Patient Information Center iX 4.2Recalls & reported problems
Philips Medizin Systeme Böblingen GmbH · Patient Information Center iX · Class 2
All models of Patient Information Center iX (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Philips Patient Information Center iX monitor is a patient bedside monitor with arrhythmia detection and alarms. Reports mostly describe alarm failures, including no audible alarm and alarms not displaying when triggered. There are 77 reports naming this device, with 6 deaths, 5 injuries and 66 malfunctions. There are 2 open recalls: one for a Care Assist mobile app custom ringtone issue and one for push notifications failing to send. Reports fell from 31 to 17 in the past 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 79
- Deaths reported
- 5
- Injuries reported
- 6
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 79 adverse event reports that list this exact device (DI 00884838121782), received between Dec 15, 2023 and Sep 30, 2026 — classified as 5 deaths, 6 injuries, 68 malfunctions.
The most frequently reported specific device problem is “No Audible Alarm” (31 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 28, 2024Class IIOpen, Classified
It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.
Z-1386-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 28, 2023Class IIOpen, Classified
Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.
Z-0731-2024 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- No Audible Alarm31
- Defective Alarm16
- Wireless Communication Problem14
- Communication or Transmission Problem6
- Device Alarm System6
- Alarm Not Visible1
- Delayed Alarm1
- Imprecise or Erratic Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions71
- Insufficient Information5
- Asystole1
- Cardiac Arrest1
- Heart Block1
- Low Oxygen Saturation1
- Tachycardia1
Recent adverse event reports
The 25 most recent of 79 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Defective Alarm
It was reported that the user requested assistance in retrieving logs following the device's failure to alarm for a cardiac arrest. No other clinical information or medical intervention was reported; therefore, good faith efforts are being performed.
Patient problems: Cardiac Arrest; Insufficient Information
MAUDE report 50943068 ↗MalfunctionSep 25, 2026Device Alarm System
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903741 ↗MalfunctionSep 25, 2026No Audible Alarm
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903596 ↗MalfunctionAug 28, 2026Imprecise or Erratic Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50497164 ↗MalfunctionJul 17, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25856743 ↗MalfunctionJun 9, 2026No Audible Alarm
PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX DEVICE INDICATING THAT THE DEVICE HAS NO AUDIBLE ALARMS. THE DEVICE WAS IN CLINICAL USE AT TIME OF EVENT, NO ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25442750 ↗MalfunctionJun 8, 2026No Audible Alarm
PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX INDICATING THAT THE THERE WERE NO ALARM SOUNDS FROM THE CONTROL PANEL. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25427394 ↗MalfunctionJun 1, 2026Defective Alarm
PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX INDICATING THAT THE PATIENT HAD ASYSTOLE AND ALARM WAS NOT DISPLAYED ON THE BED TO BED OVERVIEW. THE DEVICE WAS IN USE ON A PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25352284 ↗MalfunctionMay 12, 2026No Audible Alarm
PHILIPS RECEIVED A COMPLAINT ON A PATIENT INFORMATION CENTER IX (PIC IX) INDICATING THAT THE WARD REPORTED THAT ALARM SOUNDS WERE NOT AUDIBLE, EVEN THOUGH ALARMS WERE TRIGGERED. THE AUDIO VOLUME WAS CHECKED AND CONFIRMED NOT TO BE MUTED. THE VOLUME LEVEL ON THE CENTRAL STATION (PIC IX) WAS SET TO LEVEL THREE. IT WAS THEN ADJUSTED TO MAXIMUM, BUT NO SOUND WAS HEARD FROM THE SPEAKERS. NORMALLY, A SMALL TEST SOUND IS AUDIBLE WHEN ADJUSTING THE VOLUME, BUT NO SOUND OCCURRED. THE SPEAKER WAS REPLACED WITH A NEW ONE AND THE COMPUTER WAS RESTARTED. NORMALLY, A LOUD BEEP IS HEARD DURING SYSTEM STARTUP, BUT NO STARTUP SOUND WAS HEARD. THE SPEAKER WAS ALSO TESTED IN A DIFFERENT PORT ON THE COMPUTER, WITH NO IMPROVEMENT. A FAINT HISSING NOISE CAN BE HEARD FROM THE SPEAKER, BUT NO FUNCTIONAL AUDIO OUTPUT. THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25149839 ↗MalfunctionMay 7, 2026Incorrect Measurement
PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX INDICATING THE RHYTHM DISPLAYED IS NOT SYNCHRONIZED, SHOWING INCONSISTENT READINGS BETWEEN THE TWO MONITORS. THE TELEMETRY IS NOT SHOWING ACCURATE HR READINGS AND THAT THE VALUE KEEPS JUMPING AROUND FROM 60 TO 20 TO 30 BPM. THE DEVICE WAS IN USE AT THE TIME OF THE EVENT. NO ADVERSE EVENT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25112914 ↗MalfunctionApr 29, 2026Communication or Transmission Problem
IT WAS REPORTED THAT MULTIPLE PIC IX SURVEILLANCES ARE IN LOCAL MODE AND CANNOT CONNECT TO THE SERVER. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25027650 ↗MalfunctionApr 8, 2026No Audible Alarm
THE CUSTOMER REPORTED THAT NO AUDIBLE ALARMS AT THE ADMISSIONS CENTER IN THE EMERGENCY ROOM. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24823821 ↗MalfunctionMar 2, 2026Defective Alarm
PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX INDICATING CRITICAL ALARMS ARE NOT STAYING FLAGGED IN PATIENT SECTORS, AND ASYSTOLE ALARM DID NOT ALERT. THE DEVICE WAS IN USE AT THE TIME OF THE EVENT. NO ADVERSE EVENT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24484446 ↗MalfunctionFeb 19, 2026Communication or Transmission Problem
IT WAS REPORTED THAT RED ALARMS HAVE NOT BEEN RECEIVED ON THE LOGGED IN MOBILE DEVICES RUNNING CAREASSIST. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24390406 ↗DeathFeb 18, 2026Defective Alarm
IT WAS REPORTED AN ALARM FOR SPO2 WAS NOT GENERATED AT THE CENTRAL STATION AND THE PATIENT PASSED AWAY. THE CUSTOMER INDICATED THE ALARM WAS NOT FORWARDED TO THE CENTRAL STATION AND STATED 'IT HAPPENED AT 6 A.M., THE ALARM WAS NOT DISPLAYED, ONLY SOMETIME LATER. IT WAS NOTICED THAT THE PATIENT HAD DIED.' IT WAS INDICATED THERE IS A CENTRAL MONITORING UNIT, BUT NURSES ARE RESPONSIBLE FOR RESPONDING TO AND ACKNOWLEDGING ALARMS. THE CUSTOMER CONFIRMED THERE WAS A DELAY IN CARE DUE TO THE REPORTED ISSUE. PRELIMINARY REVIEW OF THE LOGS REVEALED THE ALARMS WERE GENERATED AND ACKNOWLEDGED. IN ADDITION, THE MONITOR REPEATEDLY GENERATED TECHNICAL ALARMS FOR THE SPO2 SENSOR IN THE FEW HOURS PRIOR TO THE EVENT. THE TECHNICAL ALARM WAS NOT RESOLVED, SO AN ADDITIONAL ALARM FOR PULSE OR SPO2 WAS NOT POSSIBLE.
Patient problems: Low Oxygen Saturation; No Clinical Signs, Symptoms or Conditions
MAUDE report 24379968 ↗MalfunctionFeb 16, 2026No Audible Alarm
PHILIPS RECEIVED A COMPLAINT ON A PATIENT INFORMATION CENTER IX INDICATING THAT SOME ALARMS ARE INTERMITTENTLY NOT PRODUCING SOUND. THE DEVICE WAS IN USE ON A PATIENT AT TIME OF EVENT; THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24344704 ↗MalfunctionJan 13, 2026No Audible Alarm
IT WAS REPORTED THAT ONLY VISUAL ALARMS WERE AVAILABLE, THE DEVICE IS NOT PRODUCING AUDIO SOUND. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24057001 ↗MalfunctionDec 10, 2025No Audible Alarm; No Audible Alarm
PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX (PIC IX) INDICATING THAT THE CENTRAL STATION WAS NOT EMITTING ANY SOUND. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770963 ↗MalfunctionDec 3, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED MULTIPLE CENTRAL STATION HOSTS WERE IN LOCAL MODE. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23708441 ↗MalfunctionSep 8, 2025No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT THE NURSES ARE UNABLE TO HEAR THE ALARMS ON THE PIC IX WHEN THEY GO OFF. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22991258 ↗MalfunctionSep 2, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THE DEVICE IS NOT GETTING ALARMS OUT OF ONE OF THE ROOMS FROM CENTRAL MONITORING UNIT. THE DEVICE WAS IN USE ON A PATIENT AT TIME OF EVENT; THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22936110 ↗MalfunctionAug 8, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THE ALARMS ARE NOT COMING THROUGH. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22751080 ↗DeathAug 7, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THE DEVICE DID NOT GENERATE AN ALARM FOR A DESATURATION AND THE PATIENT PASSED AWAY. IT WAS INDICATED THE ALARM MAY HAVE BEEN MISSED OR NOT HEARD BY THE STAFF AND THE PATIENT PASSED WAY AT APPROXIMATELY 9:42AM. BIOMED PERFORMED SIMULATED TESTING ON THE MX40 TELEMETRY DEVICE, CONFIRMING THE ALARM FUNCTIONED AND APPEARED ON BOTH THE MX40 AND THE PIC IX CENTRAL STATION. BIOMED ALSO REVIEWED THE CLINICAL AUDIT LOGS, REVEALING A DESATURATION ALARM WAS GENERATED AT 9:20AM FOR AN OXYGEN SATURATION OF 75% AND AT 9:40AM FOR AN OXYGEN SATURATION OF 74%. THE PATIENT DEATH IS BEING ADDRESSED IN RELATED MFR REPORT NUMBER 1218950-2025-000417.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22742088 ↗MalfunctionAug 7, 2025Alarm Not Visible; Alarm Not Visible
IT WAS REPORTED THAT THE ALARM NOTIFICATION IS NOT APPEARING IN THE OVERVIEW SURVEILLANCE AND IN OTHER ROOMS. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22735998 ↗MalfunctionJul 1, 2025No Audible Alarm; No Audible Alarm
THE CUSTOMER REPORTED THIS IS A BACKUP SPARE UNIT THEY WANTED TO USE; HOWEVER, THE PATIENT INFORMATION CENTER IX DID NOT HAVE SOUND. THEY PERFORMED TROUBLESHOOTING AND WHEN THEY LOOSENED THE LID, THEY IDENTIFIED THERE IS NO CABLE FROM THE SOUND CARD AND THAT IT A SCREW IS MISSING. THE DEVICE WAS REPORTED TO BE IN CLINICAL USE. NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22370039 ↗
Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page