Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884838097711), received between Sep 4, 2024 and Jan 2, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Unintended Electrical Shock” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Unintended Electrical Shock1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Clinical Signs, Symptoms or Conditions1
  • Numbness1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLEARVUE 850 ULTRASOUND SYSTEM WAS ASSOCIATED WITH A PATIENT SERIOUS INJURY EVENT. IT WAS REPORTED THE TRANSDUCER MALFUNCTIONED DURING NORMAL USE WITH THEIR CLEARVUE 850 ULTRASOUND SYSTEM. THE CUSTOMER ALLEGED THE MALFUNCTION CAUSE SERIOUS INJURY TO THE PATIENT. NO OTHER CLINICAL INFORMATION OR MEDICAL INTERVENTION WAS REPORTED. AT THIS TIME THERE IS INSUFFICIENT INFORMATION WITHIN THE EVENT DESCRIPTION TO DETERMINE IF THE DEVICE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21063193 ↗
  • InjurySep 4, 2024Unintended Electrical Shock; Unintended Electrical Shock

    IT WAS REPORTED THE USER FELT OCCASIONAL ELECTRICAL LEAKAGE FROM THE C5-2 TRANSDUCER. THE USER REPORTED SENSING A LEAK AND HAVING NUMBNESS, BUT NO MEDICAL INTERVENTION WAS REQUIRED. THERE WAS NO PATIENT HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER REPORTED THE TRANSDUCER PASSES THE LEAKAGE TEST. THE CUSTOMER¿S IMMEDIATE CONCERNS WERE ADDRESSED BY REPLACING THE TRANSDUCER. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: Numbness

    MAUDE report 20137321 ↗

Other System, imaging, pulsed echo, ultrasonic devices

All System, imaging, pulsed echo, ultrasonic devices →