Skip to content

NA 71366517Recalls & reported problems

Smith & Nephew, Inc. · FEMORAL HEAD TRIAL 12/14 TAPER40 +0 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Smith & Nephew femoral head trial component is a plastic sizing trial used during hip and knee replacement surgery to fit the implant before the real part is placed. The 35 reports mostly describe plastic tabs or locking sides snapping off or chipping during the operation, with fragments sometimes left unaccounted for until surgeons finish with a backup instrument. There are 4 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 00885556110119), received between Feb 17, 2019 and Feb 3, 2026 — classified as 4 injuries, 31 malfunctions.

The most frequently reported specific device problem is “Break” (24 reports).

4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break24
  • Fracture10
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • No Consequences Or Impact To Patient4
  • Device Embedded In Tissue or Plaque1
  • Failure of Implant1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Fracture

    IT WAS REPORTED THAT, DURING A THA SURGERY, A TINE ON THE INSIDE OF ONE (1) FEMORAL HEAD TRIAL 12/14 40 +0 BROKE OFF. IT IS UNKNOWN IF ANY PIECE FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246704 ↗
  • MalfunctionDec 5, 2024Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY, THE ENGAGEMENT AREA WERE THE HEAD SEATS IN ONE (1) FEMORAL HEAD TRIAL 12/14 40 +0 BROKE. NO PIECES WERE LEFT IN THE PATIENT, THE BROKEN PLASTIC FROM THE HEAD WAS IDENTIFIED, REMOVED WITH A PICK UP AND ACCOUNTED FOR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20854250 ↗
  • MalfunctionDec 5, 2024Break; Break

    IT WAS REPORTED THAT DURING A TKA SURGERY, THE FEMORAL HEAD TRIAL 12/14 40 +0 BROKE DURING THE RELOCATION OF THE HIP. THE PROCEDURE WAS PERFORMED, AFTER NO DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20854190 ↗
  • MalfunctionJan 20, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR, THE TABS ON THE INSIDE OF ONE (1) FEMORAL HEAD TRIAL 12/14 40 +0 BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18548859 ↗
  • MalfunctionDec 18, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING THR SURGERY, IT WAS NOTICED THAT ONE (1) FEMORAL HEAD TRIAL 12/14 40 +0 HAD THE PLASTIC CHIPPED. THE PROCEDURE HAD BEEN COMPLETED, WITHOUT ANY DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18352289 ↗
  • MalfunctionOct 30, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR SURGERY, THE INSIDE PART OF THE FEMORAL HEAD TRIAL 12/14 40 +0 CHIPPED SO IT DOES NOT SIT PROPERLY ON THE NECK OF THE TRIAL/STEM. NOTHING WAS FOUND INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18034759 ↗
  • MalfunctionJul 11, 2023Break; Break

    IT WAS REPORTED THAT DURING A THR SURGERY, ONE OF THE 4 INNER LOCKING SIDES OF A FEMORAL HEAD TRIAL 12/14 TAPER40 +0 WAS BROKEN OFF. THERE WAS A NON-SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17302855 ↗
  • MalfunctionJun 20, 2023Fracture; Fracture

    IT WAS REPORTED, THAT DURING A THR SURGERY. THE TAB BROKE OFF INSIDE THE FEMORAL HEAD TRIAL 12/14 40 +0. ALL PIECES WERE RETRIEVED. THE PROCEDURE WAS AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17162221 ↗
  • MalfunctionJun 10, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR, THE INNER TAB OF ONE (1) FEMORAL HEAD TRIAL 12/14 40 +0 BROKE DURING TRAILING. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17102155 ↗
  • MalfunctionMar 3, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, IT WAS NOTICED THAT ONE OF THE TABS ON THE INSIDE OF ONE (1) FEMORAL HEAD TRIAL 12/14 40 +0 BROKE DURING REDUCTION, THE BROKEN PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16482286 ↗
  • MalfunctionDec 6, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, A FEMORAL HEAD TRIAL 12/14 40 +0 IS CHIPPED ON INSIDE. A PICTURE OF THE REPORTED DEVICE SHOWS ONE OF THE FOUR PIECES OF THE LOCKING MECHANISM HAS FRACTURED OFF. SURGERY WAS FINISHED WITHOUT ANY DELAY WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15926028 ↗
  • MalfunctionOct 4, 2022Fracture; Fracture

    IT WAS REPORTED THAT, THE INNER LIPS OF A FEMORAL HEAD TRIAL 12/14 40 +0 WERE BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15535480 ↗
  • MalfunctionSep 15, 2022Fracture; Fracture

    IT WAS REPORTED THAT ONE (1) FEMORAL HEAD TRIAL 12/14 40 +0 WAS BROKEN. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15434658 ↗
  • InjuryJul 21, 2022Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY THE FEMORAL HEAD TRIAL 12/14 40 +0 INNER TAB BROKE DURING TRIAL REDUCTION. ALL PIECES RECOVERED, NO PATIENT AFFECTATION. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY. NO FURTHER COMPLICATION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15076050 ↗
  • MalfunctionJun 9, 2022Break; Break

    IT WAS REPORTED THAT THE FEMORAL HEAD TRIAL 12/14 40 +0 BROKE. AS THIS WAS FOUND OUT DURING FACILITY INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14644930 ↗
  • MalfunctionApr 6, 2022Break; Break

    IT WAS REPORTED THAT A FEMORAL HEAD TRIAL 12/14 40 +0 WAS FOUND BROKEN. AS THIS WAS NOTICED IN A FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14022085 ↗
  • MalfunctionMar 1, 2022Break; Break

    IT WAS REPORTED THAT THE INSIDE OF A FEMORAL HEAD TRIAL 12/14 TAPER40 +0 BROKE. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13636253 ↗
  • MalfunctionFeb 9, 2022Break; Break

    IT WAS REPORTED THAT THE FEMORAL HEAD TRIAL 12/14 40 +0 BROKE ON THE INSIDE. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13502764 ↗
  • InjuryDec 8, 2021Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY, THE INNER TABS OF FEMORAL HEAD TRIAL 12/14 40 +0 BROKE OFF INSIDE THE PATIENT. ALL PARTS ACCOUNTED FOR. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE. SURGERY WAS NOT DELAYED. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12964935 ↗
  • MalfunctionNov 15, 2021Break; Break

    IT WAS REPORTED THAT THE INTERNAL NECK OF A TRIAL FEMORAL HEAD 40 S/+0 IS BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12814196 ↗
  • MalfunctionJun 9, 2021Break; Break

    IT WAS REPORTED THAT THE FEMORAL HEAD TRIAL 12/14 40 +0 BROKE. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11968571 ↗
  • MalfunctionMay 31, 2021Break; Break

    IT WAS REPORTED THAT A FEMORAL HEAD TRIAL 12/14 40 +0 WAS BROKEN. SINCE IT WAS NOTICED DURING A FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11909322 ↗
  • InjuryApr 25, 2021Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY, THE PLASTIC INSIDE THE 40MM FEMORAL HEAD TRIAL +0 BROKE OFF WHILE BEING USED INSIDE THE PATIENT. THE PROCEDURE FINISHED WITH THE SAME DEVICE AND IT WAS DELAYED LESS THAN OR EQUAL TO 30MIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11722119 ↗
  • MalfunctionApr 19, 2021Break; Break

    IT WAS REPORTED THAT DURING INSPECTION THE INTERNAL CAPTURE PIECE OF THE FEMORAL HEAD TRIAL 12/14 40 +0 SNAPPED OFF. NO CASE INVOLVED THEREFORE NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11683761 ↗
  • MalfunctionMar 8, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT, THE FEMORAL HEAD TRIAL 12/14 40 +0 CAME BACK IN OFFICE SET WITH LARGE GASH TAKEN OUT OF IT. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431676 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.

  • Z-0878-2020 · Dec 16, 2019 · Open, Classified
    LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
  • Z-0200-2019 · Sep 18, 2018 · Terminated
    Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
  • Z-2163-2018 · May 4, 2018 · Terminated
    smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
  • Z-1902-2018 · Apr 9, 2018 · Terminated
    smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a

Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page