Skip to content

LEGION 71453203Recalls & reported problems

Smith & Nephew, Inc. · LEGION POSTERIOR STABILIZED HIGH FLEXION HIGHLY CROSS LINKED POLYETHYLENE SIZE 1-2 13MM · Class 2 · Implant

All models of LEGION (446)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00885556034804), received between Sep 18, 2017 and Feb 16, 2026 — classified as 6 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K071071, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Detachment Of Device Component1
  • Loose or Intermittent Connection1
  • Manufacturing, Packaging or Shipping Problem1
  • Material Fragmentation1
  • Misconnection1
  • Naturally Worn1
  • Positioning Failure1

Patient problems

  • Injury3
  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • No Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A FIRST REVISION SURGERY HAD BEEN PERFORMED ON (B)(6) 2024, THE PATIENT UNDERWENT A SECOND REVISION SURGERY ON (B)(6) 2026 FOR UNKNOWN REASONS, IN WHICH ONE (1) LGN PS HIGH FLEX XLPE SZ 1-2 13MM WAS EXPLANTED AND REPLACED WITH A LGN PS HIGH FLEX XLPE SZ 1-2 9MM. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 24344028 ↗
  • InjuryOct 25, 2022Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A TKA SURGERY WAS PERFORMED ON (B)(6) 2016, THE PATIENT PRESENTED TO THE DOCTOR AFTER FALL. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2022, IN WHICH A LGN PS HIGH FLEX XLPE SZ 1-2 13MM WAS EXCHANGED WITH A 1/2 X 15MM LEGION XLPE HF INSERT.. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 15665713 ↗
  • MalfunctionSep 9, 2022Positioning Failure; Material Fragmentation; Positioning Failure; Material Fragmentation

    IT WAS REPORTED THAT, DURING TOTAL KNEE ARTHROPLASTY, A LEGION POLYETHYLENE INSERT SIZE 1-2 13MM FAILED TO BE POSITIONED. UPON ATTEMPTING TO INSERT THE IMPLANT, A LITTLE FLAKY ON THE INSERT WAS CAUSED. NO FLAKES FELL INTO THE PATIENT. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15393498 ↗
  • MalfunctionJan 26, 2021Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE SURGEON REQUESTED A 1/2 13MM INSERT AND WHEN OPENING THE BOX THE STERILE INSERT WAS A 3/4 11MM INSERT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH & NEPHEW BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11231201 ↗
  • InjuryJun 16, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A TKA SURGERY HAD BEEN PERFORMED, THE PATIENT EXPERIENCED AN ACCIDENT IN WHICH THE LEGION PS HIGH FLEX XLPE INSERT FRACTURED. THE IMPLANTS WERE NOT DEEMED AS DEFECTIVE BY THE SURGEON. THE INSERT WAS EXPLANTED IN A REVISION SURGERY. THE PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Injury

    MAUDE report 10158131 ↗
  • InjuryApr 24, 2019Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT LEFT KNEE REVISION SURGERY WAS PERFORMED DUE TO LOOSENING OF THE TIBIA COMPONENT.

    Patient problems: Injury

    MAUDE report 8547008 ↗
  • InjuryMar 3, 2018Detachment Of Device Component; Detachment Of Device Component

    AS UPDATED BY CUSTOMER, PATIENT IS NOT WILLING TO HAVE THE REVISION SURGERY.

    Patient problems: No Information

    MAUDE report 7312255 ↗
  • InjurySep 18, 2017Misconnection; Misconnection

    IT WAS REPORTED THAT A RIGHT TIBIAL REVISION SURGERY WAS PERFORMED DUE TO INSERT NOT ENGAGING WITH THE TIBIAL BASEPLATE.

    Patient problems: Injury

    MAUDE report 6874409 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071071, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1507-2020 · Dec 19, 2019 · Terminated
    LEGION PS High Flex XLPE size 5-6 9mm Articular Insert
  • Z-1904-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 10 MM, REF 71453185 designed for use in patients in primary and revision surgery, where the anterior and post
  • Z-1905-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 11 MM, REF 71453122 designed for use in patients in primary and revision surgery, where the anterior and post
  • Z-1906-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 9 MM, REF 71453121 designed for use in patients in primary and revision surgery, where the anterior and poste
  • Z-1907-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 1-2 Size, 9 MM, REF 71453101designed for use in patients in primary and revision surgery, where the anterior and poster

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates