Skip to content

IntelliSpace Critical Care and Anesthesia IntelliSpace CC & Anesthesia JRecalls & reported problems

Philips Medical Systems Hsg · IntelliSpace Critical Care and Anesthesia · Class 2

All models of IntelliSpace Critical Care and Anesthesia (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838113763), received between Apr 13, 2026 and Apr 13, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Data Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Data Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2026Data Problem

    IT WAS REPORTED THAT ICCA LAB RESULTS HAD AN ISSUE REGARDING INBOUND INTERFACE INVOLVING THE HANDLING OF UPDATED SPECIMEN COLLECTION TIMESTAMPS SENT FROM THE LABORATORY INFORMATION SYSTEM (LIS). WHEN A BIOLOGIST MODIFIED THE SAMPLE DATE AND TIME (OBR 7), THE LIS GENERATED A NEW HL7 ORU^R01 MESSAGE USING THE SAME UNIQUE SPECIMEN IDENTIFIER (OBR 3.1) BUT WITH A CORRECTED TIMESTAMP IN OBR 7 OR OBX 14. ICCA DID NOT UPDATE THE PREVIOUSLY STORED SAMPLE DATE/TIME FOR THAT SPECIMEN; INSTEAD, IT UPDATED ONLY THE CHART TIME FIELD WHILE CONTINUING TO DISPLAY THE RESULT UNDER THE ORIGINAL CHART TIME COLUMN IN THE LAB FLOWSHEET. THIS BEHAVIOR PREVENTS THE RESULT FROM BEING REPOSITIONED ACCORDING TO THE CORRECTED TIMESTAMP AND MAY LEAD TO MISINTERPRETATION OF LABORATORY DATA TIMING, POSING A POTENTIAL PATIENT SAFETY RISK. HOWEVER, THERE WAS NO HARM TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24860704 ↗