Skip to content

JOURNEY 74025718Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY II INSERT XLPE DEEP DISHED RIGHT SIZE 1-2 21MM · Class 2 · Implant

All models of JOURNEY (566)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556096765), received between May 22, 2017 and May 22, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

1 other recall cover devices cleared under the same FDA submission (K113482, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Operates Differently Than Expected1
  • Unstable1

Patient problems

  • Failure of Implant1
  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 22, 2017Unstable; Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED THAT PATIENT UNDERWENT A REVISION SURGERY TO REPLACE A KNEE IMPLANT TO A HINGE KNEE DUE TO GLOBAL INSTABILITY.

    Patient problems: Failure of Implant; No Code Available

    MAUDE report 6580941 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113482, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page