Skip to content

GENESIS II 71433347Recalls & reported problems

Smith & Nephew, Inc. · GENESIS II SYMMETRICAL POSTERIOR CONDYLES REAM THROUGH CRUCIATE RETAINING TRIAL SIZE 7 LEFT · Class 2

All models of GENESIS II (462)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00885556032398), received between Aug 7, 2020 and Apr 11, 2022 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break3
  • Crack2
  • Fracture1
  • Material Deformation1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2022Crack; Crack

    IT WAS REPORTED THAT THE GII SPC REAMTHRU CR TR SZ 7 LT WAS NOTICED TO BE CRACKED WHERE THE PS CAM LOCKS ONTO IT. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14072177 ↗
  • MalfunctionFeb 1, 2022Break; Break

    IT WAS REPORTED THAT THE LATERAL LUG HOLE OF A GENESIS II SYMMETRICAL POSTERIOR CONDYLES REAM THROUGH CRUCIATE RETAINING TRIAL SIZE 7 LEFT IS BROKEN. THE CONTEXT IN WHICH THIS OBSERVATION WAS MADE IS UNSPECIFIED. THEREFORE, PATIENT INVOLVEMENT IS YET TO BE CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13415501 ↗
  • MalfunctionSep 14, 2021Fracture; Material Deformation; Mechanical Jam; Fracture; Material Deformation; Mechanical Jam

    IT WAS REPORTED THAT, DURING TKA SURGERY, A GII SPC P/S HOUS COLLET S 3-8 GOT STUCK IN THE GII SPC REAMTHRU CR TR SZ 7 LT. THE SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A BACK-UP DEVICE. THE PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM. THE INVESTIGATION FOUND THAT THE FIXATION RAILS IN THE REAM THROUGH SLOT OF THE GII SPC REAMTHRU CR TR SZ 7 LT WERE FRACTURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12467410 ↗
  • MalfunctionSep 25, 2020Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE FEMORAL TRIALS WERE FOUND CRACKED. THIS WAS DISCOVERED AFTER THE CASE. NO PATIENT WAS AFFECTED OR DELAYS WERE REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 10587970 ↗
  • MalfunctionAug 7, 2020Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE THE GENESIS II SPC REAMTHRU CR TR SIZE 7 LEFT WAS DISCOVERED TO BE CRACKED. NO HARM TO THE PATIENT OR STAFF WAS REPORTED. NO PROCEDURAL DELAY, BECAUSE A FUNCTIONING BACKUP WAS AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10382333 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.

  • Z-0878-2020 · Dec 16, 2019 · Open, Classified
    LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
  • Z-0200-2019 · Sep 18, 2018 · Terminated
    Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
  • Z-2163-2018 · May 4, 2018 · Terminated
    smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
  • Z-1902-2018 · Apr 9, 2018 · Terminated
    smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a

Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page