Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885380028710), received between May 10, 2023 and May 10, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Suction Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Suction Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 10, 2023Suction Failure; Suction Failure
THE CUSTOMER REPORTED THAT ON (B)(6) 2023 THE PATIENT BECAME LESS RESPONSIVE. RRT WAS CALLED AND SUCTION SET UP WAS COMPLETED AND ACCURATE, BUT NO SUCTION WAS ACHIEVED THROUGH THE YANKAUER. THE REGULATOR WAS SWITCHED BUT AGAIN NO SUCTION. CRASH CART SUCTION WAS THEN SET UP BY NURSING WITH THE SAME RESULT. RT NOW PRESENT AND ADVISES TO TRY A NEW YANKAUER WHICH WAS EFFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16901760 ↗
Other APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices
- StrykeFlow 0250070520Stryker Corporation · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED2 recalls250 reports
- StrykeFlow 0250070505Stryker Corporation · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall3 reports
- Sentinel Seal 8888571562Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Aqua-Seal 8888571299Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- spineX BONE DUST COLLECTOR S-4500Oto-Med, Inc. · Apparatus, Suction, Operating-Room, Wall Vacuum Powered1 recall0 reports
- Aqua-Seal 8888571406Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Altitude 8888571370Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- CRD 65651-517Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page