Patient Information Center iX 4.7Recalls & reported problems

Philips Medizin Systeme Böblingen GmbH · Patient Information Center iX · Class 2

All models of Patient Information Center iX (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00884838128897), received between May 3, 2026 and Sep 23, 2026 — classified as 1 injury, 13 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 4 reportsJul 2026: 3 reportsAug 2026: 4 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem5
  • No Audible Alarm4
  • Defective Alarm2
  • Device Alarm System2
  • Installation-Related Problem1
  • Missing Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Bradycardia1
  • Choking1
  • Low Oxygen Saturation1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873064 ↗
  • MalfunctionSep 11, 2026Device Alarm System

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50726158 ↗
  • MalfunctionAug 14, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50246630 ↗
  • MalfunctionAug 11, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50213873 ↗
  • MalfunctionAug 3, 2026Missing Test Results; Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50124483 ↗
  • MalfunctionAug 3, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50113418 ↗
  • MalfunctionJul 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50090560 ↗
  • MalfunctionJul 9, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25764882 ↗
  • MalfunctionJul 8, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25751383 ↗
  • InjuryJun 30, 2026Installation-Related Problem

    PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX INDICATING THAT THE STAFF EXPECTED ALARMS TO POP UP IN ALL UNIT ROOMS, BUT THE BEHAVIOR HAD CHANGED AFTER THE SOFTWARE UPGRADE. IT WAS DOCUMENTED THAT THE BEDSIDE MONITOR AND CENTRAL STATION ALARMED AS EXPECTED; HOWEVER, THE STAFF WAS EXPECTING POP UP ALARMS THAT DID NOT OCCUR AS THE NEW UPGRADE DOES NOT HAVE THIS FUNCTIONALITY. THE NURSE WAS WITH ANOTHER PATIENT AT THE TIME OF THE EVENT. THE EVENT STARTED AT 0259 AND LASTED UNTIL 0303 WHEN A STAFF ASSIST WAS CALLED. THE INFANT WAS CHOKING ON SPIT UP AND THE MONITOR STARTED ALARMING FOR DESATURATION AT 0259 AND LASTED FOR APPROXIMATELY ONE MINUTE BEFORE ALARMS WOKE THE MOTHER, WHO TRIED TO RECOVER THE INFANT ON HER OWN BY SUCTIONING WHILE WAITING FOR THE NURSE TO RESPOND. WHEN THE NURSE DID NOT RESPOND, THE MOTHER CALLED OUT FOR ASSISTANCE AND TWO PEDIATRIC NURSES IN A NEARBY UNIT RESPONDED. THE FOLLOWING IS THE REST OF THE EVENT DESCRIPTION FROM THE CUSTOMER, "THE PEDIATRIC NURSES INITIATED THE STAFF ASSIST ALARM BRINGING THE NICU CHARGE NURSE AND NEONATAL NURSE PRACTITIONER TO BEDSIDE AT 0303. THE INFANT WAS FOUND TO BE DUSKY AND APNEIC WITH A LARGE AMOUNT OF EMESI

    Patient problems: Bradycardia; Choking; Low Oxygen Saturation

    MAUDE report 25663855 ↗
  • MalfunctionJun 26, 2026Device Alarm System

    PHILIPS RECEIVED A COMPLAINT ON A PATIENT INFORMATION CENTER IX INDICATING THAT AFTER AN UPDATE WAS COMPLETED, THE BEDS ARE NOT ALARMING ON THE OTHER MONITORS. IT IS UNKNOWN IF THE DEVICE WAS IN USE AT TIME OF EVENT, AND THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25639211 ↗
  • MalfunctionJun 17, 2026No Audible Alarm

    IT WAS REPORTED THAT ON (B)(6) 2026 IN THE TIME SLOT 06:00-06:46 THE ALARMS, IN PARTICULAR THE RED ALARMS, DID NOT SOUND. THE DEVICE WAS IN CLINICAL USE. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25531205 ↗
  • MalfunctionJun 16, 2026No Audible Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX (PIC IX) INDICATING THAT USER WAS UNABLE TO HEAR ANY ALARMS. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25516109 ↗
  • MalfunctionMay 3, 2026Defective Alarm

    PHILIPS RECEIVED A COMPLAINT ON THE PATIENT INFORMATION CENTER IX INDICATING THAT THE DEVICE FAILED TO ALARM FOR A V TACH IN ROOM (B)(6) ON (B)(6) 2026 AT APPROXIMATELY 10:50 AM AND 10:52 AM. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25064008 ↗

Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices

All Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices →