Skip to content

R3 71339162Recalls & reported problems

Smith & Nephew, Inc. · R3 CONSTRAINED ACETABULAR LINER 62MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556021668), received between Aug 31, 2017 and Oct 17, 2023 — classified as 3 injuries.

The most frequently reported specific device problem is “Break” (1 report).

14 other recalls cover devices cleared under the same FDA submission (K083566, K111635, K122139, K211176); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Insufficient Device Problem Information1

Patient problems

  • Failure of Implant1
  • Fall1
  • Injury1
  • Joint Dislocation1
  • Medical device site infection1
  • Osteolysis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A THR REVISION PERFORMED ON (B)(6) 2015, DUE TO A HISTORY OF PROSTHETIC HIP DISLOCATIONS, THE PATIENT BEGAN TO DEVELOP CHRONIC PAIN. AFTER A THOROUGH WORKUP THEY WERE DIAGNOSED WITH A PROSTHETIC JOINT BORDETELLA HOLMESII INFECTION, FOR WHICH A TWO-STAGE REVISION WAS CARRIED OUT. STAGE ONE OF THE REVISION WAS PERFORMED ON (B)(6) 2021; WHEN THE PATIENT¿S HIP WAS ASPIRATED GROSS PURULENT FLUID WAS NOTED, THE POSTERIOR CAPSULE WAS EXTREMELY INFLAMED, AND FOCAL OSTEOLYSIS WAS FOUND IN THE ACETABULUM. PATIENT HAD A SYNOVECTOMY, WAS IRRIGATED WITH ANTIBIOTICS, HAD R3 AND SYNERGY HAD COMPONENTS EXPLANTED, AND A DE PUY PROSTALAC SPACER IMPLANTED. SECOND STAGE REVISION WAS PERFORMED ON (B)(6) 2021, DURING WHICH NO EVIDENCE OF INFECTION WAS FOUND, AND A STRYKER REVISION SYSTEM WAS IMPLANTED. THE PATIENT WAS TAKEN TO THE RECOVERY ROOM IN STABLE CONDITION.

    Patient problems: Medical device site infection; Osteolysis

    MAUDE report 17951017 ↗
  • InjuryApr 29, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT WENT THROUGH PRIMARY THR IN (B)(6) 2020. LAST WEEK THE PATIENT FELL IN THE BATHROOM AND HAD PAIN IN HIS HIP, THEREFORE, A REVISION SURGERY WAS PERFORMED. INTRAOPERATIVELY, THE LINER WAS NOTICED TO LOOKS TILTED AND THE INNER LINER RING WAS FOUND BROKEN. THE R3 INSERT CONSTRAINED 62 MM HAD TO BE DISASSEMBLED DUE TO THE COMPLEX REVISION SURGERY TO ALLOW DISLOCATION OF THE HIP. IT WAS NOT POSSIBLE TO REINSERT A NEW 62MM CONSTRAINED INLAY, SO AN ANTEVERTED INLAY WAS INSERTED. THE OUTCOME OF THE PATIENT IS UNKNOWN.

    Patient problems: Failure of Implant; Fall

    MAUDE report 11745721 ↗
  • InjuryAug 31, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    PATIENT PRESENTED TO EMERGENCY DEPARTMENT WITH DISLOCATED HIP. EMERGENCY DEPARTMENT RELOCATED HIP UNDER SEDATION. NO OPERATION NEEDED.

    Patient problems: Injury; Joint Dislocation

    MAUDE report 6833815 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083566, K111635, K122139, K211176. One submission often covers many models, so check the product and lot details in each record.

  • Z-0326-2025 · Oct 15, 2024 · Open, Classified
    Tandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
  • Z-1737-2015 · May 6, 2015 · Terminated
    R3(TM) 22 MM I.D., 58 MM O.D. CONSTRAINED LINER, REF 71339158, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthes
  • Z-1736-2015 · May 6, 2015 · Terminated
    R3(TM) 22 MM I.D., 56 MM O.D. CONSTRAINED LINER, REF 71339156, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthes
  • Z-1735-2015 · May 6, 2015 · Terminated
    R3(TM) 22 MM I.D., 52 MM O.D. CONSTRAINED LINER, REF 71339152, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthes
  • Z-1738-2015 · May 6, 2015 · Terminated
    R3(TM) 28 MM I.D., 60 MM O.D. CONSTRAINED LINER, REF 71339160, QTY: (1), STERILE EO, Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthe
  • Z-1739-2015 · May 6, 2015 · Terminated
    R3(TM) 28 MM I.D., 62 MM O.D. CONSTRAINED LINER, REF 71339162, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthes
  • Z-0083-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 28 MM I.D., 60 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339160, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabular
  • Z-0082-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 22 MM I.D., 58 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339158, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabular
  • Z-0079-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 22 MM I.D., 52 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339152, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabular
  • Z-0086-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 28 MM I.D., 66/70 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339166, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabul
  • Z-0081-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 22 MM I.D., 56 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339156, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabular
  • Z-0080-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 22 MM I.D., 54 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339154, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabular
  • Z-0084-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 28 MM I.D., 62 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339162, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabular
  • Z-0085-2011 · Aug 13, 2010 · Terminated
    R3 (TM) 28 MM I.D., 64 MM O.D., CONSTRAINED LINER, UHMWPE, CO-CR, TI-6AL-4V, REF 71339164, STERILE EO, QTY 1, Smith & Nephew, Inc., Hip prosthesis acetabular component. Acetabular

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →