R3 ACETABULAR 71338952Recalls & reported problems
Smith & Nephew, Inc. · R3 36MM ID US CERAMIC ACET LINER 52MM · Class 3 · Implant
All models of R3 ACETABULAR (24)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00885556021552), received between Sep 21, 2018 and May 22, 2023 — classified as 7 injuries.
The most frequently reported specific device problem is “Fracture” (4 reports).
6 other recalls cover devices cleared under the same FDA submission (P030022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture4
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
- Noise, Audible1
Patient problems
- Failure of Implant3
- Pain3
- Ambulation or Postural Difficulties1
- Arthralgia1
- Cellulitis1
- Extraskeletal ossification1
- Injury1
- No Code Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 22, 2023Fracture; Fracture
IT WAS REPORTED THAT, AFTER A THR HAD BEEN PERFORMED ON (B)(6) 2010, THE PATIENT SUFFERED A FALL RESULTING IN PAIN IN THE AFFECTED HIP. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2023 TO ADDRESS THIS EVENT. IN ADDITION, THERE DID NOT APPEAR TO BE ANY DAMAGE TO THE FORTE FEMORAL HEAD, HOWEVER THE R3 36MM ID US CRMC LINER 52MM WAS FRACTURED. THE COMPONENTS WERE REPLACED WITH A R3 20 DEGREE XLPE LINER AND A ZIMMER +7MM SLEEVED 36 DELTA HEAD. PATIENT WAS SCHEDULED TO BE DISCHARGED DAY OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 16975971 ↗InjuryMar 25, 2021Fracture; Fracture
US LEGAL. IT WAS REPORTED THAT A PATIENT UNDERWENT MEDICALLY INDICATED REVISION OF THE R3 SYSTEM LEFT HIP ON (B)(6) 2020. THE PATIENT REVISION SURGERY WAS PERFORMED DUE TO A FRACTURED R3 CERAMIC LINER. THE PATIENT OUTCOME IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 11568083 ↗InjuryJan 29, 2021Break; Fracture; Break; Fracture
IT WAS REPORTED THAT, PATIENT UNDERWENT BILATERAL THR ((B)(6) 2011). ON (B)(6) 2020 A REVISION SURGERY DUE TO PAIN IN THE LEFT HIP AFTER PLAYING GOLF WAS PERFORMED. A CT SCAN SHOWED A FRACTURE OF THE R3 CERAMIC LINER. THE FOLLOWING DEVICES WHERE EXPLANTED: R3 CERAMIC LINER, R3¿ NO HOLE ACETABULAR SHELL, SYNERGY POROUS PLUS STEM , UNKNOWN FEMORAL HEAD INSERT AND R3¿/REFLECTION¿ THREADED HOLE COVER. THE PROCEDURE WAS COMPLETED WITH SMITH & NEPHEW BACK-UP DEVICES. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. AFTER SURGERY AND ONCE PATIENT DISCHARGED HOME ((B)(6) 2020), IN DECEMBER THE PATIENT DEVELOPED AN AREA OF CELLULITIS AT THE DISTAL INCISION, TREATED WITH ANTIBIOTICS. AS OF ((B)(6) 2021) OFFICE VISIT, THERE WAS IMPROVEMENT IN THE CELLULITIS; THERE WAS NO OBVIOUS FLUID COLLECTION AND THE SKIN & SUBCUTANEOUS TISSUES FELT NORMAL. NO OTHER COMPLICATIONS WHERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Ambulation or Postural Difficulties; Cellulitis; Extraskeletal ossification; Failure of Implant; Pain
MAUDE report 11250345 ↗InjuryJul 16, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER A THA HAD BEEN PERFORMED, THE PATIENT EXPERIENCED PAIN. THE ADVERSE EVENT WAS ADDRESSED BY PERFORMING A REVISION SURGERY. THE PATIENT OUTCOME IS UNKNOWN.
Patient problems: Arthralgia
MAUDE report 10284229 ↗InjuryJul 14, 2019Fracture; Fracture
IT WAS REPORTED THAT THE DOCTOR REVISED A SHATTERED AN R3 CERAMIC LINER. THE HEAD AND THE LINER WERE REMOVED. A POLY LINER WAS IMPLANTED. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Injury
MAUDE report 8787667 ↗InjuryOct 2, 2018Break; Noise, Audible
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO THE CERAMIC LINER WAS SQUEAKING WHEN THE PATIENT WALKED. DURING REMOVAL THE DOCTOR REPORTED THAT THE LINER WAS BROKEN.
Patient problems: No Code Available
MAUDE report 7927972 ↗InjurySep 21, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT REVISION SURGERY WAS PERFORMED. SURGEON REVISED R3 ALUMINA ON ALUMINA. PATIENT WAS COMPLAINING OF PAIN, THOUGHT CERAMIC LINER MIGHT BE FRACTURED. IT WAS FOUND THAT THE CERAMIC LINER WAS NOT FRACTURED, IMPLANTS WERE FINE. SURGEON CHANGED LINER TO A POLY LINER AND A NEW FEMORAL HEAD WAS PUT ON. NO ISSUES, NO DELAYS, NO INJURIES.
Patient problems: Pain
MAUDE report 7899528 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030022. One submission often covers many models, so check the product and lot details in each record.
- Z-2058-2011 · Mar 11, 2011 · TerminatedR3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 62 MM O.D., REF 71338962, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
- Z-2054-2011 · Mar 11, 2011 · TerminatedR3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 32 MM O.D. 55 MM O.D., REF 71338950, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
- Z-2057-2011 · Mar 11, 2011 · TerminatedR3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 56 MM O.D., REF 71338956, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
- Z-2055-2011 · Mar 11, 2011 · TerminatedR3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 52 MM O.D., REF 71338952, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
- Z-2053-2011 · Mar 11, 2011 · TerminatedR3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 32 MM I.D. 48 MM O.D., REF 71338948, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06 Usage: Primary total
- Z-2056-2011 · Mar 11, 2011 · TerminatedR3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 54 MM O.D., REF 71338954, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.