Vysis 06N38-033Recalls & reported problems

Abbott Molecular Inc. · Vysis ALK Break Apart FISH Probe Kit · Class 2

All models of Vysis (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884999042728), received between Aug 27, 2024 and Aug 27, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “False Negative Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2024False Negative Result; False Negative Result

    THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT ON THE VYSIS ALK BREAK APART FISH. THE CUSTOMER TESTED A SAMPLE AND FOUND THAT IMMUNOHISTOCHEMISTRY (IHC) RESULTS AND PCR RESULTS WERE 90% POSITIVE, BUT THE FISH RESULTS WERE LESS THAN 5%. THE VYSIS ALK BREAK APART FISH'S MANUFACTURE DATE WAS MAY 2023 AND EXPIRATION DATE WAS APRIL 2024. THE KIT WAS EXPIRED AT THE TIME OF THE REPORTED EVENT. THE CUSTOMER REPORTED THAT THEY CAN SEE ALL FUSION SIGNAL AND DID NOT BELIEVE THE EXPIRED KIT WOULD CAUSE DIFFERENT RESULTS BETWEEN THE IHC AND PCR. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT. THE NEGATIVE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20076816 ↗