Skip to content

LEGION 71423356Recalls & reported problems

Smith & Nephew, Inc. · LEGION HINGE GUIDED MOTION TIBIAL INSERT 15MM SIZE 4-5 LEFT · Class 2 · Implant

All models of LEGION (446)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885556030271), received between Apr 17, 2017 and Oct 17, 2025 — classified as 6 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

7 other recalls cover devices cleared under the same FDA submission (K081111); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Insufficient Device Problem Information2
  • Detachment of Device or Device Component1
  • Malposition of Device1
  • Migration1
  • Migration or Expulsion of Device1

Patient problems

  • Failure of Implant2
  • Injury2
  • Unspecified Infection2
  • Discomfort1
  • Foreign Body In Patient1
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 17, 2025Migration; Migration

    IT WAS REPORTED THAT, AFTER A TKA SURGERY HAD BEEN PERFORMED ON AN UNKNOWN DATE, THE SCREW COMPONENT OF THE POLY WAS BACKING OUT. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2025, IN WHICH ONE (1) LEGION HK GD MOTION ISRT 15MM SZ 4-5 LT WAS REMOVED BUT UNFORTUNATELY THE SCREW WAS BEHIND THE KNEE OF THE PATIENT AND IT WAS LEFT IMPLANTED. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 23320056 ↗
  • InjuryJun 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A TKA SURGERY PERFORMED ON (B)(6) 2021, THE PATIENT SUSTAINED AN UNSPECIFIED INFECTION. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2023 TO EXCHANGE THE LEGION HK GD MOTION ISRT 15MM SZ 4-5 LT AND LEGION HK 15MM BOLT - SLEEVE TO ADDRESS THIS COMPLICATION. THE I & D WAS DONE TO FIGHT OFF INFECTION. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 17162139 ↗
  • InjuryJun 14, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A SURGICAL INTERVENTION WILL BE PERFORMED DUE TO KNEE INSTABILITY.

    Patient problems: Injury

    MAUDE report 7603976 ↗
  • InjuryJun 12, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO JOINT INSTABILITY.

    Patient problems: Injury

    MAUDE report 7590256 ↗
  • InjuryJul 27, 2017Detachment of Device or Device Component; Migration or Expulsion of Device; Malposition of Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT PATIENT UNDERWENT A REVISION SURGERY TO REPLACE THE ARTICULAR INSERT OF TOTAL KNEE SYSTEM. DURING CHECK UP THE DOCTOR NOTICED THE INSERT SCREW OF THE HINGE INSERT WASN'T IN ITS ORIGINAL PLACE. BECAUSE OF PAIN, LAXITY AND UNCOMFORTED FEELING THE DOCTOR DECIDED TO REVISE THE INSERT OF THE LEGION HINGE.

    Patient problems: Discomfort; Failure of Implant; Pain

    MAUDE report 6746286 ↗
  • InjuryApr 17, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED AN INFECTED HINGE WAS REVISED FOR WASHOUT AND DEBRIDEMENT, POLY CHANGE WAS DONE AS A PRECAUTION.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 6498064 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081111. One submission often covers many models, so check the product and lot details in each record.

  • Z-1105-2018 · Feb 1, 2018 · Terminated
    smith&nephew, LEGION HK FEMORAL WEDGE (WEDGE AND SCREW), 5 MM, SIZE 5, REF 71422159, QTY: (1), STERILE The Revision Knee System Components are indicated for rheumatoid arthritis;
  • Z-2094-2016 · May 26, 2016 · Terminated
    LEGION(TM) HK FEMORAL ASSEMBLY, SIZE 5 LEFT, A/P 62 MM, M/L 70 MM, REF 71421375 The LEGION Hinge Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthritis
  • Z-2090-2016 · May 26, 2016 · Terminated
    LEGION(TM) HK FEMORAL ASSEMBLY, SIZE 3 LEFT, A/P 54.5 MM, M/L 62 MM, REF 71421373 The LEGION Hinge Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthrit
  • Z-2091-2016 · May 26, 2016 · Terminated
    LEGION(TM) HK FEMORAL ASSEMBLY, SIZE 3 RIGHT, A/P 54.5 MM, M/L 62 MM, REF 71421363 The LEGION Hinge Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthri
  • Z-2093-2016 · May 26, 2016 · Terminated
    LEGION(TM) HK FEMORAL ASSEMBLY, SIZE 4 RIGHT, A/P 58.5 MM, M/L 66 MM, REF 71421364 The LEGION Hinge Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthri
  • Z-2092-2016 · May 26, 2016 · Terminated
    LEGION(TM) HK FEMORAL ASSEMBLY, SIZE 4 LEFT, A/P 58.5 MM, M/L 66 MM, REF 71421374 The LEGION Hinge Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthrit
  • Z-2095-2016 · May 26, 2016 · Terminated
    LEGION(TM) HK FEMORAL ASSEMBLY, SIZE 5 RIGHT, A/P 62 MM, M/L 70 MM, REF 71421365 The LEGION Hinge Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthriti

Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page