GENESIS II 71433346Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II SYMMETRICAL POSTERIOR CONDYLES REAM THROUGH CRUCIATE RETAINING TRIAL SIZE 6 LEFT · Class 2
All models of GENESIS II (462)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00885556032107), received between Jun 2, 2020 and Sep 5, 2023 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Crack” (5 reports).
4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Crack5
- Break3
- Difficult to Remove1
- Insufficient Device Problem Information1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement2
- Cartilage damage1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjurySep 5, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT, DURING A ROUTINE PRIMARY LEGION TKA, AFTER TRIALING WITH A GENESIS II SYMMETRICAL POSTERIOR CONDYLES REAM THROUGH CRUCIATE RETAINING TRIAL SIZE 6 LEFT AND TAKING THE KNEE AN EXTENSION-FLEXION RANGE THERE APPEARED TO BE A PIECE OF CARTILAGE TRAPPED IN THE LATERAL HOLE OF THE FEMORAL TRIAL COMPONENT THAT IS UTILIZED TO PREPARE CR LUNGS. ON INSPECTION IT WAS NOTED THAT THE PREVIOUSLY VERY HEALTHY PATELLA HAD A DEFECT IN IT THAT MATCHED THIS PIECE OF CARTILAGE. FURTHERMORE, THE SURGEON BELIEVES THAT THE HOLE HAD ESSENTIALLY ACTED AS A CHEESE GRATER IN REMOVING THIS PIECE OF CARTILAGE ON THE PATELLA DURING A TRIAL RANGE OF MOTION. THE SURGEON INSPECTED THE FEMORAL TRIAL AND COULD NOT IDENTIFY A HIGH POINT. THE SURGEON HAS REQUESTED TO CLOSELY INSPECT THE TRIAL FOR ANY DEFECT THAT COULD HAVE LED TO THE INJURY AND SUGGESTED THAT THE TRIAL LUG HOLES SHOULD BE CHANGED IN DESIGN TO AVOID THIS. AT THIS MOMENT IT REMAINS UNCLEAR IF THERE WAS A DELAY AND HOW THE SURGERY WAS COMPLETED. THE PATIENTS CURRENT HEALTH STATUS REMAINS UNCLEAR.
Patient problems: Cartilage damage
MAUDE report 17684497 ↗MalfunctionJul 21, 2021Crack; Crack
IT WAS REPORTED THAT, AFTER A TKA, DURING EQUIPMENT STERILIZATION, A GII SPC REAMTHRU CR TR SZ 6 LT WAS NOTICED TO HAVE A SMALL CRACK IN THE BACK SIDE. AS THIS WAS NOTICED AFTER PROCEDURE, THE SURGERY HAD BEEN COMPLETED WITHOUT ANY SURGICAL DELAY, WITH SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12201261 ↗MalfunctionMay 13, 2021Crack; Crack
IT WAS REPORTED THAT, BEFORE A TKA, A GII SPC REAMTHRU CR TR SZ 6 LT WAS NOTICED TO BE CRACKED. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S + N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11817296 ↗MalfunctionMay 10, 2021Crack; Crack
IT WAS REPORTED THAT, DURING A ROUTINE INSPECTION A GII SPC REAMTHRU CR TR SZ 6 LT (CASE: (B)(4)) AND A GII SPC REAMTHRU CR TR SZ 7 RT (CASE: (B)(4)) WERE FOUND CRACKED. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11803272 ↗MalfunctionMar 24, 2021Break; Break
IT WAS REPORTED THAT, DURING TKA SURGERY IT WAS NOTICED THAT THE METAL IS BREAKING AROUND THE NOTCH ON THE GII SPC REAMTHRU CR TR SZ 6 LT. INCIDENT OCCURRED OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11566388 ↗MalfunctionJan 14, 2021Crack; Naturally Worn; Crack; Naturally Worn
IT WAS REPORTED THAT A GENESIS II SPC REAMTHRU CR TR SIZE 6 LEFT IS WEARN. NO HARM OR INJURY REPORTED ON PATIENT. PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. NO DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11168685 ↗MalfunctionNov 4, 2020Crack; Crack
IT WAS REPORTED THAT, WHILE INSPECTING INSTRUMENTS, A SLIGHT CRACK WAS NOTICED IN THE TRIALS. NO PATIENT WAS INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10788358 ↗MalfunctionSep 25, 2020Break; Break
IT WAS REPORTED THAT AFTER PROCEDURE THE GII SPC REAMTHRU CR TR SZ 6 LT WAS FOUND TO BE CRACKED. NO PATIENT AFFECTATION. NO DELAY. NO BACK UP APPLICABLE FOR THIS CASE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10590453 ↗MalfunctionAug 10, 2020Break; Break
IT WAS REPORTED THAT AFTER SURGERY THE GII SPC REAMTHRU CR TR SZ 6 LT WAS FOUND BROKEN. NO DELAY. NO BACK UP NEEDED. NO INJURIES REPORTED AT THIS TIME. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10388270 ↗MalfunctionJun 2, 2020Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT DURING A TKA PROCEDURE, THE HOUSING REAMER COULD NOT BE REMOVED FROM THE FEMORAL TRIAL. THE SCREWS ON A POSTERIOR STABILIZED REAMER HOUSING SEPARATED. ANOTHER NARROW FEMORAL TRIAL SIZE 6 WAS USED TO FINISH THE PROCEDURE WITHOUT DELAYS. PATIENT WAS NOT HARMED IN ANY WAY. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10109101 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.
- Z-0878-2020 · Dec 16, 2019 · Open, ClassifiedLEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
- Z-0200-2019 · Sep 18, 2018 · TerminatedSmith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
- Z-2163-2018 · May 4, 2018 · Terminatedsmith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
- Z-1902-2018 · Apr 9, 2018 · Terminatedsmith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a
Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- R3 71363077Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER461 reports
- NA 71364009Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER174 reports
- JOURNEY 74012415Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012416Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012414Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER129 reports
- GENESIS II 71440194Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER114 reports
- GENESIS II 71440842Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER113 reports
- JOURNEY 74018901Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page