NA 71363076Recalls & reported problems
Smith & Nephew, Inc. · FEMORAL HEAD REDUCER · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00885556111024), received between Sep 3, 2019 and Dec 29, 2025 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 29, 2025Break; Break
IT WAS REPORTED THAT DURING A THR SURGERY, ONE (1) FEMORAL HEAD REDUCER BROKE UNDER NORMAL USE. THE PROCEDURE WAS COMPLETED, AFTER NO DELAY, CHANGING THE SURGICAL TECHNIQUE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23919177 ↗MalfunctionMay 23, 2025Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, ONE (1) FEMORAL HEAD REDUCER BROKE WHILE USING THE INSTRUMENT TO RELOCATE THE HEAD. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22078429 ↗MalfunctionMay 13, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A THR (POSTERIOR APPROACH) SURGERY, THE END OF ONE (1) FEMORAL HEAD REDUCER BROKE OFF DURING THE MOMENT OF REDUCTION. IT DID NOT FALL INTO THE WOUND. THE DOCTOR SIMPLY RETRIEVED THE PIECE WITH HIS FINGERS. IT WAS VERY CLEAR THAT THE BROKEN PIECE AND THE ORIGINAL PIECE FIT BACK TOGETHER, SO WE KNEW THERE WERE NO ADDITIONAL PIECES LEFT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22011070 ↗MalfunctionJul 30, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING THR SURGERY, IT WAS NOTICED THAT, THE PLASTIC WORE OUT AND THE FEMORAL HEAD REDUCER SNAPPED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19861714 ↗MalfunctionMay 15, 2024Break; Break
IT WAS REPORTED THAT, DURING A THR, ONE (1) FEMORAL HEAD REDUCER BROKE WHEN SLIDING THE HEAD INTO THE CUP. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19322630 ↗InjurySep 29, 2021Break; Break
IT WAS REPORTED DURING A THR SURGERY, THAT WHEN THE SURGEON WENT TO REDUCE THE FEMORAL HEAD USING A FEMORAL HEAD REDUCER, TOO MUCH TORQUE WAS APPLIED ON THE INSTRUMENT AND IT SNAPPED IN HALF. NO PIECES FELL INTO THE PATIENT, NO INJURY REPORTED. PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE, NO DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12551073 ↗InjuryJul 2, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, AS THE SURGEON WENT TO REDUCE THE HIP, HE LEVERED ON THE AFFECTED DEVICE CAUSING IT TO BREAK. ALL BROKEN PIECES WERE RETRIEVED. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12111780 ↗MalfunctionSep 3, 2019Break; Break
IT WAS REPORTED THAT THE ITEM WAS BROKEN DURING REDUCTION. NO IMPACT/INJURY TO PATIENT, S&N BACKUP AVAILABLE, NO PIECES/PARTS LEFT IN PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8955128 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.
- Z-0878-2020 · Dec 16, 2019 · Open, ClassifiedLEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
- Z-0200-2019 · Sep 18, 2018 · TerminatedSmith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
- Z-2163-2018 · May 4, 2018 · Terminatedsmith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
- Z-1902-2018 · Apr 9, 2018 · Terminatedsmith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a
Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- R3 71363077Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER461 reports
- NA 71364009Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER174 reports
- JOURNEY 74012416Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012415Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012414Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER129 reports
- GENESIS II 71440194Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER114 reports
- GENESIS II 71440842Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER113 reports
- JOURNEY 74018901Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page