Skip to content

N/A 2982040Recalls & reported problems

Medtronic Sofamor Danek, Inc. · TRIAL 2982040 10 X 40MM L DISTRACTOR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885074549750), received between Jun 20, 2020 and Jun 20, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult to Advance1
  • Use of Device Problem1

Patient problems

  • Blood Loss1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 20, 2020Use of Device Problem; Difficult to Advance; Use of Device Problem; Difficult to Advance

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH LUMBAR SPINAL CANAL STENOSIS; AND UNDERWENT OBLIQUE LUMBAR INTERBODY FUSION AT L2-L5. BEFORE TRIAL INSERTION, THE DISTRACTOR HAD BEEN SET TO 12 MM, AND THE TRIAL WAS STARTED FROM HEIGHT OF 10 MM. INTRA-OP, WHILE ATTEMPTING TO INSERT THE TRIAL AT L4-L5 AT THE FIRST LEVEL, THE TRIAL PROTRUDED ABOUT 20 MM FROM THE CONTRALATERAL SIDE AS A RESULT OF STRONG HAMMERING. THE DIRECTION REMAINED DIAGONAL BEFORE RAISING VERTICALLY. STRONG HAMMERING WAS PERFORMED AS THE ENTRANCE OF THE INTERVERTEBRAL DISC SPACE WAS NARROW. AS A RESULT OF TRIAL PROTRUSION INTO THE VERTEBRAL BODY, THE PATIENT HAD A LOT OF BLEEDING (AMOUNT UNKNOWN) AND HEMOSTASIS WAS PERFORMED. THEN, THE WOUND WAS CLOSED WITHOUT INSERTING THE CAGE. ORIGINALLY, IT WAS PLANNED TO PERFORM A SECOND STAGE FIXATION BUT IT COULD NOT BE PERFORMED THEN. THERE WAS A DELAY OF MORE THAN 60 MINUTES IN OVERALL PROCEDURE AS A RESULT OF THIS EVENT. THE SURGEON FEELS THAT IT MAY HAVE BEEN NECESSARY TO START THE TRIAL FROM A SMALLER SIZE. ALTHOUGH THE EXACT PLACE OF BLEEDING WAS UNKNOWN, THE SURGEON SUSPECTED THE BLEEDING TO BE FROM VENOUS PLEXUS NEAR THE CONTRALATERAL ROUTE. THE SURGEON ALSO BELIEVES

    Patient problems: Blood Loss

    MAUDE report 10177132 ↗

Other SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices

All SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices →

Sources and updates