EPIQ Elite Diagnostic Ultrasound System 795234 (UDI 00884838102514)Recalls & reported problems
Philips Ultrasound LLC · Class 2
All models of EPIQ Elite Diagnostic Ultrasound System (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00884838102514), received between Jan 2, 2025 and Mar 31, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 7, 2026Class IIOpen, Classified
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
Z-3249-2026 · Root cause: Software Design Change · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Integrity Problem4
- Unintended System Motion4
- Material Split, Cut or Torn3
- Break1
- Electrical /Electronic Property Problem1
- Fracture1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2026Fracture
IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THIS WAS ASSOCIATED WITH AN EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE. THERE WAS NO PATIENT OR USER HARM. PHILIPS HAS STARTED AN INVESTIGATION AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24744670 ↗MalfunctionMar 11, 2026Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS SERVICE ENGINEER REPAIRED THE SYSTEM BY ADJUSTING THE BUSHING WHICH ALLOWED THE LOCKING MECHANISM TO BE FULLY ENGAGED. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24568802 ↗MalfunctionMar 10, 2026Unintended System Motion
IT WAS REPORTED THAT, WHEN RELEASING THE CONTROL PANEL LOCK ON THE EPIQ ELITE ULTRASOUND SYSTEM, THE CONTROL PANEL LOWERED FASTER THAN EXPECTED. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, DURING PREVENTATIVE MAINTENANCE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE PHILIPS SERVICE ENGINEER HAS INITIATED THE REPAIR PROCESS TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERN. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24563001 ↗MalfunctionMar 4, 2026Material Split, Cut or Torn
A CUSTOMER REPORTED THEIR C10-3V INTRACAVITY TRANSDUCER EXHIBITED SOME DAMAGE TO THE LENS TIP AREA EXPOSING A PORTION OF THE METAL LAYER UNDERNEATH. THIS PROBE IS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM. THE CUSTOMER'S COMPLAINT WAS ADDRESSED BY PROVIDING INFORMATION FOR A REPLACEMENT TRANSDUCER. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24507626 ↗MalfunctionFeb 4, 2026Break
IT WAS REPORTED THAT THE C10-3V TRANSDUCER LENS HAD A HOLE, RESULTING IN EXPOSED METAL. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ ELITE DIAGNOSTIC ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. A PHILIPS REMOTE SERVICE ENGINEER INITIATED THE REPLACEMENT PROCESS TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24264169 ↗MalfunctionJan 29, 2026Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS FIELD SERVICE ENGINEER PERFORMED A SYSTEM INSPECTION AND, AS A CORRECTION, RESEATED AND REAFFIXED THE REAR ARM BUSHING ASSEMBLY TO RESOLVE THE ISSUE. A THOROUGH ENGINEERING INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING. THE SYSTEM HAS RETURNED TO SERVICE WITH NO SIMILAR ISSUES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24217731 ↗MalfunctionJan 13, 2026Material Integrity Problem
A CUSTOMER REPORTED THEIR 3D9-3V TRANSDUCER WAS LEAKING OIL. THIS PROBE IS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE, AND THERE WAS NO PATIENT OR USER HARM. A REPLACEMENT TRANSDUCER WAS SHIPPED DIRECTLY TO THE CUSTOMER TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24057320 ↗MalfunctionNov 24, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
A CUSTOMER REPORTED THEIR C10-3V TRANSDUCER¿S LENS WAS DAMAGED, RESULTING IN EXPOSED METAL. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE OCCURRED OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. A PHILIPS FIELD SERVICE ENGINEER REPLACED THE TRANSDUCER TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23639401 ↗MalfunctionOct 31, 2025Material Integrity Problem; Material Integrity Problem
A CUSTOMER REPORTED THEIR C10-3V TRANSDUCER HAD DAMAGED ON THE LENS EXPOSING CRYSTALS. THIS PROBE IS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE, AND THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER EVALUATED THE TRANSDUCER TO CONFIRM THE REPORTED PROBLEM AND REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434158 ↗MalfunctionOct 27, 2025Electrical /Electronic Property Problem; Material Integrity Problem; Electrical /Electronic Property Problem; Material Integrity Problem
IT WAS REPORTED THE 3D9-3V TRANSDUCER HAD LENS DAMAGE THAT FAILED THE ELECTRICAL SAFETY TEST. THE TRANSDUCER WAS ASSOCIATED WITH AN EPIQ ELITE ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. A REPLACEMENT TRANSDUCER WAS SHIPPED DIRECTLY TO THE CUSTOMER TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23392890 ↗MalfunctionSep 15, 2025Material Integrity Problem; Material Integrity Problem
A CUSTOMER REPORTED THEIR C10-3V TRANSDUCER HAD A HOLE IN THE LENS EXPOSING ELECTRONICS. THIS PROBE IS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE, AND THERE WAS NO PATIENT OR USER HARM. A REPLACEMENT PROBE WAS SHIPPED DIRECTLY TO THE CUSTOMER TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23057080 ↗MalfunctionSep 8, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
A CUSTOMER REPORTED THE C10-3V TRANSDUCER HAD DAMAGE TO THE LENS SURFACE EXPOSING METAL UNDERNEATH. THE TRANSDUCER WAS ASSOCIATED WITH AN EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. INFORMATION WAS PROVIDED REGARDING REPLACEMENT OF THE TRANSDUCER. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22994679 ↗MalfunctionAug 27, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE INTRACAVITY 3D9-3V PROBE WAS LEAKING OIL FROM TIP. THE PROBE WAS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT TRANSDUCER TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22903897 ↗MalfunctionJan 2, 2025Unintended System Motion; Unintended System Motion
IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL ARM WILL NOT LOCK INTO POSITION WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. NO PATIENT OR USER WAS HARMED BECAUSE OF THE ISSUE. A PHILIPS SERVICE ENGINEER REMOVED THE CONTROL PANEL ARM COVERS AND FOUND THE BUSHING AROUND THE LOWER SWIVEL LOCK LATCH WAS OUT OF PLACE. THE PHILIPS SERVICE ENGINEER CLEANED THE GREASE AND RESIDUE FROM THE HOLE WHERE THE BUSHING SEATS AND APPLIED LOCTITE 480 TO THE BUSHING AND SEATED IT FIRMLY IN PLACE TO RESOLVE THE ISSUE. A THOROUGH ENGINEERING INVESTIGATION HAS BEEN PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21057342 ↗
Other Transducer, ultrasonic, diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer L15-7io 989605387852Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page