SPATIAL FRAME 71070364Recalls & reported problems
Smith & Nephew, Inc. · EQUATOR CLAMP · Class 2 · Implant
All models of SPATIAL FRAME (19)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00885556119181), received between Feb 24, 2020 and Feb 28, 2022 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Break” (13 reports).
1 other recall cover devices cleared under the same FDA submission (K093047); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break13
- Solder Joint Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- No Patient Involvement2
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 28, 2022Break; Break
IT WAS REPORTED THAT THE EQUATOR CLAMP BROKE IN HALF. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13625561 ↗MalfunctionFeb 9, 2022Break; Break
IT WAS REPORTED THAT THE EQUATOR CLAMP WAS BROKEN. SINCE THIS WAS NOTICED DURING A FIELD INSPECTION, NO CASE WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13500032 ↗MalfunctionNov 17, 2021Break; Solder Joint Fracture; Break; Solder Joint Fracture
IT WAS REPORTED THAT A EQUATOR CLAMP WAS FOUND TO BE BROKEN. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12830783 ↗MalfunctionAug 9, 2021Break; Break
IT WAS REPORTED THAT A WIRE CUTTER HAVE WEAR AND DAMAGE AND A EQUATOR CLAMP IS BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12297599 ↗MalfunctionJul 13, 2021Break; Break
IT WAS REPORTED THAT AN EQUATOR CLAMP WAS FOUND TO BE BROKEN INSIDE THE PACKAGE. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12159254 ↗MalfunctionJun 23, 2021Break; Break
IT WAS REPORTED THAT, DURING AN EXTERNAL FIXATION SURGERY, THE KNOB TO TIGHTEN AN EQUATOR CLAMP BROKE OFF WHILE INSTRUMENTS WERE OUTSIDE THE PATIENT. THE SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12053811 ↗MalfunctionSep 30, 2020Break; Break
IT WAS REPORTED THAT WAS WHILE PREPARING THE INSTRUMENTATION ON THE BACK TABLE BEFORE THE CASE STARTED THE EQUATOR CLAMP BROKE. PATIENT NOT INJURED. DELAY REPORTED LESS OR EQUAL THAN 30 MINUTES. S+N BACK-UP DEVICE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10609699 ↗MalfunctionSep 11, 2020Break; Break
IT WAS REPORTED THAT THE PLASTIC PIECE THAT HOLDS THE METAL PIN BROKE OFF. NO INJURY, NO DELAY, S & N BACKUP AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10521937 ↗MalfunctionJun 29, 2020Break; Break
IT WAS REPORTED THAT THE DEVICE WAS FOUND BROKEN. IT IS UNKNOWN WHEN THIS HAPPEN OR IF THERE WAS A CASE INVOLVED. NO INJURY REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10209745 ↗MalfunctionJun 6, 2020Break; Break
IT WAS REPORTED THAT THE EQUATOR CLAMP BROKE WHILE BEING CLEANED AFTER PROCEDURE. NO BACKUP DEVICE WAS REQUIRED AND NO SURGICAL DELAY REPORTED SINCE THE PROCEDURE WAS ALREADY FINISHED AS INTENDED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 10126560 ↗MalfunctionFeb 24, 2020Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE EQUATOR CLAMP WAS FOUND BROKEN IN THE SET, SO REGULAR BOLT WAS USED TO SERVE ITS FUNCTION (WHICH IS NOT INTENDED FOR THIS REASON). THERE WERE NO DELAYS IN THE PROCEDURE AND NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9747266 ↗MalfunctionFeb 24, 2020Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE EQUATOR CLAMP WAS FOUND BROKEN IN THE SET, SO REGULAR BOLT WAS USED TO SERVE ITS FUNCTION (WHICH IS NOT INTENDED FOR THIS REASON). THERE WERE NO DELAYS IN THE PROCEDURE AND NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9747259 ↗MalfunctionFeb 24, 2020Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE EQUATOR CLAMP WAS FOUND BROKEN IN THE SET, SO REGULAR BOLT WAS USED TO SERVE ITS FUNCTION (WHICH IS NOT INTENDED FOR THIS REASON). THERE WERE NO DELAYS IN THE PROCEDURE AND NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9747255 ↗
FDA clearance and classification
Cleared through 510(k) K093047 on Sep 27, 2010 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew Circular Fixation System”. The FDA review took 362 days. FDA record ↗
The FDA regulates product code KTT (appliance, fixation, nail/blade/plate combination, multiple component) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K093047. One submission often covers many models, so check the product and lot details in each record.
- Z-0245-2022 · Jun 17, 2021 · Open, ClassifiedIlizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
Common questions about the SPATIAL FRAME 71070364
Has the SPATIAL FRAME 71070364 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00885556119181), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the SPATIAL FRAME 71070364?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between Feb 24, 2020 and Feb 28, 2022. In those reports, the most common specific device problems are “Break” (13) and “Solder Joint Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SPATIAL FRAME 71070364?
None of the reports are classified as deaths, 0 as injuries and 13 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SPATIAL FRAME 71070364 FDA approved?
It was cleared through 510(k) K093047 on Sep 27, 2010. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the SPATIAL FRAME 71070364 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the SPATIAL FRAME 71070364, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand SPATIAL FRAME, model 71070364). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page