SPRINTER® SPR1515WRecalls & reported problems
Medtronic, Inc. · BALLOON SPR1515W SPRINTER OTW 1.5X15MM · Class 2
All models of Sprinter (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885074197029), received between Oct 19, 2017 and Oct 19, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
Patient problems
- Device Embedded In Tissue or Plaque1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 19, 2017Burst Container or Vessel; Detachment of Device or Device Component; Burst Container or Vessel; Detachment of Device or Device Component
IT WAS INTENDED TO USE A SPRINTER OTW BALLOON CATHETER TO TREAT A SEVERELY TORTUOUS, SEVERELY CALCIFIED DISTAL LESION EXHIBITING 95% STENOSIS IN A PERIPHERAL ARTERY (BELOW THE KNEE NEAR THE FOOT). THERE WAS NO DAMAGE NOTED TO PACKAGING OR ISSUES REMOVING THE DEVICE FROM THE HOOP/TRAY. NO STYLET REMOVAL DIFFICULTIES. THE DEVICE WAS INSPECTED AND PREPPED WITH NO ISSUES NOTED. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE HOWEVER EXCESSIVE FORCE WAS NOT USED. THE DEVICE WAS NOT KINKED AND RE-STRAIGHTENED DURING USE. IT IS REPORTED THAT UPON INFLATION, THE BALLOON BURST AT 10 ATM. THERE WAS A SUDDEN DROP IN PRESSURE FROM INFLATION DEVICE. THE BALLOON WAS NOT REPOSITIONED BEFORE BURST. THE BALLOON THEN SEPARATED FROM THE SHAFT WITH THE SINGLE MARKER INTACT AND WHEN THE PHYSICIAN REMOVED THE SYSTEM FROM THE PATIENT THE DETACHED PORTION REMAINED IN THE VERY DISTAL TIBIAL ARTERY. AN ATTEMPT WAS MADE TO SNARE WITH A LOOP SNARE HOWEVER THE SNARE COULD NOT GET ALL THE WAY DOWN TO THE FOOT. A SURGICAL CONSULT DECIDED THAT IT WAS NOT AN OPTION TO CUT DOWN AND REMOVE THE BALLOON. THE DETACHED PORT
Patient problems: Device Embedded In Tissue or Plaque; Patient Problem/Medical Problem
MAUDE report 6960660 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page