CD HORIZON® Spinal System 7577545Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 7577545 LEGACY 5.5 CANN MAS 7.5X45 · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00885074032047), received between Oct 9, 2018 and Mar 30, 2026 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Back pain1
- Failure of Implant1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Detachment of Device or Device Component
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO UNDERWENT LUMBAR FUSION FOR DEGENERATIVE DISC DISEASE. IT WAS REPORTED THAT WHEN THE PATIENT RETURNED FOR A FOLLOW-UP VISIT SEVEN MONTHS POST-OP, THE PHYSICIAN DISCOVERED THAT THE HEAD AND SHANK OF OUR SCREW HAD SEPARATED/DISLODGED. PATIENT WAS STILL IN BACK PAIN. THE DOCTOR'S CUR RENT EVALUATION IS TO KEEP THE PATIENT UNDER OBSERVATION. THERE IS NO NEED TO SCHEDULE A SECOND SURGERY YET. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: Back pain
MAUDE report 24713437 ↗MalfunctionNov 28, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING TRANSFORAMINAL LUMBAR INTERBODY FUSION. IT WAS REPORTED THAT THE SCREW HEAD WAS BROKEN FROM THE SHAFT. THERE WAS NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23670342 ↗InjuryJun 16, 2021Break; Break
INFORMATION WAS RECEIVED FROM A HEALTH CARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A SCREW AND ROD IMPLANTED IN A PATIENT. PROCEDURE PERFORMED WAS MIS TLIF WITH PERCUTANEOUS SCREW FIXATION. PRE-OP DIAGNOSIS WAS LUMBAR STENOSIS WITH SPONDYLOLISTHESIS. IT WAS REPORTED THAT POST-OPERATIVELY, SURGEON FOUND THE SCREW HEAD WAS SEPARATED WITH SCREW DURING THE CASE FOLLOW UP. THE C RING IN THE BOTTOM OF THE SCREW HEAD WAS BROKE. THERE WAS NO FRAGMENT OF THE IMPLANT OR INSTRUMENT REMAINING IN THE PATIENT. THE ROD WAS INTACT. SURGEON REMOVED THE LOOSENING SCREW AND ROD AND IMPLANTED NEW SCREW AND ROD. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12010369 ↗InjuryMay 20, 2020Break; Break
IT WAS REPORTED THAT ABOUT 3 YEARS BACK, THE PATIENT UNDERWENT MINIMAL INVASIVE TRANSFORAMINAL LUMBAR INTERBODY FUSION (MIS-TLIF) AT L5-S1. CAGE (FROM VELOFIX VENDOR), SCREWS AND RODS WERE IMPLANTED IN THE PATIENT DURING THIS SURGERY. ON AN UNKNOWN DATE, A YEAR AND 4 MONTHS POST-OP, THE PATIENT PRESENTED WITH NON-UNION. REPORTEDLY, THE PATIENT STARTED TO HEAR NOISES (CLICKS) IN HIS BACK AND THE PATIENT HAD A LOT OF PAIN IN HIS BACK. REPORTEDLY, THERE WAS FUSION AT ONLY ONE SIDE; AND LIKELY, THE CAGE WAS SMALL DUE TO WHICH THERE WAS A LITTLE GAP BETWEEN VERTEBRAE. HENCE, THE PATIENT UNDERWENT AN ANTERIOR LUMBAR INTERBODY FUSION (ALIF) TO REPLACE THE CAGE IMPLANTED IN THE PREVIOUS SURGERY. IN THIS SURGERY, THE CAGE WAS REPLACED WITH A NEWER AND A BIGGER ONE, IN WHICH MORE AMOUNT OF BONE MATERIAL WAS PLACED. THE PATIENT STARTED TO FEEL BETTER AFTER ALIF, BUT THERE WAS STILL PAIN IN THE PATIENT'S BACK. THE PATIENT STILL HEARD CLICK NOISES IN HIS BACK. SO, A YEAR LATER, ANOTHER SURGERY WAS PERFORMED, IN WHICH REMOVAL OF SCREWS AND RODS WAS PERFORMED. IT WAS FOUND THAT THE AREA WHERE THE PATIENT WAS EXPERIENCING PAIN, WAS THE SAME PLACE AT WHICH THE SCREW BROKE. THE BROKEN SCREW COULD NO
Patient problems: Failure of Implant; Pain
MAUDE report 10075301 ↗InjuryMay 17, 2020Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) AT L4/5, L5/S1 DUE TO STENOSIS. THREE MONTHS POST-OP, THE IMPLANTED SCREW WAS BROKEN. AS A RESULT OF THIS EVENT PATIENT UNDERWENT REVISION SURGERY. THE PATIENT WENT TO ANOTHER HOSPITAL FOR REMOVAL OF THE SCREW, WHERE SCREW WAS REPLACED WITH ANOTHER MANUFACTURER SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10064044 ↗MalfunctionOct 9, 2018Break; Break
IT WAS REPORTED THAT INTRA-OP, A SCREW BROKE. IT WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7946172 ↗MalfunctionOct 9, 2018Break; Break
IT WAS REPORTED THAT INTRA-OP, A SCREW BROKE. IT WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7946169 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page