EPIQ CVx Diagnostic Ultrasound System 795231Recalls & reported problems

Philips Ultrasound LLC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Philips EPIQ CVx ultrasound system has 119 adverse event reports, with 4 injuries and no deaths. Many describe the control panel arm dropping or the swivel not locking, and some describe software freezing or a probe that would not articulate. Many reports note that no patient or user was harmed.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
119
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 119 adverse event reports that list this exact device (DI 00884838107526), received between Oct 7, 2022 and May 28, 2026 — classified as 4 injuries, 115 malfunctions.

The most frequently reported specific device problem is “Unintended System Motion” (71 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 5 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 5 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 4 reportsMar 2025: 4 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 4 reportsAug 2025: 3 reportsSep 2025: 3 reportsOct 2025: 7 reportsNov 2025: 2 reportsDec 2025: 4 reportsJan 2026: 3 reportsFeb 2026: 3 reportsMar 2026: 7 reportsApr 2026: 5 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended System Motion71
  • Application Program Freezes or Fails to Launch14
  • Mechanical Jam5
  • Electrical /Electronic Property Problem4
  • Material Integrity Problem4
  • Adverse Event Without Identified Device or Use Problem3
  • Material Split, Cut or Torn3
  • Unintended Application Program Shut Down3

Patient problems

  • No Clinical Signs, Symptoms or Conditions115
  • Perforation of Esophagus2
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 119 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. A PHILIPS SERVICE ENGINEER REPAIRED THE SYSTEM BY RESEATING CONTROL PANEL ARM BUSHING, WHICH ALLOWED THE LOCKING MECHANISM TO BE FULLY ENGAGED. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25316613 ↗
  • MalfunctionMay 19, 2026Unintended System Motion

    IT WAS REPORTED THE CONTROL PANEL ARM ON AN EPIQ CVX ULTRASOUND SYSTEM DROPPED QUICKLY. THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER REPLACED THE CONTROL PANEL ARM GAS STRUT TO RESOLVE THE CUSTOMER¿S IMMEDIATE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219911 ↗
  • MalfunctionMay 19, 2026Mechanical Jam

    IT WAS REPORTED THAT AN X8-2T TRANSDUCER HAD AN ARTICULATING PROBLEM DURING USE, IT DID NOT TURN RIGHT OR LEFT. THE PROBE WAS ASSOCIATED WITH THE EPIQ CVX ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER EVALUATED THE TRANSDUCER TO CONFIRM ARTICULATION PROBLEM. THE SERVICE ENGINEER INITIATED THE REPLACEMENT PROCESS TO RESOLVE THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219560 ↗
  • MalfunctionApr 30, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. A PHILIPS SERVICE ENGINEER REPAIRED THE SYSTEM BY RESEATING THE BUSHING, WHICH ENSURED THE LOCKING MECHANISM FULLY ENGAGED. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25044310 ↗
  • MalfunctionApr 24, 2026Unintended System Motion

    IT WAS REPORTED THAT THE CONTROL PANEL ARM ON AN EPIQ CVX ULTRASOUND SYSTEM DROPPED RAPIDLY. THIS ISSUE WAS FOUND OUTSIDE OF CLINICAL USE, AND THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER REPLACED THE CONTROL PANEL ARM GAS STRUTS TO RESOLVE THE CUSTOMER¿S IMMEDIATE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24987193 ↗
  • MalfunctionApr 23, 2026Unintended System Motion

    IT WAS REPORTED THE CONTROL PANEL ARM ON AN EPIQ CVX ULTRASOUND SYSTEM DROPPED RAPIDLY. THIS ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER REPLACED THE CONTROL PANEL ARM GAS STRUTS TO RESOLVE THE CUSTOMER¿S IMMEDIATE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966232 ↗
  • MalfunctionApr 21, 2026Unintended System Motion

    A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE PHILIPS SERVICE ENGINEER REPLACED THE BEARING TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERN. A THOROUGH ENGINEERING INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING. THE SYSTEM HAS RETURNED TO SERVICE WITH NO SIMILAR ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24942793 ↗
  • MalfunctionApr 6, 2026Material Integrity Problem

    IT WAS REPORTED THE X8-2T TRANSDUCER HAD A DAMAGED AND DETACHED SURFACE. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. PHILIPS HAS STARTED AN INVESTIGATION AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789886 ↗
  • MalfunctionMar 24, 2026Material Integrity Problem

    A CUSTOMER REPORTED THAT THE TRANSESOPHAGEAL X8-2T TRANSDUCER¿S SEALANT ON THE TIP WAS COMING OFF. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ CVX ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24673478 ↗
  • MalfunctionMar 23, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE SYSTEM WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OF USER WAS HARMED. THE SERVICE ENGINEER RESEATED AND SECURED THE ARTICULATION ARM BUSHING TO REPAIR THE SYSTEM. THE ENGINEER TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663489 ↗
  • MalfunctionMar 18, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626757 ↗
  • MalfunctionMar 13, 2026Unintended System Motion

    A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE SOLENOID WAS REPLACED TO RESOLVE THE CUSTOMER'S IMMEDIATE CONCERNS. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. RESULTS OF THE INVESTIGATION WILL BE INCLUDED IN A FOLLOW UP REPORT UPON ITS COMPLETION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590563 ↗
  • MalfunctionMar 10, 2026Unintended System Motion

    A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK AS EXPECTED WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS FIELD SERVICE ENGINEER PUSHED THE COLLET BACK INTO PLACE TO RESOLVE THE CUSTOMER'S IMMEDIATE CONCERNS. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING. THE SYSTEM HAS RETURNED TO SERVICE WITH NO SIMILAR ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24564003 ↗
  • MalfunctionMar 10, 2026Material Integrity Problem

    A CUSTOMER REPORTED THEIR X8-2T TRANSDUCER SEAL WAS MISSING. THIS PROBE IS ASSOCIATED WITH THE EPIQ CVX ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE, AND THERE WAS NO PATIENT OR USER HARM. THE REMOTE ENGINEER EVALUATED THE TRANSDUCER TO CONFIRM THE REPORTED PROBLEM, AND INFORMATION HAS BEEN PROVIDED TO REPLACE THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24562451 ↗
  • MalfunctionMar 6, 2026Unintended System Motion

    A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS FIELD SERVICE ENGINEER REPLACED A CONTROL PANEL GROMMET BUSHING TO RESOLVE THE ISSUE. A THOROUGH ENGINEERING INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING. THE SYSTEM HAS RETURNED TO SERVICE WITH NO SIMILAR ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24533103 ↗
  • MalfunctionFeb 27, 2026Material Split, Cut or Torn

    IT WAS REPORTED THE X7-2T TRANSDUCER HAD A SMALL TEAR IN THE LENS THAT WAS LEAKING FLUID. THIS WAS ASSOCIATED WITH AN EPIQ CVX ULTRASOUND SYSTEM. THIS ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM. THE TRANSDUCER WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24472399 ↗
  • MalfunctionFeb 20, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF AN EPIQ CVX ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE SERVICE ENGINEER RESEATED THE BUSHING TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401226 ↗
  • MalfunctionFeb 11, 2026Unintended System Motion

    A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A QUALIFIED SERVICE ENGINEER REPLACED THE BEARING TO RESOLVE THE CUSTOMER¿S IMMEDIATE ISSUE. A THOROUGH ENGINEERING INVESTIGATION HAS BEEN PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24314039 ↗
  • MalfunctionJan 28, 2026Unintended System Motion

    IT WAS REPORTED THAT THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE PHILIPS FIELD SERVICE ENGINEER CLEANED AND RESEATED THE BUSHING WITH LOCTITE ADHESIVE TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERN. A THOROUGH ENGINEERING INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING. THE SYSTEM HAS RETURNED TO SERVICE WITH NO SIMILAR ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24205563 ↗
  • MalfunctionJan 15, 2026Unintended System Motion

    THE CUSTOMER REPORTED WHEN RELEASING THE FRONT CONTROL PANEL HANDLE LOCK ON THE EPIQ CVX ULTRASOUND SYSTEM, THE CONTROL PANEL STARTED TO LOWER IMMEDIATELY AND FASTER THAN EXPECTED. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. NOTHING DETACHED FROM THE SYSTEM. THE PHILIPS SERVICE ENGINEER REPLACED THE CONTROL PANEL ARM GAS STRUTS TO RESOLVE THE CUSTOMER¿S ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24083023 ↗
  • MalfunctionJan 9, 2026Detachment of Device or Device Component

    A CUSTOMER REPORTED THERE WAS A DETACHMENT OF THE CONTROL PANEL TO THE BASE OF THE EPIQ CVX ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THIS ISSUE WAS DISCOVERED OUTSIDE CLINICAL USE. THE PHILIPS FIELD SERVICE ENGINEER RE-INSTALLED THE SCREWS TO SECURE THE CONTROL PANEL TO THE SYSTEM TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24024428 ↗
  • MalfunctionDec 18, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE TRANSESOPHAGEAL X8-2T TRANSDUCER¿S SEALANT ON THE TIP WAS COMING OFF. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ CVX ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERN. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23846039 ↗
  • MalfunctionDec 18, 2025Unintended System Motion; Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS FIELD SERVICE ENGINEER DISASSEMBLED AND CLEANED THE SOLENOID TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23842096 ↗
  • MalfunctionDec 5, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE TRANSESOPHAGEAL X8-2T TRANSDUCER WAS LEAKING FLUID FROM THE SCAN TIP. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ CVX ULTRASOUND SYSTEM AT THE CUSTOMER'S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23727913 ↗
  • MalfunctionDec 3, 2025Unintended System Motion; Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS FIELD SERVICE ENGINEER REPLACED THE SOLENOID TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23705631 ↗