 2024High Readings; High Readings
IT WAS REPORTED THAT WHILE A PAIR OF TWINS WAS BEING MONITORED, ONE OF THE TRANSDUCERS PICKED UP A ABNORMAL RHYTHM. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20833380 ↗MalfunctionOct 3, 2024Incorrect Measurement; Incorrect Measurement
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THE DEVICE ACQUIRES AN INCORRECT MEASUREMENT. THE DEVICE WAS IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20359966 ↗MalfunctionSep 17, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT THERE WAS A VALUE PROBLEM. IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20235560 ↗MalfunctionAug 21, 2024High Readings; High Readings
THE CUSTOMER REPORTED THAT THE MONITOR HOOKED UP TO THE TWINS, SHOWED THE HEART RATE KEEPS GOING TO 200. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20013685 ↗MalfunctionAug 13, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN MONITORING TWINS, THE FETAL HEART RATE (FHR) MEASURED BY THE US TRANSDUCER IS INACCURATE. THE HEART RATE IS HIGHER THAN THE VALUE MEASURED BY A STETHOSCOPE OR AN ECHOGRAPHY. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED. THE INVESTIGATION HAS BEEN ESCALATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19977000 ↗MalfunctionAug 13, 2024High Readings; High Readings
IT WAS REPORTED THAT WHEN MONITORING TWINS, THE FETAL HEART RATE (FHR) MEASURED BY THE US TRANSDUCER IS INACCURATE. THE HEART RATE IS HIGHER THAN THE VALUE MEASURED BY A STETHOSCOPE OR AN ECHOGRAPHY. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED. THE INVESTIGATION HAS BEEN ESCALATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19974705 ↗MalfunctionAug 6, 2024Incorrect Measurement; High Test Results; Incorrect Measurement; High Test Results
THE CUSTOMER REPORTED THAT THE MONITOR HOOKED UP TO THE TWINS, SHOWED THE HEART RATE KEEPS GOING TO 200. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19920697 ↗MalfunctionAug 5, 2024Incorrect Measurement; High Readings; Incorrect Measurement; High Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCE INDICATING THAT THE SYSTEM CONTINUES TO SHOW TOO HIGH VALUES FOR FETAL HEART RATES 180BPM". THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT AND NO ADVERSE EVENT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19906612 ↗MalfunctionJul 19, 2024No Audible Alarm; No Audible Alarm
BIOMED STATES THE US TRANSDUCER DOES NOT HAVE SOUND; NEEDS TO ORDER A REPLACEMENT TRANSDUCER. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19780247 ↗MalfunctionJul 11, 2024High Readings; High Readings
IT WAS REPORTED THAT THE TWIN PREGNANCY IS RECORDING THE INCORRECT FHR (180BPM). THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.THE CUSTOMER HAS BEEN NOTIFIED THAT THIS IS A KNOWN ISSUE AND FCO86202016A IS EXPECTED TO BE IMPLEMENTED IN THE NEAR FUTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19724302 ↗MalfunctionJul 1, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings
PHILIPS RECEIVED A COMPLAINT ON THE AVALON US TRANSDUCER INDICATING THAT "SYSTEM CONTINUES TO SHOW TOO HIGH VALUES FOR FETAL HEART RATES 180BPM". THE DEVICE WAS IN CLINICAL USE AT THE TIME OF EVENT AND NO ADVERSE EVENT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19647051 ↗DeathJul 1, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THE SITE HAD TWO SENTINEL EVENTS DESCRIBED AS DEATH WITH UNKNOWN ROOT CAUSE. THE CUSTOMER WAS STILL COMPILING THEIR REPORT SO THEY DID NOT HAVE MUCH ADDITIONAL INFORMATION BUT THEY WERE WONDERING IF THE TRANSDUCER FIELD ACTION IMPACTED SINGLE WIRED TRANSDUCERS OR CABLELESS TRANSDUCERS. BASED ON THIS INFORMATION, IT CANNOT BE DETERMINED WHETHER THERE WAS ANY HARM RELATED TO PHILIPS DEVICES. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 19643577 ↗
Other System, Monitoring, Perinatal devices
- Avalon Fetal Monitor FM30 Avalon FM30 Fetal MonitorPhilips Medizin Systeme Böblingen GmbH · SYSTEM, MONITORING, PERINATAL1 recall11 deaths142 reports
- Avalon Fetal Monitor FM20 Avalon FM20 Fetal MonitorPhilips Medizin Systeme Böblingen GmbH · SYSTEM, MONITORING, PERINATAL1 recall126 reports
- Avalon Fetal Monitor FM50 Avalon FM50 Fetal MonitorPhilips Medizin Systeme Böblingen GmbH · SYSTEM, MONITORING, PERINATAL9 deaths51 reports
- Central Monitoring System MFM-CNS LiteEdan Instruments, Inc. · System, Monitoring, Perinatal1 recall0 reports
- Fetal & maternal monitor F6 ExpressEdan Instruments, Inc. · System, Monitoring, Perinatal1 recall0 reports
- Fetal & maternal monitor F9Edan Instruments, Inc. · System, Monitoring, Perinatal1 recall0 reports
- Central Monitoring System MFM-CNSEdan Instruments, Inc. · System, Monitoring, Perinatal1 recall0 reports
- Fetal & maternal monitor F6Edan Instruments, Inc. · System, Monitoring, Perinatal1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page